Materials (Basel)
July 2023
In situ fenestration of endovascular stent-grafts has become a mainstream bailout technique to treat complex emergent aneurysms while maintaining native anatomical visceral and aortic arch blood supplies. Fabric tearing from creating the in situ fenestration using balloon angioplasty may extend beyond the intended diameter over time. Further tearing may result from the physiologic pulsatile motion at the branching site.
View Article and Find Full Text PDFJ Endovasc Ther
April 2021
Purpose: To evaluate the response of various stent-grafts after laser fenestration and dilation with noncompliant balloons to determine the optimal therapeutic combination for this treatment technique.
Materials And Methods: Five aortic stent-grafts were evaluated ex vivo: the Bolton RelayPlus, Jotec E-vita Thoracic 3G, Medtronic Valiant, Cook Zenith Alpha, and Vascutek Anaconda. Small holes were created using an excimer laser with the grafts submerged in saline.
The damage caused to the fabric of endovascular stent-grafts most often occurs at the contact zones between the fabrics where they are attached to the apices of Z-shaped stents as a result of normal physiologic pulsatile movement within angulated vessels in vivo. Although design improvements were made over the years, the risks were not fully eliminated even with the newer M-shaped stent reconfiguration. In this study, we proposed to create and manufacture a novel fabric for stent-grafts with specifically designed reinforced zones to enhance resistance to fabric abrasion.
View Article and Find Full Text PDFPurpose: To examine the effects of in situ laser fenestration and subsequent balloon dilation (noncompliant vs cutting) on the graft fabric of 4 aortic stent-graft models.
Method: In an in vitro setup, the Zenith TX2, Talent, Endurant, and Anaconda aortic stent-grafts (all made of polyester graft material) were subjected to laser fenestration with a 2.3-mm-diameter probe at low and high energy in a physiologic saline solution followed by balloon dilation of the hole.
In situ fenestration of stent-grafts allows patients with life threatening aortic pathologies to be amenable to emergent "off the shelf indications for use" percutaneous treatments as a bail out technique. Three types of aortic stent-grafts were subjected to laser fenestration in a physiological saline solution followed by balloon angioplasty using 8, 10 or 12 mm in diameter noncompliant balloons. The morphology and the size of fenestrations were observed under optical and scanning electron microscopy.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
March 2018
The Bentall procedure introduced in 1968 represents an undisputed cure to treat multiple pathologies involving the aortic valve and the ascending thoracic aorta. Over the years, multiple modifications have been introduced as well as a standardized approach to the operation with the goal to prevent long-term adverse events. The Gelweave Valsalva graft provides a novel manner to more efficiently reconstruct the anatomy of the aortic root either with a valve-sparing procedure or with the implantation of a valved conduit (bioprosthesis or mechanical valve).
View Article and Find Full Text PDFMaterials (Basel)
February 2016
Whereas buckling can cause type III endoleaks, long-term twisting of a stent-graft was investigated here as a mechanism leading to type V endoleak or endotension. Two experimental device designs supported with Z-stents having strut angles of 35° or 45° were compared to a ringed control under accelerated twisting. Damage to each device was assessed and compared after different durations of twisting, with focus on damage that may allow leakage.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
April 2014
In this study, we aimed to investigate changes to the fabric of Talent stent-grafts following implantation of aortic endografts and to determine the possible causes of fatigue and/or failure of the grafts. Six devices were explanted at reoperation (N=5) and autopsy (N=1). Selected segments were assessed nondestructively by gross observation and destructively by analyzing textile characteristics and chemical properties.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
April 2014
The biodurability of the Nitinol wires used in stent-grafts retrieved from reoperations and autopsy was analyzed to assess the possible causes of fracture and/or corrosion of the stents. The Nitinol wires of six explanted devices presented a corrosion-free surface after in vivo service. The drawing lines in the control wires were still present, but neither burrs nor flakes were observed.
View Article and Find Full Text PDFJ Biomater Appl
March 2014
Stent-grafts externally fitted with a Z-shaped stents were compared to devices fitted with ringed stents in an in vitro oscillating fatigue machine at 200 cycles per minute and a pressure of 360 mmHg for scheduled durations of up to 1 week. The devices fitted with Z-stents showed a considerably lower endurance limit to buckling compared to the controls. The contact between the apexes of adjacent Z-stents resulted in significant damage to the textile scaffolds and polyester fibers due to the sharp angle of the Z-stents.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
April 2014
J Long Term Eff Med Implants
March 2012
Intra-operative fenestrations of stent grafts make more frail and elderly patients amenable to endovascular surgery but require further assessment of the viability of currently used experimental techniques. Four types of polyester fabrics currently employed in stent grafts were exposed in vitro to various protocols of fenestration: cutting, trocaring, and cantering. The resulting fenestrations were examined by gross observation, light microscopy, and scanning electron microscopy.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
August 2012
Six Talent stent-grafts were harvested at reoperations (N=5) and autopsy (N=1). The explants were observed nondestructively, including gross morphology, X-rays, CT scans and closed pressure system analysis. The Nitinol frames in three devices harvested at reoperations and another harvested at autopsy were intact.
View Article and Find Full Text PDFThanks to minimally invasive procedures, frail and elderly patients can also benefit from innovative technologies. More than 14 million implanted pacemakers deliver impulses to the heart muscle to regulate the heart rate (treating bradycardias and blocks). The first human implantation of defibrillators was performed in early 2000.
View Article and Find Full Text PDFPurpose: To report the use of a branched, balloon-deployable stent-graft to treat abdominal aortic aneurysm (AAA) in the setting of a solitary kidney.
Case Report: A 72-year-old man with a solitary intrapelvic kidney and multiple comorbid conditions was diagnosed with an asymptomatic 5.3-cm abdominal aortic aneurysm (AAA); the renal artery emerged from the aneurysm sac.
J Biomed Mater Res B Appl Biomater
May 2010
Fluoropassivation and gelatin coating have been applied to polyethylene terephthalate (PET) vascular prosthesis to combine the advantages of both polytetrafluoroethylene (PTFE) and PET materials, and to eliminate the preclotting procedure. The morphological, chemical, physical, and mechanical properties of such prostheses were investigated and compared with its original model. Fluoropassivation introduced -OCF(3), -CF(3), and -CFCF(2)- structures onto the surface of the polyester fibers.
View Article and Find Full Text PDFPurpose: To determine the evolution of the proximal aortic neck diameter in mid- to long-term follow-up after endovascular aneurysm repair of abdominal aortic aneurysm (AAA) with a balloon-expandable stent-graft.
Methods: Thirty patients (27 men; average age 71 years, range 56-87) with infrarenal AAAs were treated with the SETA-Latecba balloon-expandable stent-graft (6 aortomonoiliac and 24 bifurcated configurations). Follow-up ranged from 4 to 8 years (mean 73.
Numerous endovascular stent grafts to treat intrarenal aortic aneurysms are now commercially available, and many new concepts are currently in development worldwide. In order to objectively quantify their outcomes, we propose a detailed protocol to examine a reference device that was harvested from a patient who died a few hours after endovascular stent-graft deployment for an abdominal aortic aneurysm according to the 3Bs rule (biocompatibility, biofunctionality, and biodurability). Relevant patient history of this 63-year-old man included radiotherapy treatment for lung cancer.
View Article and Find Full Text PDFJ Long Term Eff Med Implants
February 2010
A patient was fitted with an Anaconda stent graft for which there was a persistent type II endoleak. Two subsequent attempts at embolization were unable to resolve the endoleak. The diameter of the aneurysm varied initially from 5.
View Article and Find Full Text PDFFurther to the rapid enlargement of an aneurysm to 5.6 cm in diameter after 3 years of surveillance, a 79-year-old patient was fitted with a Vanguard modular stent graft and monitored on a regular basis for 6 years. Two years later, the aneurysmal sac ruptured.
View Article and Find Full Text PDFBackground: The study was conducted to demonstrate improved survival (30-day mortality) after the introduction of an emergency endovascular therapy protocol for ruptured abdominal aortic aneurysms (rAAA). Numerous authors have successfully demonstrated reduced mortality in patients with rAAA using endovascular techniques. Comparison of endovascular aneurysm repair (EVAR) with open repair for rAAA may be misleading, however, because EVAR cannot be performed on all patients, and selection bias may explain the superior performance of any given surgical or endovascular strategy.
View Article and Find Full Text PDFBackground: Numerous repairs exist for direct inguinal hernias. These repairs are limited by the shortcomings of their respective technique. Reported recurrence rates for all currently employed hernia repairs for direct inguinal hernias range from 1% to 10%.
View Article and Find Full Text PDF