A new generation cone-beam computed tomography (CBCT) system with new hardware design and advanced image reconstruction algorithms is available for radiation treatment simulation or adaptive radiotherapy (HyperSight CBCT imaging solution, Varian Medical Systems-a Siemens Healthineers company). This study assesses the CBCT image quality metrics using the criteria routinely used for diagnostic CT scanner accreditation as a first step towards the future use of HyperSight CBCT images for treatment planning and target/organ delineations. Image performance was evaluated using American College of Radiology (ACR) Program accreditation phantom tests for diagnostic computed tomography systems (CTs) and compared HyperSight images with a standard treatment planning diagnostic CT scanner (Siemens SOMATOM Edge) and with existing CBCT systems (Varian TrueBeam version 2.7 and Varian Halcyon version 2.0). Image quality performance for all Varian HyperSight CBCT vendor-provided imaging protocols were assessed using ACR head and body ring CT phantoms, then compared to existing imaging modalities. Image quality analysis metrics included contrast-to-noise (CNR), spatial resolution, Hounsfield number (HU) accuracy, image scaling, and uniformity. All image quality assessments were made following the recommendations and passing criteria provided by the ACR. The Varian HyperSight CBCT imaging system demonstrated excellent image quality, with the majority of vendor-provided imaging protocols capable of passing all ACR CT accreditation standards. Nearly all (8/11) vendor-provided protocols passed ACR criteria using the ACR head phantom, with the Abdomen Large, Pelvis Large, and H&N vendor-provided protocols produced HU uniformity values slightly exceeding passing criteria but remained within the allowable minor deviation levels (5-7 HU maximum differences). Compared to other existing CT and CBCT imaging modalities, both HyperSight Head and Pelvis imaging protocols matched the performance of the SOMATOM CT scanner, and both the HyperSight and SOMATOM CT substantially surpassed the performance of the Halcyon 2.0 and TrueBeam version 2.7 systems. Varian HyperSight CBCT imaging system could pass almost all tests for all vendor-provided protocols using ACR accreditation criteria, with image quality similar to those produced by diagnostic CT scanners and significantly better than existing linac-based CBCT imaging systems.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11087172 | PMC |
http://dx.doi.org/10.1002/acm2.14299 | DOI Listing |
J Appl Clin Med Phys
September 2025
Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.
Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning.
View Article and Find Full Text PDFMed Phys
September 2025
Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, USA.
Background: Online adaptive radiation therapy (ART) offers a paradigm shift in radiotherapy by enabling adjustments to the planned dose based on daily anatomical variation. In the context of cone-beam computed tomography (CBCT) for online ART on a standard linac, thoracic and abdominal treatment sites in particular present unique challenges due to the typically large treatment volumes, mobile anatomy, scatter-induced image quality degradation, and hounsfield unit (HU) limitations. A recent hardware and software upgrade for a standard linac, Varian TrueBeam (TB) v4.
View Article and Find Full Text PDFPhys Med
August 2025
Abano Terme Hospital, Department of Radiation Oncology, Abano Terme, Padua, Italy. Electronic address:
Background/introduction: This study evaluates the dosimetric performance of HyperSight CBCT (HS-CBCT) images for CBCT-only adaptive radiotherapy workflows.
Methods: We evaluated Hounsfield unit (HU) accuracy and dose calculation performance using both Feldkamp-Davis-Kress (FDK) and Acuros reconstructions in phantom and clinical datasets (brain, breast, pelvis). HU deviations were reported relative to planning CT.
Z Med Phys
July 2025
Department of Radiotherapy and Radiation Oncology, University Medical Center Mannheim, Heidelberg University, Theodor‑Kutzer‑Ufer 1‑3, 68167 Mannheim, Germany.
Purpose: Due to the prevalence of daily cone-beam computed tomography (CBCT) imaging in radiation therapy, radiomics analysis has great potential to detect early radiation induced tissue changes. Clinical applications of radiomics using CBCT imaging have been hindered by lack of stability in radiomics features and comparably poor image quality. Novel CBCT imaging devices promise improved quality comparable to those of fan-beam CTs.
View Article and Find Full Text PDFJ Appl Clin Med Phys
August 2025
Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.
Introduction: Using diagnostic CT for radiotherapy (RT) planning can bypass traditional CT simulation but introduces challenges in patient positioning and Hounsfield unit (HU) fidelity, affecting dose accuracy. Here a Varian Ethos adaptive-driven CT direct-to-treatment (DtT) multi-fraction stereotactic body radiation therapy (SBRT) workflow is presented.
Methods: This study employed institutional diagnostic PET-CT images to generate an initial reference Ethos treatment plan.