Article Abstract

Background: Intraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality. IRRAflow® (IRRAS AB) is a new technology introduced to accelerate IVH clearance by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline, while simultaneously monitoring and maintaining a stable intracranial pressure (ICP). We addressed important aspects of the device implementation and intracranial lavage.

Method: To allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup. We evaluated 1) compatibility between the IRRAflow® catheter and the Silverline f10 bolt (Spiegelberg), 2) the physiological and hydrodynamic effects of varying the IRRAflow® settings, 3) the accuracy of the IRRAflow® injection volumes, and 4) the reliability of the internal ICP monitor of the IRRAflow®.

Results: The IRRAflow® catheter was not compatible with Silverline bolt fixation, which was associated with leakage and obstruction. Design space exploration of IRRAflow® settings revealed that appropriate settings included irrigation rate 20 ml/h with a drainage bag height at 0 cm, irrigation rate 90 ml/h with a drainage bag height at 19 cm and irrigation rate 180 ml/h with a drainage bag height at 29 cm. We found the injection volume performed by the IRRAflow® to be stable and reliable, while the internal ICP monitor was compromised in several ways. We observed a significant mean drift difference of 3.16 mmHg (variance 0.4, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (variance 0.28) in the pressures measured by the IRRAflow® compared to 0.5 mmHg (variance 1.12) in the Raumedic measured pressures.

Conclusion: Bolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11020765PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0297131PLOS

Publication Analysis

Top Keywords

drainage bag
16
irraflow®
12
irraflow® catheter
12
icp monitor
12
irrigation rate
12
ml/h drainage
12
bag height
12
mmhg variance
12
irraflow® system
8
irraflow® settings
8

Similar Publications

Background: Passive subdural drainage is used to reduce the risk of chronic subdural hematoma (cSDH) recurrence and mortality. However, the effects of patient positioning on intracranial pressure (ICP) during passive drainage remain unclear.

Objective: To examine how subdural drainage and patient positions influence postoperative ICP in cSDH patients with a subdural passive drainage system.

View Article and Find Full Text PDF

Equipoise exists regarding the optimal method to drain pleural fluid during thoracentesis. While several institutions use wall-based automated suction, others point to the risk of excessively high suction pressures and therefore elevated barotrauma risk as a reason to avoid it. We first performed in vitro experiments involving drainage of a 1-liter saline bag using standard thoracentesis apparatus, a digital manometer, and either manual drainage (using a 60 mL syringe) or automated drainage (using wall suction at the maximum setting).

View Article and Find Full Text PDF

Objective: To compare the effects of normal-pressure drainage (norPD) and negative-pressure drainage (negPD) after laparoscopic radical prostatectomy (LRP) in the treatment of PCa.

Methods: We retrospectively analyzed the clinical and follow-up data on 87 cases of PCa treated by LRP from October 2019 to October 2022, 46 receiving norPD and the other 41 negPD postoperatively. We indwelt an F20 pelvic drainage tube for each of the patients, which were connected to an anti-reflux drainage bag for those in the norPD group and a 200 ml negative pressure ball for those in the negPD group, and removed the tubes at the drainage volume ≤30 ml/d.

View Article and Find Full Text PDF

[Interrupting the spread of germs : The art of emptying urinary drainage bags].

Urologie

July 2025

, Polizeimeister-Kaspar-Straße 44, 60320, Frankfurt/Main, Deutschland.

Background: Due to the development of antibiotic resistance in many bacterial species, the demands in nursing care have changed significantly. In handling urine catheters in home care, ESKAPE (Enterococcus faecium, Staphylococcus aureus, Klebsiella pneumoniae, Acinetobacter baumannii, Pseudomonas aeruginosa, and Enterobacter spp.) pathogens may present as urinary tract infections or in bacteriuria.

View Article and Find Full Text PDF

Background: Late subluxation of an intraocular lens (IOL) within the capsular bag is a rare complication of cataract surgery, primarily associated with progressive zonular weakening in conditions such as pseudoexfoliation syndrome, high myopia, prior vitreoretinal surgery, connective tissue disorders, or ocular trauma. While IOL-capsular bag subluxation is well-documented, its impact on glaucoma drainage devices remains poorly understood. We present a rare case of Valsalva-induced IOL subluxation obstructing an Ahmed glaucoma valve, resulting in acute intraocular pressure (IOP) elevation in a patient with pseudoexfoliative glaucoma.

View Article and Find Full Text PDF