Article Abstract

Background: Cardiovascular magnetic resonance (CMR) is the current gold standard for the assessment of left ventricular (LV) function. Repeated breath-holds are needed for standard multi-slice 2D cine steady-state free precession sequences (M2D-SSFP). Accelerated single breath-hold techniques suffer from low contrast between blood pool and myocardium. In this study an intravascular contrast agent was prospectively compared to an extravascular contrast agent for the assessment of LV function using a single-breath-hold 3D-whole-heart cine SSFP sequence (3D-SSFP).

Methods: LV function was assessed in fourteen patients on a 1.5 T MR-scanner (Philips Healthcare) using 32-channel coil technology. Patients were investigated twice using a 3D-SSFP sequence (acquisition time 18-25 s) after Gadopentetate dimeglumine (GdD, day 1) and Gadofosveset trisodium (GdT, day 2) administration. Image acquisition was accelerated using sensitivity encoding in both phase encoding directions (4xSENSE). CNR and BMC were both measured between blood and myocardium. The CNR incorporated noise measurements, while the BMC represented the coeffiancy between the signal from blood and myocardium [1]. Contrast to noise ratio (CNR), blood to myocardium contrast (BMC), image quality, LV functional parameters and intra-/interobserver variability were compared. A M2D-SSFP sequence was used as a reference standard on both days.

Results: All 3D-SSFP sequences were successfully acquired within one breath-hold after GdD and GdT administration. CNR and BMC were significantly (p < 0.05) higher using GdT compared to GdD, resulting in an improved endocardial definition. Using 3D-SSFP with GdT, Bland-Altman plots showed a smaller bias (95% confidence interval LVEF: 9.0 vs. 23.7) and regression analysis showed a stronger correlation to the reference standard (R2 = 0.92 vs. R2 = 0.71), compared to 3D-SSFP with GdD.

Conclusions: A single-breath-hold 3D-whole-heart cine SSFP sequence in combination with 32-channel technology and an intravascular contrast agent allows for the accurate and fast assessment of LV function.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3438073PMC
http://dx.doi.org/10.1186/1532-429X-14-53DOI Listing

Publication Analysis

Top Keywords

single breath-hold
12
contrast agent
12
blood myocardium
12
accelerated single
8
extravascular contrast
8
cnr bmc
8
contrast
6
breath-hold assessment
4
assessment cardiac
4
function
4

Similar Publications

We report the management of a 64-year-old male with newly diagnosed bulbar-onset myasthenia gravis (MG) who was hospitalized with acute neuromuscular respiratory insufficiency. This case highlights the challenges in monitoring respiratory function in MG patients, especially in the presence of bulbar and nuchal weakness, and emphasizes the potential utility of single breath-hold time (SBHT) over forced vital capacity (FVC) as a reliable bedside monitoring tool. Despite initial stabilization with intravenous immunoglobulin (IVIG), the patient deteriorated, requiring escalation to the intensive care unit (ICU), and the clinical worsening corresponded with the SBHT rather than with FVC.

View Article and Find Full Text PDF

Purpose: Presenting a technique to quantify the transverse relaxation time T, which is associated with the diffusion of water molecules through the internal magnetic field gradients of the lung in-vivo.

Methods: A Half-Fourier-Acquired Single-shot Turbo spin-Echo (HASTE) pulse sequence with Hahn-echo preparation was implemented and used for image acquisition. Quantification of T was performed by acquiring multiple images with identical TE, but with a different number of refocusing pulses between excitation and signal acquisition.

View Article and Find Full Text PDF

Purpose: To develop a multiparametric free-breathing three-dimensional, whole-liver quantitative maps of water T, water T, fat fraction (FF) and R*.

Methods: A multi-echo 3D stack-of-spiral gradient-echo sequence with inversion recovery and T-prep magnetization preparations was implemented for multiparametric MRI. Fingerprinting and a neural network based on implicit neural representation (FINR) were developed to simultaneously reconstruct the motion deformation fields, the static images, perform water-fat separation, and generate T, T, R*, and FF maps.

View Article and Find Full Text PDF

Effect of warming gadoxetic acid on arterial-phase artifacts in dynamic liver MRI: A prospective single-center study.

Eur J Radiol

August 2025

Departments of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1, Maidashi, Higashi-ku, Fukuoka 812-8582, Japan.

Objectives: To investigate whether warming gadoxetic acid affects the frequency and degree of artifacts in the arterial phase of MRI.

Methods: This prospective study included patients who underwent gadoxetic acid-enhanced MRI (April 2016-November 2017 and June-November 2018) at a single center. Either warmed (37 °C, n = 134) or non-warmed (24 °C, n = 137) gadoxetic acid was intravenously injected at a dose of 0.

View Article and Find Full Text PDF

A Systematic Approach to the Evaluation of Lumbosacral Plexus MRI: Indications, Protocol Anatomy, and Pathology.

AJNR Am J Neuroradiol

August 2025

From Tu Lab for Diagnostic Research (SR, OZ, CKH, LHT) and Department of Radiology and Biomedical Imaging (SR, CH, IDDS, JKN, OZ, CKH, CFEK, LHT), Yale School of Medicine, New Haven, CT, USA.

Magnetic resonance imaging (MRI) of the lumbosacral plexus (LSP) is a critical tool for evaluating pelvic and lower extremity peripheral nerve disorders. The rarity of this exam type and the complexity of underlying anatomy can produce challenges for interpretation. This pictorial review outlines a structured approach for evaluating lumbosacral plexus MRI (LSP MRI), emphasizing key imaging anatomy, MRI protocols, and a spectrum of commonly encountered pathologies to aid radiologists in interpretating LSP MRI.

View Article and Find Full Text PDF