Purpose: Experimental measurements of two-dimensional (2D) dose rate distributions in proton pencil beam scanning (PBS) FLASH radiation therapy (RT) are currently lacking. In this study, we characterize a newly designed 2D strip-segmented ionization chamber array (SICA) with high spatial and temporal resolution and demonstrate its applications in a modern proton PBS delivery system at both conventional and ultrahigh dose rates.

Methods: A dedicated research beamline of the Varian ProBeam system was employed to deliver a 250-MeV proton PBS beam with nozzle currents up to 215 nA. In the research and clinical beamlines, the spatial, temporal, and dosimetric performances of the SICA were characterized and compared with measurements using parallel-plate ion chambers (IBA PPC05 and PTW Advanced Markus chamber), a 2D scintillator camera (IBA Lynx), Gafchromic films (EBT-XD), and a Faraday cup. A novel reconstruction approach was proposed to enable the measurement of 2D dose and dose rate distributions using such a strip-type detector.

Results: The SICA demonstrated a position accuracy of 0.12 ± 0.02 mm at a 20-kHz sampling rate (50 μs per event) and a linearity of R  > 0.99 for both dose and dose rate with nozzle beam currents ranging from 1 to 215 nA. The 2D dose comparison to the film measurement resulted in a gamma passing rate of 99.8% (2 mm/2%). A measurement-based proton PBS 2D FLASH dose rate distribution was compared to simulation results and showed a gamma passing rate of 97.3% (2 mm/2%).

Conclusions: The newly designed SICA demonstrated excellent spatial, temporal, and dosimetric performances and is well suited for commissioning, quality assurance, and a wide range of clinical applications in proton PBS clinical and FLASH-RT.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15706DOI Listing

Publication Analysis

Top Keywords

dose rate
16
proton pbs
16
spatial temporal
12
ionization chamber
8
chamber array
8
proton pencil
8
pencil beam
8
beam scanning
8
dose
8
rate distributions
8

Similar Publications

Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning.

View Article and Find Full Text PDF

Remimazolam is an ultrashort-acting benzodiazepine approved for procedural sedation in 2020 by the US Food and Drug Administration; however, dosing information originating from the US is limited. No existing literature details infusion rates of remimazolam delivered by manually adjusted infusion pumps. This case report describes the administration of manually infused remimazolam to an 86-year-old man for the surgical extraction of third molars under procedural sedation.

View Article and Find Full Text PDF

Crowding can result in greater disease transmission, yet crowded hosts may also remove infectious propagules from the environment, thereby lowering the encounter rate and infectious dose received by conspecifics. We combined experimental and modelling work to examine the impact of crowding of butterfly larvae on the per-capita risk of infection by a protozoan that is transmitted via the larval food plant, and the resulting infection load in adult butterflies. We reared larvae at different densities and exposed them to low and high doses of parasites.

View Article and Find Full Text PDF

The flavonoid rutin protects against imidacloprid-induced osmotic and electric disruptions in Africanized honey bees.

PLoS One

September 2025

Departamento de Biología, Escuela de Ciencias e Ingeniería, Universidad del Rosario, Bogotá, Colombia.

Honey bees (Apis mellifera) are essential pollinators threatened by sublethal effects of pesticides such as imidacloprid, a widely used neonicotinoid that disrupts the central nervous system. However, many of the systemic effects are poorly understood, especially on the physiological homeostasis of the honey bee. We evaluated the effects of oral administration of imidacloprid and the flavonol rutin on the properties of extracellular fluid (ECF) in Apis mellifera.

View Article and Find Full Text PDF