Article Abstract

Purpose: Creatine chemical exchange saturation transfer (CrCEST) MRI is used increasingly in muscle imaging. However, the CrCEST measurement depends on the RF saturation duration (Ts) and relaxation delay (Td), and it is challenging to compare the results of different scan parameters. Therefore, this study aims to evaluate the quasi-steady-state (QUASS) CrCEST MRI on clinical 3T scanners.

Methods: T and CEST MRI scans of Ts/Td of 1 s/1 s and 2 s/2 s were obtained from a multi-compartment creatine phantom and 5 healthy volunteers. The CrCEST effect was quantified with asymmetry analysis in the phantom, whereas 5-pool Lorentzian fitting was applied to isolate creatine from phosphocreatine, amide proton transfer, combined magnetization transfer and nuclear Overhauser enhancement effects, and direct water saturation in four major muscle groups of the lower leg. The routine and QUASS CrCEST measurements were compared under two different imaging conditions. Paired Student's t-test was performed with p-values less than 0.05 considered statistically significant.

Results: The phantom study showed a substantial influence of Ts/Td on the routine CrCEST quantification (p = 0.02), and such impact was mitigated with the QUASS algorithm (p = 0.20). The volunteer experiment showed that the routine CrCEST, amide proton transfer, and combined magnetization transfer and nuclear Overhauser enhancement effects increased significantly with Ts and Td (p < 0.05) and were significantly smaller than the corresponding QUASS indices (p < 0.01). In comparison, the QUASS CrCEST MRI showed little dependence on Ts and Td, indicating its robustness and accuracy.

Conclusion: The QUASS CrCEST MRI is feasible to provide fast and accurate muscle creatine imaging.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.29223DOI Listing

Publication Analysis

Top Keywords

crcest mri
8
quass crcest
8
amide proton
8
proton transfer
8
transfer combined
8
combined magnetization
8
magnetization transfer
8
transfer nuclear
8
nuclear overhauser
8
overhauser enhancement
8

Similar Publications

Investigation of Energy Metabolism Alterations in Diabetic Muscles Using CEST MRI at 5 T.

NMR Biomed

August 2025

Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.

Diabetes mellitus (DM) causes skeletal muscle damage. Energy metabolism alterations in diabetic muscles remain to be elucidated. Chemical exchange saturation transfer (CEST) MRI is increasingly adopted in imaging muscle energy metabolism.

View Article and Find Full Text PDF

Brain temperature mapping based on chemical exchange saturation transfer signal at 2 ppm.

Quant Imaging Med Surg

January 2025

Paul. C. Lauterbur Research Centers for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.

Background: Brain temperature signifies the thermal homeostasis of the tissue, and may serve as a marker for neuroprotective therapy. Currently, it remains challenging to map the human brain temperature with high spatial resolution. The thermal dependence of chemical exchange saturation transfer (CEST) effects of endogenous labile protons may provide a promising mechanism for the absolute brain temperature imaging.

View Article and Find Full Text PDF

Elucidating metabolite and pH variations in stroke through guanidino, amine and amide CEST MRI: A comparative multi-field study at 9.4T and 3T.

Neuroimage

January 2025

F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Research Institute, Baltimore, MD, USA; Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD, USA. Electronic address:

This study aims to investigate the variations in guanidino (Guan), amine and amide chemical exchange saturation transfer (CEST) contrasts in ischemic stroke using permanent middle cerebral artery occlusion (pMCAO) and transient MCAO (tMCAO) models at high (9.4T) and clinical (3T) MRI fields. CEST contrasts were extracted using the Polynomial and Lorentzian Line-shape Fitting (PLOF) method.

View Article and Find Full Text PDF

Background: Skeletal muscle mitochondrial oxidative phosphorylation (mtOXPHOS) is important for ATP generation and its dysfunction leads to exercise intolerance. Phosphorus magnetic resonance spectroscopy (P-MRS) is a useful, noninvasive technique for mtOXPHOS assessment but has limitations. Creatine-weighted chemical exchange saturation transfer (CrCEST) MRI is a potential alternative to assess muscle bioenergetics.

View Article and Find Full Text PDF