Article Abstract

Objective: The aim of this study was to investigate low-dose kilovoltage cone-beam CT (kV-CBCT) for image-guided radiotherapy, with a particular focus on the accuracy of image registration with low-dose protocols.

Methods: Imaging doses were measured with a NOMEX semiconductor detector positioned at the front of head, thorax, and pelvis human body phantoms while kV-CBCT scans were acquired at different tube currents. Aspects of image quality (spatial resolution, noise, uniformity, contrast, geometric distortion, and Hounsfield unit sensitivity) and image registration accuracy using bone and soft tissue were evaluated.

Results: With preset and the lowest tube currents, the imaging doses were 0.16 and 0.08 mGy, 5.29 and 2.80 mGy, and 18.23 and 2.69 mGy for head, thorax, and pelvis, respectively. Noise was the only quality aspect directly dependent on tube current, being increased by 1.5 times with a tube current half that of the preset in head and thorax, and by 2.2 times with a tube current 1/8 of the preset in the pelvis. Accurate auto-bone matching was performed within 1 mm at the lowest tube current. The auto-soft tissue matching could not be performed with the lowest tube current; however, manual-soft tissue matching could still be performed within 2 mm or less.

Conclusion: Noise was the only image quality aspect dependent on the imaging dose. Auto-bone and manual-soft tissue matching could still be performed at the lowest imaging dose.

Advances In Knowledge: When optimizing kV-CBCT imaging dose, the impact on bone and soft tissue image registration accuracy should be evaluated.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592495PMC
http://dx.doi.org/10.1259/bjro.20190028DOI Listing

Publication Analysis

Top Keywords

tube current
20
matching performed
16
registration accuracy
12
image registration
12
head thorax
12
lowest tube
12
tissue matching
12
kilovoltage cone-beam
8
imaging doses
8
thorax pelvis
8

Similar Publications

Complete detachment of the aortic root following a Bentall procedure is an exceptionally rare complication. The vast majority of reported cases are secondary to prosthetic valve endocarditis or underlying vasculitis. Currently, the most reliable treatment for aortic root dehiscence-particularly in the context of prosthetic valve endocarditis-is repeat root replacement, typically via a second Bentall procedure or with the use of a homograft or allograft.

View Article and Find Full Text PDF

SiO NP promotes allergic gastritis induced by degranulation of mouse MC9 cell through AQP4-mediated impairment of SIRT3-TFAM deacetylation and mitochondrial autophagy.

J Hazard Mater

September 2025

College of Veterinary Medicine, Northeast Agricultural University, Harbin 150030, PR China; Key Laboratory of the Provincial Education Department of Heilongjiang for Common Animal Disease Prevention and Treatment, College of Veterinary Medicine, Northeast Agricultural University, Harbin 150030, PR C

Silicon dioxide nanoparticles (SiO NPs) are a novel material with a wide range of applications whose cumulative effects in the body pose certain health risks. The types of gastric injuries caused by different-sized SiO NPs and their mechanisms, however, remain unclear. Based on this, we established a mouse subchronic exposure model (10 mg/kg/d, 21 consecutive days of tube-feeding) with different SiO NP sizes (50, 300, and 1000 nm) in conjunction with in vitro MC9 and BMMCs models (160 μg/mL exposure for 24 h) to explore the gastric injury mechanisms.

View Article and Find Full Text PDF

Background: Current neurovascular unit isolation requires processing brain microvascular endothelial cells (BMECs) and neurons from separate animals, preventing concurrent analysis of neurovascular crosstalk within identical genetic/physiological contexts.

New Methods: We developed an enzymatic digestion/bovine serum albumin density gradient technique that enables the simultaneous isolation of neural tissue and microvascular segments from individual mice. The neural tissue was filtered and centrifuged for primary cortical neuron culture on poly-L-lysine-coated plates.

View Article and Find Full Text PDF

Competency framework for patients with surgical jejunostomy: from theory to educational practice.

Ann Pharm Fr

September 2025

Nantes Université, CHU Nantes, Pharmacie, F-44000 Nantes, France; Nantes Université, CHU Nantes, INSERM, Regenerative Medicine and Skeleton, RMeS, UMR 1229, F-44000 Nantes, France.

Objectives: Patients with a jejunostomy use their tube daily in most cases, which exposes them to various complications. However, to our knowledge, the current literature is limited regarding available educational interventions or tools for these patients, and no specific competency framework for surgical jejunostomy has been developed yet. This project aimed to develop a unique and adapted competency framework for surgical jejunostomy patients through a multiprofessional and consensus-driven approach and to create a fun educational tool based on this framework to promote an effective educational approach.

View Article and Find Full Text PDF

Introduction: Endotracheal tube (ETT) cuff pressures that exceed 20-30 cmHO may lead to iatrogenic adverse effects such as cough, sore throat, and tracheal edema or more serious complications including tracheal stenosis, recurrent laryngeal nerve injury, and tracheal rupture. The current study evaluates a novel technique, titration of the ratio of expiratory to inspiratory tidal volumes (TV), to regulate intracuff pressure.

Methods: This prospective, cross-over trial measured intracuff pressure in a cohort of pediatric patients presenting for general anesthesia with an ETT.

View Article and Find Full Text PDF