Background: This study aimed to evaluate the imaging performance and diagnostic value of a photon-counting detector (PCD) computed tomography (CT) compared to an energy-integrating detector (EID) and flat panel detector (FPD) for cadaveric wrist arthrographies.

Methods: Following ethics committee approval, ten cadaveric wrists were injected with diluted iodinated contrast agent. CT arthrographies using PCD-, EID-, and FPD-CT were performed. Six dose protocols between 0.1 mGy (using a tin filter) and 6 mGy, ultrahigh-resolution-mode, and two reconstruction kernels were used for the PCD-CT and EID-CT. FPD-CT images were reconstructed using a "normal" and "sharp" kernel. Signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated and analyzed using analysis of variance (ANOVA) and post hoc tests. Three blinded radiologists independently rated image quality concerning trabecular, cartilage, and intrinsic structures. Intraclass correlation coefficients (ICC) were calculated, followed by a Friedman and post hoc test.

Results: At 1.5 mGy, 3 mGy, and 6 mGy with the Br89 kernel, the PCD-CT yielded up to 2.35 times higher SNR and up to 7 times higher CNR than dose-equivalent and higher dose EID-CT scans. Subjective ratings favored the PCD-CT over the EID-CT and occasionally the FPD-CT, with a combined ICC of 0.942. Applying sharper kernels, SNR did not differ significantly between the PCD-CT (1.5 mGy, 3 mGy, and 6 mGy) and the FPD-CT.

Conclusion: Using sharp kernels, the PCD-CT provided superior image quality to the EID-CT and achieved comparable or better quality than the FPD at certain parameters. Thus, the PCD-CT could be considered a possible alternative in clinical routine for evaluating wrist injuries.

Relevance Statement: This study demonstrates the potential of the PCD-CT as a valuable tool in diagnosing wrist injuries. Its superior image quality compared to the EID-CT can increase confidence in diagnosing subtle bone pathologies and additionally yields the possibility of radiation exposure reduction.

Key Points: The technical advantages of the PCD-CT allow for dose reduction while generating high-quality images. PCD-CT showed superior image quality over EID-CT and was comparable to the FPD-CT. PDC-CT offers improved visualization of fine joint structures in wrist arthrography and should be considered in clinical routine.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12397008PMC
http://dx.doi.org/10.1186/s41747-025-00604-yDOI Listing

Publication Analysis

Top Keywords

image quality
16
superior image
12
pcd-ct
9
energy-integrating detector
8
cadaveric wrist
8
wrist arthrography
8
tin filter
8
dose reduction
8
kernels pcd-ct
8
pcd-ct eid-ct
8

Similar Publications

Purpose: Real‑time magnetic resonance-guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.

View Article and Find Full Text PDF

Introduction: Medical physicists play a critical role in ensuring image quality and patient safety, but their routine evaluations are limited in scope and frequency compared to the breadth of clinical imaging practices. An electronic radiologist feedback system can augment medical physics oversight for quality improvement. This work presents a novel quality feedback system integrated into the Epic electronic medical record (EMR) at a university hospital system, designed to facilitate feedback from radiologists to medical physicists and technologist leaders.

View Article and Find Full Text PDF

Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning.

View Article and Find Full Text PDF

Background: Fetal MRI is increasingly used to investigate fetal lung pathologies, and super-resolution (SR) algorithms could be a powerful clinical tool for this assessment. Our goal was to investigate whether SR reconstructions result in an improved agreement in lung volume measurements determined by different raters, also known as inter-rater reliability.

Materials And Methods: In this single-center retrospective study, fetal lung volumes calculated from both SR reconstructions and the original images were analyzed.

View Article and Find Full Text PDF

Background And Purpose: To date, no consensus guidelines have been published that systematically guide delineation of primary and nodal Clinical Target Volumes (CTVs) in patients who require post-operative radiotherapy (PORT) for mucosal Head and Neck squamous cell carcinoma (HNSCC). As a result, significant individual, institutional and national variation exists in the way that CTVs are delineated in the post-operative setting, leading to considerable heterogeneity in radiotherapy treatment.

Methods: A multi-disciplinary group of experts convened by the European Society for Radiotherapy and Oncology (ESTRO) set-out principles for the multi-disciplinary management of oral cavity squamous cell carcinoma (OCSCC).

View Article and Find Full Text PDF