Background: Diving is a diffused recreational activity, and the number of divers carrying cardiac implanted devices is similarly growing. Due to the lack of guidelines or technical indications, the suitability of such devices for diving or the fitness to dive for these patients still needs to be determined.

Objective: This work summarizes implantable cardiac devices' suitability for recreational diving, technical vulnerability factors, and recommendations to improve implanted divers' safety.

Methods: Between May 1, 2021, and March 20, 2022, three interventional cardiologists retrieved the technical documentation of selected implantable cardiac devices. In particular, any suitability and tests conducted in hyperbaric environments were tracked.

Results: Technical documentation was recovered for four companies. Most devices were tested in hyperbaric conditions in single, prolonged, or repeated exposures to pressurized air; underwater tests were not mentioned. No company expressly disclosed the suitability of the devices for underwater activities.

Conclusion: In the absence of technical indications or guidelines, a multidisciplinary evaluation between cardiology, diving medicine, and sports medicine is essential to establish the suitability for underwater sports in implanted patients. Before each diving trip, device control is advisable, and underwater physiological adaptations should be considered, especially in the cardiovascular domain. Stressors other than water and pressure must be considered during diving, such as lead strain caused by arm movements and pressure exerted by suits or buoyancy control devices on the chest. Future directions point towards a follow-up of implanted, active divers and developing leadless devices and underwater telemonitoring.

Download full-text PDF

Source

Publication Analysis

Top Keywords

implantable cardiac
12
devices
8
cardiac devices
8
recreational diving
8
technical indications
8
suitability devices
8
technical documentation
8
devices underwater
8
diving
7
technical
6

Similar Publications

Long Term Outcomes of Tricuspid Valve Repair or Replacement in Patients with Continuous Flow Left Ventricular Assist Devices.

J Thorac Cardiovasc Surg

September 2025

Division of Cardiac, Thoracic & Vascular Surgery, Department of Surgery, Columbia University Irving Medical Center, NewYork-Presbyterian Hospital, New York, NY. Electronic address:

Objective: Our objective was to determine the long-term outcomes of concomitant tricuspid valve procedures (TVP) during continuous-flow left ventricular assist device (LVAD) implantation.

Methods: We retrospectively reviewed patients who received HeartMate II or 3 from 2004 to 2023. Nine patients who had a previous TVP were excluded.

View Article and Find Full Text PDF

Background: Peripartum cardiomyopathy (PPCM) is a rare, life-threatening form of heart failure occurring in late pregnancy or postpartum, with variable clinical course and outcomes. We report preliminary clinical and echocardiographic findings from a national Italian registry of PPCM patients METHODS: The study was approved by the institutional Ethics Committee and registered at ClinicalTrials.gov (NCT05878041).

View Article and Find Full Text PDF

Background: Current recommendations for a prophylactic (primary prevention) implantable cardioverter defibrillator (ICD) in patients with both ischemic and non-ischemic heart failure with reduced ejection fraction (HFrEF) originate from clinical trials conducted in selected patients over 20 years ago that showed an overall statistically significant survival benefit associated with a primary prevention ICD in the range of 23%-34%. The recent introduction of angiotensin receptor-neprilysin inhibitors [ARNI] and sodium glucose co-transporter 2 inhibitors [SGLT2i]) was shown to further reduce the risk of sudden cardiac death (SCD) in patients with HFrEF. Thus, there is an unmet need appropriately designed comparative effectiveness clinical trials aimed to reassess the survival benefit of a primary prevention ICD in contemporary patients with HFrEF.

View Article and Find Full Text PDF

Spanish cardiac catheterization and coronary intervention registry. 34 official report of the Interventional Cardiology Association of the Spanish Society of Cardiology (1990-2024).

Rev Esp Cardiol (Engl Ed)

September 2025

Centro de Investigación Biomédica en Red de Enfermedades Cardiovasculares (CIBERCV), Instituto de Salud Carlos III, Madrid, España; Servicio de Cardiología, Hospital Clínico de Santiago de Compostela, Santiago de Compostela, España.

Introduction And Objectives: This report presents the 2024 activity data from the Interventional Cardiology Association of the Spanish Society of Cardiology (ACI-SEC).

Methods: All interventional cardiology laboratories in Spain were invited to complete an online survey. Data analysis was conducted by an external company and then reviewed and presented by the ACI-SEC board.

View Article and Find Full Text PDF

Myocardial Blood Flow Quantification in Patients with an Implanted Cardiodefibrillator during Stress and at Rest using a Wideband Perfusion Pulse Sequence: An Initial Feasibility Study.

J Cardiovasc Magn Reson

September 2025

Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, IL, USA; Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, IL, USA.

Background: Although a recently developed wideband perfusion sequence has shown diagnostically acceptable image quality and accurate myocardial blood flow (MBF) quantification at rest in patients with cardiac implanted electronic devices (CIEDs), its performance during vasodilator stress remains unproven. This study aims to determine whether the sequence produces diagnostically acceptable image quality during stress and is capable of quantitatively detecting abnormal stress MBF and myocardial perfusion reserve (MPR) in patients with implanted cardiodefibrillators (ICDs).

Methods: We enrolled 29 patients with an ICD (mean age = 63 ± 15 years, 17 males, 12 females) and 11 control patients (mean age = 50 ± 17 years, 6 males, 5 females; negative coronary artery disease; negative stress perfusion CMR; and no cardiac event one year post CMR) with an ICD taped below the left clavicle to mimic image artifacts.

View Article and Find Full Text PDF