Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To compare corneal biomechanical parameters and two measures of intraocular pressure (IOP) in eyes before and after excimer laser refractive surgery, with the Ocular Response Analyser (ORA).

Materials And Methods: Eighty normal eyes of 41 patients undergoing excimer laser refractive surgery in Birmingham, U.K. were recruited into three groups: Laser Assisted-Epithelial Keratomileusis (LASEK) (Myopes), Laser Assisted in Situ Keratomileusis (LASIK) (myopes) and LASIK (hyperopes). The preop and 3 months postop Goldmann correlated IOP (IOPg), corneal compensated IOP (IOPcc), corneal hysteresis (CH), and corneal resistance factor (CRF) were measured by the ORA. Central corneal thickness (CCT) was measured using ultrasonic pachymeter. The differences of the changes in IOPg, IOPcc, CH, CRF and CCT between the three groups were estimated. A General Linear Model was selected to investigate the influence of gender, age, initial conditions (CH, CRF, CCT, IOPcc and IOPg) and changes in CCT on the measured IOP.

Results: The differences between the mean IOPg, CH and CRF after refractive surgery were statistically significant for all three groups. The hyperopic LASIK group had a significantly smaller change compared to the other groups (which had no statistical significance). The preop IOPg, preop CH and gender were significant predictors of the changes in measured pressure and biomechanical parameters after surgery in the myopic groups only.

Conclusion: CH and CRF were found to decrease after both myopic and hyperopic refractive surgery. CH and CRF measurement may prove important tools to clarify the role of corneal biomechanics for refractive surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clae.2009.02.005DOI Listing

Publication Analysis

Top Keywords

refractive surgery
24
excimer laser
12
laser refractive
12
three groups
12
ocular response
8
response analyser
8
corneal hysteresis
8
eyes excimer
8
biomechanical parameters
8
cct measured
8

Similar Publications

Purpose: To report a case of acute and transient accommodative insufficiency after laser in situ keratomileusis (LASIK) due to coronavirus disease 2019 (COVID-19).

Methods: Case report and literature review.

Results: A 36-year-old man complained of acute blurred near vision 7 days after uneventful bilateral hyperopic LASIK, concurrent with the onset of COVID-19 infection.

View Article and Find Full Text PDF

Purpose: To analyze the effects of ablation interruption on ablation depths and clinical refractive outcomes to characterize the impact of ambient temperature changes and ablation interruption on ocular surface temperature (OST) during excimer laser ablation.

Methods: This prospective study was conducted on laser ablations in polymethylmethacrylate (PMMA) plates and porcine corneas to simulate laser in situ keratomileusis (LASIK) treatments using the EX500 laser (Alcon Laboratories, Inc) at ambient temperatures of 18, 20, and 22 °C. Ablation interruption was performed for 1, 2, 3, 4, and 5 seconds at the 10th second of the treatment of -9.

View Article and Find Full Text PDF

Purpose: To evaluate axis-dependent visual and refractive outcomes of small incision lenticule extraction (SMILE) in patients with interocular astigmatic axis discordance.

Methods: Seventy-five patients (150 eyes) with interocular astigmatic axis discordance were included in the study. Based on interocular axis combinations, patients were stratified into three cohorts: with-the-rule (WTR)/against-the-rule (ATR) (n = 19), WTR/oblique astigmatism (OA) (n = 39), and ATR/OA (n = 17).

View Article and Find Full Text PDF

Purpose: To discuss the technique and outcome of what the authors called the "flap-in-flap" technique and report its safety as a procedure for correction of post-laser in situ keratomileusis (LASIK) myopic regression.

Methods: Seven eyes of 4 patients were included in this study. All patients had previously undergone LASIK for compound myopic astigmatism using the Moria M2 micro-keratome (Moria) 8 to 12 years prior to presentation.

View Article and Find Full Text PDF

Purpose: To report the refractive outcome of femtosecond laser-assisted lenticule intrastromal keratoplasty (LIKE) in correcting moderate to high hyperopia. Intraoperative effective optical zone (EOZ), centration offset, and postoperative higher order aberrations (HOAs) were analyzed to better understand factors affecting postoperative outcomes.

Methods: This was a prospective, consecutive case series study of LIKE for correcting hyperopia in one department from 2018 to 2023.

View Article and Find Full Text PDF