Correlation between vertebral bone mineral density and multi-level virtual non-calcium imaging parameters from dual-layer spectral detector computed tomography.

Quant Imaging Med Surg

Department of Diagnostic Radiology, Graduate Collaborative Training Base of Hunan Cancer Hospital, Hengyang Medical School, University of South China, Hengyang, China.

Published: June 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Virtual non-calcium (VNCa) imaging based on dual-energy computed tomography (CT) plays an increasingly important role in diagnosing spinal diseases. However, the utility of VNCa technology in the measurement of vertebral bone mineral density (BMD) is limited, especially the VNCa CT value at multiple calcium suppression levels and the slope of VNCa curve. This retrospective cross-sectional study aimed to explore the correlation between vertebral BMD and new VNCa parameters from dual-layer spectral detector CT.

Methods: The dual-layer spectral detector CT and quantitative CT (QCT) data of 4 hydroxyapatite (HAP) inserts and 667 vertebrae of 234 patients (132 male and 102 female) who visited a university teaching hospital between April and May 2023 were retrospectively analyzed. The BMD values of 3 vertebrae (T12, L1, and L2) and inserts were measured using QCT, defined as QCT-BMD. The VNCa CT values and the slope λ of the VNCa attenuation curve of vertebrae and inserts were recorded. The correlations between VNCa parameters (VNCa CT value, slope λ) and QCT-BMD were analyzed.

Results: For the vertebrae, the correlation coefficient ranged from -0.904 to 0.712 (all P<0.05). As the calcium suppression index (CaSI) increased, the correlation degree exhibited a decrease first and then increased, with the best correlation (r=-0.904, P<0.001) observed at the index of 25%. In contrast, the correlation coefficient for the inserts remained relatively stable (r=-0.899 to -1, all P<0.05). For the vertebrae, the values of 3 slopes λ (λ1, λ2, and λ3) derived from the VNCa attenuation curve were 6.50±1.99, 3.75±1.15, and 2.04±0.62, respectively. Regarding the inserts, the λ1, λ2, and λ3 values were 11.56 [interquartile range (IQR): 2.40-22.62], 6.68 (IQR: 1.39-13.49), and 3.63 (IQR: 0.75-7.8), respectively. For the vertebrae, all 3 correlation coefficients between 3 slopes λ and QCT-BMD were 0.956 (all P<0.05). For the inserts, the 3 correlation coefficients were 0.996, 0.998, and 1 (all P<0.05), respectively.

Conclusions: A promising correlation was detected between VNCa CT parameters and QCT-BMD in vertebrae, warranting further investigation to explore the possibility of VNCa imaging to assess BMD.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11151250PMC
http://dx.doi.org/10.21037/qims-23-1543DOI Listing

Publication Analysis

Top Keywords

dual-layer spectral
12
spectral detector
12
vnca
9
correlation vertebral
8
vertebral bone
8
bone mineral
8
mineral density
8
virtual non-calcium
8
parameters dual-layer
8
computed tomography
8

Similar Publications

Objectives: To evaluate whether dual-layer spectral computed tomography, compared with conventional CT, improves diagnostic accuracy for osteolytic vertebral metastases. Furthermore, to investigate the influence of dual-layer CT on the subjective visibility of metastases.

Materials And Methods: In this single-center retrospective study, consecutive patients with an untreated primary tumor who underwent dual-layer CT and either MRI or PET-CT as reference standard within 14 days were included.

View Article and Find Full Text PDF

Background: Accurate preoperative T staging is essential for determining optimal treatment strategies in colorectal cancer (CRC). Low-keV virtual monoenergetic images (VMIs) have been shown to enhance lesion conspicuity. This study aimed to assess the diagnostic value of dual-layer spectral computed tomography (CT)-derived VMIs, in combination with multiplanar reformation (MPR) and evaluation of peritumoral fat stranding (PFS), for improving the accuracy of T staging in CRC.

View Article and Find Full Text PDF

Objectives: Reducing motion artifacts in coronary computed tomography angiography (CCTA) is essential for accurate coronary artery disease assessment. We evaluated the clinical performance of a motion-compensated reconstruction (MCR) using subjective image quality (SIQ) and interpretability of CCTA at varying heart rates (HR).

Materials And Methods: We retrospectively identified 150 patients, grouped by HR (≤ 60, 60-69, ≥ 70 bpm, n = 50 each), referred for prospective ECG-gated CCTA on a spectral dual-layer CT.

View Article and Find Full Text PDF

Background: This study aimed to compare the diagnostic performance of radiomic features derived from dual-layer spectral detector computed tomography (DLSCT) and a deep learning (DL) model applied to conventional CT images in the differentiation of osteoblastic bone metastases (OBM) from bone islands (BI).

Methods: This retrospective study included patients with osteogenic lesions who underwent DLSCT examinations between March 2023 and September 2023. We extracted first-order radiomic features (e.

View Article and Find Full Text PDF

Background: Computed tomography (CT) is the preferred imaging modality for evaluating adrenal lesions; however, the associated radiation exposure remains a significant concern. Dual-layer spectral detector CT (SDCT)-derived virtual noncontrast (VNC) images may reduce radiation exposure by eliminating dedicated noncontrast scans, yet their agreement with true noncontrast (TNC) imaging remains debated. This study aimed to quantitatively evaluate the agreement and image quality of VNC images [reconstructed from the arterial phase (VNCa) and portal venous phase (VNCp)] compared to TNC images in adrenal adenomas stratified by lipid content, and to assess the radiation dose reduction.

View Article and Find Full Text PDF