Article Abstract

Introduction: Left bundle branch pacing (LBBP) in bipolar pacing with anodal capture produces a more balanced ventricular activation than conventional unipolar pacing but need high pacing output. The present study aimed to compare the electrophysiological characteristics of LBBP in different bilateral electrode pacing vector configurations.

Methods: A total of 57 patients who met the criteria for left bundle branch (LBB) capture and underwent three bilateral electrode pacing vector configuration test were enrolled. The electrocardiogram (ECG) and electrogram (EGM) parameters were evaluated and other electrophysiological characteristics were analyzed using a three-electrode configuration test.

Results: Seven capture modes [right ventricular septal (RVS) + left ventricular septal (LVS) + LBB, RVS + LBB, LVS + LBB, RVS + LVS, RVS, LVS, and LBB] were utilized in the study. The thresholds of full fusion mode (RVS + LVS + LB) in Bilateral Cathodes and Ring Bipolar were all lower than that in Tip Bipolar (1.2 ± 0.5 V vs. 2.7 ± 1.0 V,  < 0.001; 1.6 ± 0.6 V vs. 2.7 ± 1.0 V,  < 0.001). Full fusion mode had the shortest P-QRS (116.9 ± 12.8 ms) and V1 RWPT (94.5 ± 12.3 ms), V6 RWPT remain the shortest (64.9 ± 9.7 ms).

Conclusion: Changing the bilateral electrode pacing vector configuration to Bilateral Cathodes and Ring Bipolar can reduce the full fusion mode capture threshold compared to conventional bipolar pacing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11936911PMC
http://dx.doi.org/10.3389/fcvm.2025.1500196DOI Listing

Publication Analysis

Top Keywords

left bundle
12
bundle branch
12
bilateral electrode
12
electrode pacing
12
pacing vector
12
pacing
8
branch pacing
8
electrophysiological characteristics
8
ventricular septal
8
comparison electrophysiological
4

Similar Publications

Background: During left bundle branch area pacing (LBBAP), several markers of electrical synchrony, (V6 R-wave peak time (RWPT), aVL-RWPT, and the V6-V1 interpeak interval), are commonly used to assess left bundle branch (LBB) capture. Nevertheless, the relationship between these electrocardiographic (ECG) measurements and mechanical synchrony remains poorly understood.

Objective: We aimed to analyze the association between electrical parameters from the paced QRS complex and mechanical performance assessed through 2D strain and myocardial work (MW) indices, following LBBAP implantation.

View Article and Find Full Text PDF

Author's reply to Do all pacing system analyzer models fit one size for the conduction system pacing?

Heart Rhythm

September 2025

Electrophysiology Unit. Cardiovascular Division. Hospital del Mar. Passeig Maritim de la Barceloneta. Barcelona. Spain; Department of Medicine, Universitat Autònoma de Barcelona, Bellaterra, Spain.

View Article and Find Full Text PDF

Objective: To investigate two conditions that have been poorly investigated in the medical literature before in the context of atrial fibrillation: the coexistence and association of right or left bundle branch block and axis deviation in patients with permanent atrial fibrillation compared to the control group of healthy subjects with sinus rhythm.

Material And Methods: We conducted an analytic, retrospective observational study performed at Ziv Medical Center, Safed, Israel, collecting data from medical history records of all patients that have been diagnosed with permanent atrial fibrillation versus healthy controlled patients with normal sinus rhythm. We analyzed their ECGs in order to assess the presence of any bundle branch block and/or axis deviation.

View Article and Find Full Text PDF