Article Abstract

Background: Clinical experience during trans-patch ventricular ablations shows difficult lesion formation when using standard RF ablation technology. The aim of this exploratory study is to test if ultralow temperature cryoblation (ULTC) technology could be suitable for trans-patch ventricular ablations.

Materials And Methods: Ex vivo ablations were performed on 16 porcine tenderloin specimens of 5 × 7 × 3 cm, eight with and eight without pericardial patch (Supple Peri-Guard Repair Patch, Baxter, USA). RF energy was applied for 1 min, with a power of 40W. ULTC was applied as a 2 × 2 min freeze. Temperature was measured using an array of thermocouples ranging between 0.5 and 13.1 mm below the tissue surface.

Results: For RF, the maximum temperatures without and with patch respectively reached 58.9 °C and 48.5 °C. The thermocouple closest to the tissue surface displayed a 41% temperature range drop when RF energy was applied trans-patch, whereas the same thermocouple showed a 16% and 13% drop for the first and second ULTC freeze cycle when ULTC was applied trans-patch. The 45 °C isotherm depth was reduced by 55% for RF ablations, whereas the - 25 °C isotherm depth was reduced by 23% and 19% for the first and second freeze cycles, respectively.

Conclusion: Both energy modalities show reduced efficacy for trans-patch ablations when considering absolute temperature differences and estimated lesion depths. However, when considering relative differences, ULTC is more effective than RF for trans-patch lesions. Additionally, ULTC did not show visible damage to either the patch or the ex vivo specimen, while RF resulted in visible damage to both the tissue and patch.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10840-024-01952-2DOI Listing

Publication Analysis

Top Keywords

ultralow temperature
8
standard ablation
8
trans-patch ventricular
8
energy applied
8
ultc applied
8
applied trans-patch
8
isotherm depth
8
depth reduced
8
visible damage
8
trans-patch
6

Similar Publications

Study on the high viscosity and gel-prone properties of glycine-amidated pectin and its regulatory role on the freeze-thaw stability of sea bass surimi.

Food Chem

September 2025

The Engineering Technological Center of Mushroom Industry, School of Biological Science and Biotechnology, Minnan Normal University, Zhangzhou 363000, PR China; Food Science Research Institute of Zhangzhou, Minnan Normal University, Zhangzhou 363000, China. Electronic address:

An ultra-low temperature (-5 °C) enzymatic method was employed to prepare glycine-amidated pectin (Gly-Pe) and evaluate its physicochemical properties and freeze-thaw protection mechanism in surimi. After glycine grafting (12.77 %), amide bonds disrupted pectin's crystalline structure and enhanced molecular chain flexibility.

View Article and Find Full Text PDF

Homo-layer flexible BiTe-based films with high thermoelectric performance.

Sci Adv

September 2025

Department of Physics, State Key Laboratory of Quantum Functional Materials, and Guangdong Basic Research Center of Excellence for Quantum Science, Southern University of Science and Technology, Shenzhen 518055, China.

Here, we demonstrate unconventional scalable and sustainable manufacturing of flexible n-type BiTe films via physical vapor deposition and homo-layer fusion engineering. The achieved ultrahigh power factor of up to 30.0 microwatts per centimeter per square kelvin and ultralow lattice thermal conductivity of 0.

View Article and Find Full Text PDF

The development of flexible gas sensors is of growing interest in wearable electronics. However, developing a gas sensor with low operating temperature, high sensitivity, and rapid response remains a huge challenge. Herein, we first develop a polyacrylamide-sodium acrylate-sodium citrate (PAM-Na-SC) hydrogel electrolyte, and design a hydrogel-based nitrogen dioxide (NO) gas sensor enabled by zinc-air batteries (ZABs).

View Article and Find Full Text PDF

Continuous Proton Channels in Vanadium Hexacyanoferrate Enable All-Round Improved Ultra-Low-Temperature Energy Conversion.

Small

September 2025

Key Laboratory for New Functional Materials of Ministry of Education, College of Materials Science and Engineering, Beijing University of Technology, Beijing, 100124, China.

Proton-based batteries emerge as a promising candidate for low-temperature energy conversion and storage, yet their practical implementation is constrained by cathode performance limitations. Herein, [Fe(CN)] vacancies-mediated vanadium hexacyanoferrate (VFeCN-VHCF) with contiguous proton channels is pioneered as a proton cathode material for all-round improved ultra-low-temperature energy conversion. Through operando characterization and density functional theory calculations, we reveal that the [Fe(CN)] vacancies induce the formation of V═O bonds, which serve as the active sites to store protons and endow V-VHCF to reach a high capacity.

View Article and Find Full Text PDF

Fundamentally understanding lattice dynamics and thermal transport behavior in liquid-like, partially occupied compounds remains a long-standing challenge in condensed matter physics. Here, the microscopic mechanisms are investigated underlying the ultralow thermal conductivity in ordered/liquid-like CuBiS by combining experimental methods with first-principles calculations. The ordered structure and liquid-like are first experimentally synthesized and characterized, partially Cu-atom occupied CuBiS structure with increasing temperature.

View Article and Find Full Text PDF