Impact of screw tip design on screw anchorage: mechanical testing and numerical simulation.

J Orthop Surg Res

Department of Orthopedics, Luzhou Key Laboratory of Orthopedic Disorders, The Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University, Luzhou, Sichuan Province, 646000, P.R. China.

Published: July 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Screw loosening is a commonly reported issue following spinal screw fixation and can lead to various complications. The initial cause of screw loosening is biomechanical deterioration. Previous studies have demonstrated that modifications in screw design can impact the local biomechanical environment, specifically the stress distribution on bone-screw interfaces. There are several different designs of screw tips available for clinically used pedicle screws; however, it remains unclear whether these variations affect the local stress distribution and subsequent screw anchorage ability.

Methods: This study conducted comprehensive biomechanical research using polyurethane foam mechanical tests and corresponding numerical simulations to investigate this topic. Models of pedicle screw-fixed osteoporotic polyurethane foam were created with two different clinically used screw tip designs (flat and steep) featuring varying tip lengths, taper angles, and diameters, as well as identical flank overlap areas and thread designs. The anchorage ability of the different models was assessed through toggle and pull-out test. Additionally, numerical mechanical models were utilized to compute the stress distributions at the screw and bone-screw interfaces in the different models.

Results: Mechanical tests revealed superior anchorage ability in models utilizing flat-tipped screws. Furthermore, numerical modeling indicated improved anchorage ability and reduced stress concentration tendency in these models.

Conclusion: Changes in screw tip design can significantly impact the biomechanical anchoring capability of screws. Specifically, flatter tip pedicle screws may mitigate the risk of screw loosening by alleviating stress concentration on bone-screw interfaces.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11287948PMC
http://dx.doi.org/10.1186/s13018-024-04841-6DOI Listing

Publication Analysis

Top Keywords

screw design
12
screw loosening
12
bone-screw interfaces
12
anchorage ability
12
screw
11
screw anchorage
8
design impact
8
stress distribution
8
pedicle screws
8
polyurethane foam
8

Similar Publications

Two-Phase Material Shape Optimization of an Additively Manufactured Integrated Metal and Ceramic Resin Implant-Supported Dental Crown.

Int J Numer Method Biomed Eng

September 2025

Department of Industrial and Manufacturing Engineering, Pennsylvania State University, University Park, Pennsylvania, USA.

The screw-retained implant-supported crown is a durable, aesthetic restoration, but debonding between the crown and abutment remains a challenge to survivability. The purpose of this work was to devise an abutment shape that can be embedded into the crown while the crown is being additively manufactured. The result was a mechanically retained, no-adhesive abutment and crown unit that is mounted to the implant fixture.

View Article and Find Full Text PDF

Study Design: This is a retrospective single-center study.

Purpose: The purpose is to investigate the incidence of distal junctional kyphosis (DJK) when fused proximal to the stable sagittal vertebra (SSV) in adolescent idiopathic scoliosis (AIS) patients undergoing selective thoracic fusion.

Methods: We retrospectively reviewed a consecutive cohort of surgically treated AIS patients with Lenke 1-2 A/B curves between 2011 and 2022 with a minimum of 2 years of follow-up.

View Article and Find Full Text PDF

Bio-functional properties of extruded corn gluten meal after simulated gastrointestinal digestion.

Food Res Int

November 2025

Ciência e Tecnologia de Alimentos, Centro de Ciências Agrárias, Universidade Estadual de Londrina, Celso Garcia Cid, PR-445, Km 380 - University Campus, Londrina, PR 86057-970, Brazil. Electronic address:

The objective of the research was to employ extrusion to increase the yield of simulated gastrointestinal digestion of protein corn gluten meal (CG). A single-screw extruder and a full factorial design with two center points were used. The optimal extrusion parameters were 40 % sample moisture, 140 °C and 54 rpm, resulting in a gastrointestinal digestion yield of 37.

View Article and Find Full Text PDF

Plasticity Mechanisms in Nanostructured Cubic Boron Nitride: Internal Defects and Amorphous Layers.

ACS Appl Mater Interfaces

September 2025

School of Materials Science and Engineering, Beihang University, Beijing 100191, P. R. China.

Nanostructured cubic boron nitride (NS-cBN) has attracted significant attention due to its high hardness and excellent thermal stability, yet a systematic strategy to balance strength and toughness through atomically structural design remains elusive. Here, we integrate plasticity theory with large-scale atomistic simulations to elucidate the size-dependent roles of internal defects, i.e.

View Article and Find Full Text PDF

Posterior lumbar screw fixation is the most common surgical method for lumbar disc herniation, but patients often face multiple complications postoperatively. The occurrence of screw track loosening can lead to fusion failure and even life-threatening screw track extrusion. However, there is currently a lack of animal models specifically targeting changes in the screw track following lumbar screw fixation.

View Article and Find Full Text PDF