Article Abstract

Background And Objective: The use of bone allograft reconstructions after tumor resection can introduce significant complications. Stable fixation is required to decrease the incidence of mechanical complications of segmental bone allografts. The purpose of the present study is to compare plating fixation methods of diaphyseal allografts after intercalary resection of the femur.

Methods: We created four defined fixation models using plates and/or intramedullary polymethylmethacrylate (PMMA) to simulate typical bone tumor resection with intercalary allograft reconstruction. One angularly stable plate (DFP) with 13 locking screws and fresh frozen allografts (labeled "I") were used for bone reconstruction. Three modified reconstructions were created: "II" included a supplementary plate (SP) with four locking screws, "III" was augmented with intramedullary PMMA in the allograft, and "IV" combined intramedullary PMMA and both plates. We applied a load model that simulates partial weight bearing on the lower limb to simulate the load during postoperative rehabilitation.

Results: The highest stress in the DFP occurred at the allograft-bone transition, with variant IV reaching 297 MPa. PMMA augmentation reduced median interfragmentary motion (IFM) and sliding distances, with variant III achieving the lowest distal sliding distance (0.9 μm) in the distal area. Supplementary plate fixation reduced maximal and median proximal IFM distances (86.9 μm in variant II vs. 116.0 μm in variant I) but increased sliding distances (23.7 μm in variant II vs. 0.6 μm in variant I).

Conclusions: PMMA augmentation reduces IFM and sliding distances, enhancing rigidity, particularly in the distal area. Supplementary plate fixation decreases IFM distances in the proximal area but increases sliding distances in the same region. Variants III and IV demonstrate lower IFM and sliding distances in the distal area overall. Variant III shows very low sliding distances in both distal and proximal areas. Variant IV combines improved firmness with slightly higher stress levels.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11801617PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0316719PLOS

Publication Analysis

Top Keywords

sliding distances
24
supplementary plate
16
μm variant
16
tumor resection
12
plate fixation
12
ifm sliding
12
distal area
12
bone allograft
8
allograft reconstruction
8
locking screws
8

Similar Publications

Latitudinal-environmental variations driving the local adaptation of stocks along the Chinese coast.

Mar Life Sci Technol

August 2025

Laboratory of Marine Organism Taxonomy and Phylogeny, Qingdao Key Laboratory of Marine Biodiversity and Conservation, and The Key Laboratory of Experimental Marine Biology, Centre for Ocean Mega-Science, Institute of Oceanology, Chinese Academy of Sciences, Qingdao, 266000 China.

Unlabelled: The distribution of (Euphrasen, 1788) spans a pronounced latitudinal-environmental gradient from the subtropical to the subpolar zones. The species is reported to have multiple stocks along coastal China, exhibiting different spawning behaviors and habitat preferences. Such ecological variations might imply potential genetic divergence and local adaptation.

View Article and Find Full Text PDF

Objective: The study aimed to evaluate the effectiveness of hard court slide braking technique of elite college tennis players and to analyze the factors that influence the use of hard court slide braking technique.

Methods: A total of 100 elite tennis players (58 males and 42 females) participated in tests involving hard court hit-to-return and Forcedecks Dual Force Plates System-based physical function. Independent samples t-tests, paired samples t-tests and binary logistic regression were employed in data analysis.

View Article and Find Full Text PDF

Cavitation erosion could lead to material loss and structural damage in the flow components of hydraulic machinery, affecting the stable operation and efficiency of the equipment. This study established a cavitation erosion distance model for the gap in the cavitation field. High-speed cameras were used to capture the bubble distribution under different cavitation erosion distances.

View Article and Find Full Text PDF

Background: To reduce the risk of local recurrence of papillary thyroid carcinoma (PTC) after thermal ablation, in clinical practice, the ablation area is typically extended 2-3 mm beyond the original tumor margin. However, there is no definitive evidence to support the validity of this extended margin.

Objective: The objective of this study was to measure the distance from the edge of a PTC to the farthest point of tumor infiltration using digital pathology and to analyze the associated risk factors.

View Article and Find Full Text PDF

StaDis: Stability distance to detecting out-of-distribution data in computational pathology.

Med Image Anal

August 2025

School of Computer Science and Technology, Xi'an Jiaotong University, Xi'an, 710049, China; Shaanxi Provincial Key Laboratory of Big Data Knowledge Engineering, Xi'an Jiaotong University, Xi'an, 710049, China. Electronic address:

Modern Computational pathology (CPath) models aim to alleviate the burden on pathologists. However, once deployed, these models may generate unreliable predictions when encountering data types not seen during training, potentially causing a trust crisis within the computational pathology community. Out-of-distribution (OOD) detection, acting as a safety measure before model deployment, demonstrates significant promise in ensuring the reliable use of models in real clinical application.

View Article and Find Full Text PDF