Article Abstract

Purpose: This work aims to develop a novel distortion-free 3D-EPI acquisition and image reconstruction technique for fast and robust, high-resolution, whole-brain imaging as well as quantitative mapping.

Methods: 3D Blip-up and -down acquisition (3D-BUDA) sequence is designed for both single- and multi-echo 3D gradient recalled echo (GRE)-EPI imaging using multiple shots with blip-up and -down readouts to encode B field map information. Complementary k-space coverage is achieved using controlled aliasing in parallel imaging (CAIPI) sampling across the shots. For image reconstruction, an iterative hard-thresholding algorithm is employed to minimize the cost function that combines field map information informed parallel imaging with the structured low-rank constraint for multi-shot 3D-BUDA data. Extending 3D-BUDA to multi-echo imaging permits mapping. For this, we propose constructing a joint Hankel matrix along both echo and shot dimensions to improve the reconstruction.

Results: Experimental results on in vivo multi-echo data demonstrate that, by performing joint reconstruction along with both echo and shot dimensions, reconstruction accuracy is improved compared to standard 3D-BUDA reconstruction. CAIPI sampling is further shown to enhance image quality. For mapping, parameter values from 3D-Joint-CAIPI-BUDA and reference multi-echo GRE are within limits of agreement as quantified by Bland-Altman analysis.

Conclusions: The proposed technique enables rapid 3D distortion-free high-resolution imaging and mapping. Specifically, 3D-BUDA enables 1-mm isotropic whole-brain imaging in 22 s at 3T and 9 s on a 7T scanner. The combination of multi-echo 3D-BUDA with CAIPI acquisition and joint reconstruction enables distortion-free whole-brain mapping in 47 s at 1.1 × 1.1 × 1.0 mm resolution.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10072851PMC
http://dx.doi.org/10.1002/mrm.29578DOI Listing

Publication Analysis

Top Keywords

joint hankel
8
structured low-rank
8
rapid distortion-free
8
distortion-free high-resolution
8
image reconstruction
8
whole-brain imaging
8
blip-up -down
8
field map
8
parallel imaging
8
caipi sampling
8

Similar Publications

Fast and ultra-high shot diffusion MRI image reconstruction with self-adaptive Hankel subspace.

Med Image Anal

May 2025

Department of Radiology, Zhongshan Hospital, Fudan University (Xiamen Branch), Fujian Province Key Clinical Specialty Construction Project (Medical Imaging Department), Xiamen Key Laboratory of Clinical Transformation of Imaging Big Data and Artificial Intelligence, Xiamen 361006, PR China. Electron

Multi-shot interleaved echo planar imaging is widely employed for acquiring high-resolution and low-distortion diffusion weighted images (DWI). These DWI images, however, are easily affected by motion artifacts induced by inter-shot phase variations which could be removed by enforcing the low-rankness of a huge 2D block Hankel matrix of the k-space. Successful applications have been evidenced on 4∼8 shots DWI but failure was observed on ultra-high shots, e.

View Article and Find Full Text PDF

Purpose: EPI with blip-up/down acquisition (BUDA) can provide high-quality images with minimal distortions by using two readout trains with opposing phase-encoding gradients. Because of the need for two separate acquisitions, BUDA doubles the scan time and degrades the temporal resolution when compared to single-shot EPI, presenting a major challenge for many applications, particularly fMRI. This study aims at overcoming this challenge by developing an echo-shifted EPI BUDA (esEPI-BUDA) technique to acquire both blip-up and blip-down datasets in a single shot.

View Article and Find Full Text PDF

Purpose: This work aims to develop a novel distortion-free 3D-EPI acquisition and image reconstruction technique for fast and robust, high-resolution, whole-brain imaging as well as quantitative mapping.

Methods: 3D Blip-up and -down acquisition (3D-BUDA) sequence is designed for both single- and multi-echo 3D gradient recalled echo (GRE)-EPI imaging using multiple shots with blip-up and -down readouts to encode B field map information. Complementary k-space coverage is achieved using controlled aliasing in parallel imaging (CAIPI) sampling across the shots.

View Article and Find Full Text PDF

Trapezoid-kinoform zone plate lens - a solution for efficient focusing in hard X-ray optics.

J Synchrotron Radiat

March 2022

Shanghai Synchrotron Radiation Facility, Shanghai Advanced Research Institute, Chinese Academy of Sciences, Shanghai 201210, People's Republic of China.

X-ray microscopes are powerful tools in the nano-inspection of materials owing to their ultra-high resolution at the molecular level. However, the focusing efficiency of binary zone plate lenses as key components in such probes is merely 5% in practice, hindering their application in advanced scientific research. Although kinoform zone plate lenses are in principle supposed to possess high efficiency beyond binary ones, little progress has been reported so far due to the shortage of both a theoretical calculation approach and greyscale lithography for generating fine three-dimensional (3D) kinoform zones of the lenses.

View Article and Find Full Text PDF

We propose a multi-slice acquisition with orthogonally alternating phase encoding (PE) direction and subsequent joint calibrationless reconstruction for accelerated multiple individual 2D slices or multi-slice 2D Cartesian MRI. Specifically, multi-slice multi-channel data are first acquired with random or uniform PE undersampling while orthogonally alternating PE direction between adjacent slices. They are then jointly reconstructed through a recently developed low-rank multi-slice Hankel tensor completion (MS-HTC) approach.

View Article and Find Full Text PDF