Article Abstract

Background: MAR algorithms have not been productized in interventional imaging because they are too time-consuming. Application of a beam hardening filter can mitigate metal artifacts and doesn't increase computational burden. We evaluate the ability to reduce metal artifacts of a 0.5 mm silver (Ag) additional filter in a Multidetector Computed Tomography (MDCT) scanner during CT-guided biopsy procedures.

Methods: A biopsy needle was positioned inside the lung field of an anthropomorphic phantom (Lungman, Kyoto Kagaku, Kyoto, Japan). CT acquisitions were performed with beam energies of 100 kV, 120 kV, 135 kV, and 120 kV with the Ag filter and reconstructed using a filtered back projection algorithm. For each measurement, the CTDIvol was kept constant at 1 mGy. Quantitative profiles placed in three regions of the artifact (needle, needle tip, and trajectory artifacts) were used to obtain metrics (FWHM, FWTM, width at - 100 HU, and absolute error in HU) to evaluate the blooming artifact, artifact width, change in CT number, and artifact range. An image quality analysis was carried out through image noise measurement. A one-way analysis of variance (ANOVA) test was used to find significant differences between the conventional CT beam energies and the Ag filtered 120 kV beam.

Results: The 120 kV-Ag is shown to have the shortest range of artifacts compared to the other beam energies. For needle tip and trajectory artifacts, a significant reduction of - 53.6% (p < 0.001) and - 48.7% (p < 0.001) in the drop of the CT number was found, respectively, in comparison with the reference beam of 120 kV as well as a significant decrease of up to - 34.7% in the artifact width (width at - 100 HU, p < 0.001). Also, a significant reduction in the blooming artifact of - 14.2% (FWHM, p < 0.001) and - 53.3% (FWTM, p < 0.001) was found in the needle artifact. No significant changes (p > 0.05) in image noise between the conventional energies and the 120 kV-Ag were found.

Conclusions: A 0.5 mm Ag additional MDCT filter demonstrated consistent metal artifact reduction generated by the biopsy needle. This reduction may lead to a better depiction of the target and surrounding structures while maintaining image quality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10926571PMC
http://dx.doi.org/10.1186/s42490-024-00076-yDOI Listing

Publication Analysis

Top Keywords

beam energies
12
metal artifacts
8
needle trajectory
8
trajectory artifacts
8
needle
5
artifacts
5
needle artifact
4
artifact reduction
4
reduction interventional
4
interventional procedures
4

Similar Publications

Dynamic dual-mode terahertz device with nonvolatile switching for integrated on-chip and free-space applications.

Microsyst Nanoeng

September 2025

Center for Terahertz Waves, College of Precision Instrument and Optoelectronics Engineering, and the Key Laboratory of Optoelectronics Information and Technology (Ministry of Education), Tianjin University, Tianjin, 300072, China.

Terahertz communication systems demand versatile devices capable of simultaneously controlling propagating waves and surface plasmon polaritons (SPPs) in far-field (FF) and near-field (NF) channels, yet existing solutions are constrained by volatile operation, single-function limitations, and the inability to integrate NF and FF functionalities. Here, we present a nonvolatile reconfigurable terahertz metasurface platform leveraging the phase-change material GeSbTe(GST) to achieve on-demand dual-channel modulation-a first in the terahertz regime. By exploiting the stark conductivity contrast of GST between amorphous and crystalline states, our design enables energy-efficient switching between NF-SPP manipulation and FF-wavefront engineering without requiring continuous power input.

View Article and Find Full Text PDF

A novel medium-current (up to 20 mA), low normalized beam emittance (<1 π mm mrad) electron cyclotron resonance microwave H+ ion source has been developed at the Center for Energy Research in Budapest, Hungary. This high-stability design targets an energy ripple below 1% while delivering a continuous or pulsed proton beam with adjustable pulse duration (0.1-10 ms) and frequency (0.

View Article and Find Full Text PDF

We introduce a novel method using a kilohertz (kHz) amplified 800 nm laser for the first experimental confinement of microparticles within a single beam. This study demonstrates that high-energy kHz pulses can confine 1-μm-radius polystyrene beads in water within ∼26 μm. This approach utilizes the unique properties of high-energy pulsed lasers, distinct from continuous-wave and megahertz pulsed lasers traditionally used in optical trapping.

View Article and Find Full Text PDF

The High-Intensity Proton Accelerator Facility at the Paul Scherrer Institute (PSI) accelerates protons to an energy of 590 MeV with currents up to 2.4 mA, i.e.

View Article and Find Full Text PDF

Ion irradiation has routinely been used to create defects or even pattern two-dimensional (2D) materials. For efficient defect engineering, that is, choosing the proper ion fluence to achieve the desired concentration of defects, it is of paramount importance to know the probability of creating defects as a function of ion energy. Atomistic simulations of ion impacts on 2D targets can provide such information, especially for free-standing systems, but in the case of supported 2D materials, the substrate can strongly affect defect production.

View Article and Find Full Text PDF