Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Measurement accuracy has been assessed for many different blood glucose monitoring systems (BGMS) over the years by different study groups. However, the choice of the comparison measurement procedure may impact the apparent level of accuracy found in such studies.

Materials And Methods: Measurement accuracy of 18 different BGMS was assessed in a setting based on ISO 15197 using two different comparison methods in parallel: a glucose oxidase (GOD)-based and a hexokinase (HK)-based method. Accuracy limits of ISO 15197 were applied, and additional analyses were performed, including bias, linear regression, and mean absolute relative difference (MARD) to assess the impact of possible differences between comparison methods on the apparent level of accuracy.

Results: While ≈80% of BGMS met the accuracy criteria of ISO 15197 when compared with the respective manufacturers' reference measurement procedure, only two-thirds did so against both comparison methods. The mean relative bias ranged from -6.6% to +5.7% for the analysis against the GOD-based method and from -11.1% to +1.3% for the analysis against the HK-based method, whereas MARD results ranged from 3.7% to 9.8% and from 2.3% to 10.5%, respectively. Results of regression analysis showed slopes between 0.85 and 1.08 (GOD-based method) and between 0.81 and 1.01 (HK-based method).

Conclusions: The results of this study indicate that there are systematic differences between the reference measurement procedures used for BGMS calibration as well as for system accuracy assessment. Because of the potential impact on therapy of patients with diabetes resulting from these differences, further steps toward harmonization of the measurement procedures' results are important.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9445332PMC
http://dx.doi.org/10.1177/1932296820948873DOI Listing

Publication Analysis

Top Keywords

reference measurement
12
iso 15197
12
comparison methods
12
measurement procedures
8
system accuracy
8
blood glucose
8
glucose monitoring
8
monitoring systems
8
measurement accuracy
8
measurement procedure
8

Similar Publications

Background: As populations age, more knowledge is needed on people who extend their working lives. The aim of this study was to explore if prior sickness absence (> 14 days) and/or disability pension (SADP) in mental and/or somatic diagnoses were associated with time until work exit after ages 65-69 and ≥ 70, respectively, among women and men.

Methods: This prospective population-based cohort study included all 65-69-year-olds (cohort65, n = 201,263) and ≥ 70-year-olds (cohort70, n = 93,751) who were in paid work in Sweden in 2014.

View Article and Find Full Text PDF

Background: Sedentary behavior (SB) and the absence of physical activity (PA) have become increasingly prevalent in modern societies due to changes in physical and social-environmental conditions, particularly in university students. This cross-sectional study aimed to describe and identify the prevalence and correlates of self-reported and accelerometer-determined SB and PA of German university students.

Methods: A convenience sample of 532 students participated in a questionnaire survey during the lecture period in the summer term 2018.

View Article and Find Full Text PDF

Radon (Rn) is a naturally occurring radioactive gas produced by the decay of uranium-bearing minerals in rocks and soils. Long-term exposure to elevated radon levels in drinking water is associated with an increased risk of stomach and lung cancers. This study aims to assess the concentration of radon in groundwater and evaluate its potential health risks in six cancer-affected districts, i.

View Article and Find Full Text PDF

Background: Clinical communication is central to the delivery of effective, timely, and safe patient care. The use of text-based tools for clinician-to-clinician communication-commonly referred to as secure messaging-has increased exponentially over the past decade. The use of secure messaging has a potential impact on clinician work behaviors, workload, and cognitive burden.

View Article and Find Full Text PDF

Background: Direct admission can help reduce emergency department crowding, improve patient satisfaction, and decrease costs, yet there is opportunity to improve standardized processes to do so safely and efficiently. We designed and implemented a new process for urgent direct admission (UDA) at our children's hospital with the SMART (specific, measurable, achievable, relevant, time-bound) aim to increase the number of UDAs between transfer to an intensive care unit (ICU) within 12 hours from direct admission by 50% in 12 months.

Methods: We compared unanticipated ICU transfers within 12 hours of admission (outcome) before and after implementing a standardized UDA process.

View Article and Find Full Text PDF