Article Abstract

Purpose: To determine whether frameless thermoplastic mask-based immobilization is adequate for image-guided cranial radiosurgery.

Methods And Materials: Cone-beam CT localization data from patients with intracranial tumors were studied using daily pre- and posttreatment scans. The systems studied were (1) Type-S IMRT (head only) mask (Civco) with head cushion; (2) Uni-Frame mask (Civco) with head cushion, coupled with a BlueBag body immobilizer (Medical Intelligence); (3) Type-S head and shoulder mask with head and shoulder cushion (Civco); (4) same as previous, coupled with a mouthpiece. The comparative metrics were translational shift magnitude and average rotation angle; systematic inter-, random inter-, and random intrafraction positioning error was computed. For strategies 1-4, respectively, the analysis for interfraction variability included data from 20, 9, 81, and 11 patients, whereas that for intrafraction variability included a subset of 7, 9, 16, and 8 patients. The results were compared for statistical significance using an analysis of variance test.

Results: Immobilization system 4 provided the best overall accuracy and stability. The mean interfraction translational shifts (± SD) were 2.3 (± 1.4), 2.2 (± 1.1), 2.7 (± 1.5), and 2.1 (± 1.0) mm whereas intrafraction motion was 1.1 (± 1.2), 1.1 (± 1.1), 0.7 (± 0.9), and 0.7 (± 0.8) mm for devices 1-4, respectively. No significant correlation between intrafraction motion and treatment time was evident, although intrafraction motion was not purely random.

Conclusions: We find that all frameless thermoplastic mask systems studied are viable solutions for image-guided intracranial radiosurgery. With daily pretreatment corrections, symmetric PTV margins of 1 mm would likely be adequate if ideal radiation planning and targeting systems were available.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2010.06.022DOI Listing

Publication Analysis

Top Keywords

frameless thermoplastic
12
intrafraction motion
12
thermoplastic mask-based
8
mask-based immobilization
8
data patients
8
systems studied
8
mask civco
8
civco head
8
head cushion
8
head shoulder
8

Similar Publications

Objective: The Leksell Gamma Knife (LGK) Icon™ facilitates frameless (mask-based) fixation through its high-definition motion management (HDMM) system. However, the HDMM only records the intra-fractional motion values of patient marker without specifying changes along the X, Y, and Z axes. This study investigates the coordinate shifts in patient markers using the HDMM system during frameless gamma knife radiosurgery (GKRS) in patients with meningioma and metastases.

View Article and Find Full Text PDF

Uveal melanoma (UM) is the most common primary intraocular malignancy, traditionally managed with episcleral plaque brachytherapy or enucleation. Single-fraction stereotactic radiosurgery (SRS) using a linear accelerator (LINAC) is an emerging alternative offering high precision and eye preservation. However, existing SRS/stereotactic radiation therapy (SRT) techniques often rely on mechanical immobilization or patient-maintained fixation.

View Article and Find Full Text PDF

While the treatment of unruptured arteriovenous malformation of the brain (bAVM) is controversial, active intervention is usually recommended for patients who have had a previous intracerebral hemorrhage. Stereotactic radiosurgery (SRS) is the preferred treatment for bAVM when the patient is unsuitable for surgery or embolization. Currently, linear accelerator-based cranial SRS is gaining popularity, and this case report describes the multidisciplinary approach in treating a military sniper with a 3.

View Article and Find Full Text PDF

Hypofractionated stereotactic radiotherapy using coplanar VMAT for single small brain metastasis: dosimetric analysis and clinical outcomes.

Front Oncol

February 2025

Thoracic Oncology Ward, Cancer Center and State Key Laboratory of Biotherapy, West China Hospital, Sichuan University, Chengdu, China.

Purpose: To evaluate the dosimetric parameters and clinical outcomes of hypofractionated stereotactic radiotherapy (HSRT) for small brain metastases [BMs; planning target volume (PTV) ≤ 4 cm) via coplanar volumetric modulated arc therapy (C-VMAT).

Methods: Between March 2019 and February 2023, 68 patients with a single BM treated with Linac-based HSRT (24-39 Gy in three fractions) via C-VMAT and a 3-mm PTV margin were enrolled in this retrospective analysis. A frameless head-neck-shoulder thermoplastic mask, whose immobilization accuracy is inferior to that of specialized mask fixation systems, was used to immobilize patients.

View Article and Find Full Text PDF

Purpose: To commission and assess the clinical performance of a new commercial surface imaging (SI) system by analyzing intra-fraction motion from the initial cohort of patients treated with frameless stereotactic radiosurgery (fSRS).

Methods: The IDENTIFY SI system was commissioned for clinical use on an Edge (Varian Medical Systems, Palo Alto, CA) linear accelerator. All patients who received intracranial radiotherapy with HyperArc (Varian Medical Systems, Palo Alto, CA) were immobilized with the Encompass (Qfix, Avondale, PA) thermoplastic mask and monitored for intra-fraction motion with SI.

View Article and Find Full Text PDF