Article Abstract

Purpose: The study aimed to assess the feasibility and image quality of dual-source photon-counting detector computed tomography (PCD-CT) in evaluating small-sized coronary artery stents with respect to different acquisition modes in a phantom model.

Methods: Utilizing a phantom setup mimicking the average patient's water-equivalent diameter, we examined six distinct coronary stents inflated in a silicon tube, with stent sizes ranging from 2.0 to 3.5 mm, applying four different CT acquisition modes of a dual-source PCD-CT scanner: "high-pitch," "sequential," "spiral" (each with collimation of 144 × 0.4 mm and full spectral information), and "ultra-high-resolution (UHR)" (collimation of 120 × 0.2 mm and no spectral information). Image quality and diagnostic confidence were assessed using subjective measures, including a 4-point visual grading scale (4 = excellent; 1 = non-diagnostic) utilized by two independent radiologists, and objective measures, including the full width at half maximum (FWHM).

Results: A total of 24 scans were acquired, and all were included in the analysis. Among all CT acquisition modes, the highest image quality was obtained for the UHR mode [median score: 4 (interquartile range (IQR): 3.67-4.00)] ( = 0.0015, with 37.5% rated as "excellent"), followed by the sequential mode [median score: 3.5 (IQR: 2.84-4.00)], = 0.0326 and the spiral mode [median score: 3.0 (IQR: 2.53-3.47), > 0.05]. The lowest image quality was observed for the high-pitch mode [median score: 2 (IQR: 1- 3), = 0.028]. Similarly, diagnostic confidence for evaluating stent patency was highest for UHR and lowest for high-pitch ( < 0.001, respectively). Measurement of stent dimensions was accurate for all acquisition modes, with the UHR mode showing highest robustness (FWHM for sequential: 0.926 ± 0.061 vs. high-pitch: 0.990 ± 0.083 vs. spiral: 0.962 ± 0.085 vs. UHR: 0.941 ± 0.036, = non-significant, respectively).

Conclusion: Assessing small-sized coronary stents using PCD-CT technology is feasible. The UHR mode offers superior image quality and diagnostic confidence, while all modes show consistent and accurate measurements.

Clinical Significance: These findings highlight the potential of PCD-CT technology, particularly the UHR mode, to enhance non-invasive coronary stent evaluation. Confirmatory research is necessary to influence the guidelines, which recommend cardiac CT only for stents of 3 mm or larger.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11701693PMC
http://dx.doi.org/10.4274/dir.2024.242893DOI Listing

Publication Analysis

Top Keywords

image quality
24
acquisition modes
16
uhr mode
16
mode [median
16
[median score
16
coronary stents
12
diagnostic confidence
12
score iqr
12
dual-source photon-counting
8
computed tomography
8

Similar Publications

Purpose: Real‑time magnetic resonance-guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.

View Article and Find Full Text PDF

Introduction: Medical physicists play a critical role in ensuring image quality and patient safety, but their routine evaluations are limited in scope and frequency compared to the breadth of clinical imaging practices. An electronic radiologist feedback system can augment medical physics oversight for quality improvement. This work presents a novel quality feedback system integrated into the Epic electronic medical record (EMR) at a university hospital system, designed to facilitate feedback from radiologists to medical physicists and technologist leaders.

View Article and Find Full Text PDF

Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning.

View Article and Find Full Text PDF

Background: Fetal MRI is increasingly used to investigate fetal lung pathologies, and super-resolution (SR) algorithms could be a powerful clinical tool for this assessment. Our goal was to investigate whether SR reconstructions result in an improved agreement in lung volume measurements determined by different raters, also known as inter-rater reliability.

Materials And Methods: In this single-center retrospective study, fetal lung volumes calculated from both SR reconstructions and the original images were analyzed.

View Article and Find Full Text PDF

Background And Purpose: To date, no consensus guidelines have been published that systematically guide delineation of primary and nodal Clinical Target Volumes (CTVs) in patients who require post-operative radiotherapy (PORT) for mucosal Head and Neck squamous cell carcinoma (HNSCC). As a result, significant individual, institutional and national variation exists in the way that CTVs are delineated in the post-operative setting, leading to considerable heterogeneity in radiotherapy treatment.

Methods: A multi-disciplinary group of experts convened by the European Society for Radiotherapy and Oncology (ESTRO) set-out principles for the multi-disciplinary management of oral cavity squamous cell carcinoma (OCSCC).

View Article and Find Full Text PDF