Article Abstract

Purpose: Aim of this study was to evaluate the use of computer-aided design (CAD) models for attenuation correction (AC) of hardware components in positron emission tomography/magnetic resonance (PET/MR) imaging.

Methods: The technical feasibility and quantitative impact of CAD-AC compared to computer tomography (CT)-based AC (reference) was investigated on a modular phantom consisting of 19 different material samples (plastics and metals arranged around a cylindrical emission phantom) typically used in phantoms, patient tables, and radiofrequency (RF) coils in PET/MR. The clinical applicability of the CAD-AC method was then evaluated on a 16-channel RF breast coil in a PET/MR patient study. The RF breast coil in this study was specifically designed PET compatible. Using this RF breast coil, the impact on clinical PET/MR breast imaging was systematically evaluated in breast phantom measurements and, furthermore, in n = 10 PET/MR patients with breast cancer. PET data were reconstructed three times: (1) no AC (NAC), (2) established CT-AC, and (3) CAD-AC. For both phantom measurements, a scan without attenuating hardware components (material probes or RF breast coil) was acquired serving as reference. Relative differences in PET data were calculated for all experiments.

Results: In all phantom and patient measurements, significant gains in PET signal compared to NAC data were measurable with CT and CAD-AC. In initial phantom experiments, mean relative differences of -0.2% for CT-AC and 0.2% for CAD-AC were calculated compared to reference measurements without the material probes. The application to a RF breast coil depicts that CAD-AC results in significant gains compared to NAC data (10%) and a slight underestimation in PET signal of -1.3% in comparison to the no-coil reference measurement. In the patient study, a total of 15 congruent lesions in all 10 patients with a mean relative difference of 14% (CT and CAD-AC) in standardized uptake value compared to NAC data could be detected.

Conclusions: To ensure best possible PET image quality and accurate PET quantification in PET/MR imaging, the AC of hardware components such as phantoms and RF coils is important. In initial phantom experiments and in clinical application to an RF breast coil, it was found that CAD-based AC results in significant gains in PET signal compared to NAC data and provides comparably good results to the established method of CT-based AC.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15284DOI Listing

Publication Analysis

Top Keywords

breast coil
28
compared nac
16
nac data
16
hardware components
12
pet signal
12
breast
10
attenuation correction
8
clinical application
8
16-channel breast
8
patient study
8

Similar Publications

Purpose: Supine breast MRI has the potential to improve over standard prone breast magnetic resonance imaging (MRI) in terms of efficiency and image quality, image alignment with diagnostic and treatment procedures, and overall accessibility. This study aims to characterize potential technical challenges of imaging in the supine position: (i)  field inhomogeneities, (ii)  variations, (iii) respiratory-induced breast motion, and (iv) supine breast geometry.

Methods: Ten healthy subjects were scanned at 3T in both prone and supine positions to quantify and compare (i) and (ii) between both positions, and to assess (iii) in the supine position.

View Article and Find Full Text PDF

Recent evidence has highlighted immune checkpoint inhibitors as among the most promising immunotherapies for various malignancies. However, a significant proportion of HCC patients exhibit poor responses. Lipid metabolic heterogeneity is considered a key driver of cancer progression.

View Article and Find Full Text PDF

Wireless radiofrequency coils offer a valuable low cost solution for various MR applications due to several benefits, such as cable-free connection and compatibility with MR platforms of different vendors. Namely, for the purpose of clinical high-field human breast imaging several wireless transceiver coils are known to the date, those operational principle is based on inductive coupling with a body coil. These coils are commonly consist of a several volume resonators to perform bilateral breast imaging.

View Article and Find Full Text PDF

Design and synthesis of ROCK2-selective inhibitors bearing 6-indazolamino-2-pyrrolidinylpyridine for the treatment of breast cancer metastasis.

Bioorg Chem

August 2025

Key Laboratory of Structure-Based Drug Design and Discovery, Ministry of Education, Shenyang Pharmaceutical University, Shenyang 110016, China. Electronic address:

Rho-associated coiled-coil kinase (ROCK) serves as a critical driver of cellular metastasis, demonstrating therapeutic potential against breast cancer cell metastasis. To further enhance the ROCK2 inhibitory activity of belumosudil (ROCK2 IC = 0.153 μM), a novel series of 6-indazolamino-2-pyrrolidylpyridine derivatives were designed and synthesized.

View Article and Find Full Text PDF

Rational Design of 6-Amino-2-Piperazinylpyridine-Based ROCK2 Inhibitors With Anti-Breast Cancer Metastatic Efficiency.

Arch Pharm (Weinheim)

August 2025

Key Laboratory of Structure-Based Drug Design and Discovery, Ministry of Education, School of Pharmaceutical Engineering, Shenyang Pharmaceutical University, Shenyang, China.

Rho-associated coiled coil kinase (ROCK) plays a pivotal role in regulating actin cytoskeleton remodeling and cellular motility, establishing it as a promising therapeutic target for metastatic breast cancer. Guided by the lead compound belumosudil, a novel series of ROCK2 inhibitors based on a 6-amino-2-piperazine pyridine scaffold was designed and synthesized. The Kinase-Glo assay identified compound 14r as the most potent analog, showing an IC value of 82.

View Article and Find Full Text PDF