Article Abstract

Purpose: Present knowledge of patient setup and alignment errors in image guided radiation therapy (IGRT) relies on voluntary reporting, which is thought to underestimate error frequencies. A manual retrospective patient-setup misalignment error search is infeasible owing to the bulk of cases to be reviewed. We applied a deep learning-based misalignment error detection algorithm (EDA) to perform a fully automated retrospective error search of clinical IGRT databases and determine an absolute gross patient misalignment error rate.

Methods And Materials: The EDA was developed to analyze the registration between planning scans and pretreatment cone beam computed tomography scans, outputting a misalignment score ranging from 0 (most unlikely) to 1 (most likely). The algorithm was trained using simulated translational errors on a data set obtained from 680 patients treated at 2 radiation therapy clinics between 2017 and 2022. A receiver operating characteristic analysis was performed to obtain target thresholds. DICOM Query and Retrieval software was integrated with the EDA to interact with the clinical database and fully automate data retrieval and analysis during a retrospective error search from 2016 to 2017 and from 2021 to 2022 for the 2 institutions, respectively. Registrations were flagged for human review using both a hard-thresholding method and a prediction trending analysis over each individual patient's treatment course. Flagged registrations were manually reviewed and categorized as errors (>1 cm misalignment at the target) or nonerrors.

Results: A total of 17,612 registrations were analyzed by the EDA, resulting in 7.7% flagged events. Three previously reported errors were successfully flagged by the EDA, and 4 previously unreported vertebral body misalignment errors were discovered during case reviews. False positive cases often displayed substantial image artifacts, patient rotation, and soft tissue anatomy changes.

Conclusions: Our results validated the clinical utility of the EDA for bulk image reviews and highlighted the reliability and safety of IGRT, with an absolute gross patient misalignment error rate of 0.04% ± 0.02% per delivered fraction.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2024.02.065DOI Listing

Publication Analysis

Top Keywords

misalignment error
16
patient misalignment
12
radiation therapy
12
error search
12
misalignment
8
misalignment errors
8
cone beam
8
beam computed
8
retrospective error
8
absolute gross
8

Similar Publications

An interferometer used for displacement measurement is typically adjusted to the center or another appropriate point of the interferometer fringe as a working point to yield maximum detection sensitivity and linearity. The interferometer is prone to varying misalignment in the course of measurements, most noticeable due to thermal drift affecting the interferometer dimensions. We introduce an automatic correction mechanism based on a proportional/integral (PI) control loop to remove any error in the alignment of the fiber interferometer, specifically long-term drift.

View Article and Find Full Text PDF

Assessment and correction of the respiratory phase misalignment using different signals in 4-dimensional imaging and free-breathing gated radiotherapy.

Radiother Oncol

August 2025

Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing 100142, China. Electronic address:

Objective: To propose and validate a correction method for the respiratory phase shift of using different motion management systems in the 4-dimensional (4D) imaging and free-breathing gated radiotherapy.

Materials And Methods: Synchronized cone-beam CT (CBCT) and optical surface images (OSI) of 30 patients at two institutions were included. Reference diaphragm-signals were extracted from CBCT projections using Amsterdam-Shroud (AS) method.

View Article and Find Full Text PDF

Robotic assembly of electrical connectors enables the automation of high-efficiency production of electronic products. A rigid gripper is adopted as the end-effector by the majority of existing works with a force-torque sensor installed at the wrist, which suffers from very limited perception capability of the manipulated objects. Moreover, the grasping and movement actions, as well as the inconsistency between the robot base and the end-effector frame, tend to result in angular misalignment, usually leading to assembly failure.

View Article and Find Full Text PDF

Steering in the dark: The impact of environmental luminance on driver behavior through optical flow analysis.

Atten Percept Psychophys

August 2025

School of Mechanical Engineering, Tianjin University, No.135 Yaguan Road, Jinnan District, Tianjin, 300350, China.

The visual perception and steering behavior of drivers are known to be influenced by environmental lighting, but the underlying perception mechanisms, particularly the role of optical flow under low-luminance conditions, remain insufficiently understood. In a simulated driving experiment, 32 participants were exposed to five controlled luminance levels while their eye movements and driving performance were recorded. The results indicate that lower environmental luminance leads to prolonged gaze duration, a wider distribution of gaze points, and an increase in lateral steering errors.

View Article and Find Full Text PDF

Leveraging Large Language Models to Advance Certification, Physician Learning, and Diagnostic Excellence.

J Am Board Fam Med

August 2025

From the American Board of Family Medicine, Lexington, KY (TW, DWP, AWB); Department of Family Medicine, University of Colorado School Anschutz School of Medicine, Aurora, CO (DWP).

Diagnostic errors are a significant challenge in health care, often resulting from gaps in physicians' knowledge and misalignment between confidence and diagnostic accuracy. Traditional educational methods have not sufficiently addressed these issues. This commentary explores how large language models (LLMs), a subset of artificial intelligence, can enhance diagnostic education by improving learning transfer and physicians' diagnostic accuracy.

View Article and Find Full Text PDF