Article Abstract

Purpose: To investigate the mechanism of action and consistency in flow characteristics of the Ahmed glaucoma valve (AGV) under simulated physiological conditions in-vitro and to evaluate whether resistance during priming has any effect on performance of the device.

Methods: Each newly opened AGV device was connected to a digital manometer and was primed with normal saline. The device was then placed in a saline bath and connected to an open manometer, a digital manometer, and an infusion pump. Saline was infused at a rate of 3 μL/min for 24 hours. Digital manometer readings were recorded at 4 Hz.

Results: Data obtained from 9 devices are presented as medians (ranges). The priming pressure was 1130 (835, 1625) mm Hg. Pressure versus time curves showed two distinct phases; transient and steady phases. The transient phase peak pressure was 24 (13, 45) mm Hg. In the steady phase, opening and closing pressures were 13 (10, 17) and 7 (4, 9) mm Hg, respectively; the valve leaflets briefly opened every 73.9 (51, 76.6) minutes and the fluctuation of pressure (difference between opening and closing pressures) was 6 (3, 9) mm Hg. The Spearman correlation coefficient between priming and opening and priming and closing pressure was ρ = -0.13 ( = 0.72) and ρ = -0.36 ( = 0.33), respectively.

Conclusions: The device showed functionality like a valve. The resistance during priming did not affect opening and closing pressures of the AGV. This study showed variable in vitro performance of the AGV.

Translational Relevance: These laboratory findings might, at least partly, explain the variability in the clinical outcome of the device.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6016431PMC
http://dx.doi.org/10.1167/tvst.7.3.19DOI Listing

Publication Analysis

Top Keywords

digital manometer
12
opening closing
12
closing pressures
12
ahmed glaucoma
8
glaucoma valve
8
resistance priming
8
phases transient
8
priming
5
pressure
5
valve
4

Similar Publications

Equipoise exists regarding the optimal method to drain pleural fluid during thoracentesis. While several institutions use wall-based automated suction, others point to the risk of excessively high suction pressures and therefore elevated barotrauma risk as a reason to avoid it. We first performed in vitro experiments involving drainage of a 1-liter saline bag using standard thoracentesis apparatus, a digital manometer, and either manual drainage (using a 60 mL syringe) or automated drainage (using wall suction at the maximum setting).

View Article and Find Full Text PDF

Low-cost single foot operated mechanical suction machine for rural health centers and hospitals.

BMC Biomed Eng

August 2025

School of Biomedical Engineering, Jimma Institute of Technology, Jimma University, Jimma, Ethiopia.

Introduction: Healthcare practitioners in low and middle-income countries encounter numerous challenges, including insufficient staffing, an unreliable electrical infrastructure, and constrained resources. Unstable availability of electricity constitutes a significant impediment to the efficacy of public health initiatives that depend on technology requiring electrical power. More than 95% of medical apparatus is procured from developed nations, with an only 13% of medical device manufacturers situated within LMIC.

View Article and Find Full Text PDF

Purpose: Despite the benefits over classical microdiscectomy, Full endoscopic lumbar discectomy (FELD) has been associated with complications attributed to changes in intracranial pressure (ICP). This study aimed to assess intracranial and epidural pressure changes at different spinal levels induced by continuous irrigation during FELD.

Materials And Methods: Eight fresh cadavers (ages 41-78) devoid of cranial or spinal trauma were utilized.

View Article and Find Full Text PDF

Airstacking is a technique to improve lung compliance and maximum insufflation capacity (MIC) in patients with neuromuscular disorders by sequentially inflating the lungs using a manual resuscitation bag. Traditional methods lack standardization and rely on subjective feedback. A pilot study established optimal pressure ranges using a digital manometer, suggesting its potential to standardize airstacking.

View Article and Find Full Text PDF

Comparative efficacy of LifeVac® and Heimlich maneuver in simulated airway obstruction.

J Pediatr (Rio J)

April 2025

Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS), School of Medicine, Porto Alegre, RS, Brazil; Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS), Infant Center Research Group, Porto Alegre, RS, Brazil. Electronic address:

Objectives: Foreign body airway obstruction is a significant cause of morbidity and mortality, especially in infants and young children. This study aims to compare the efficacy of the Heimlich maneuver and LifeVac® in a simulated environment.

Methods: A prospective experimental study was conducted using the Choking Charlie (Laerdal®) mannequin, which simulates the trunk from an adult male and is considered suitable for simulating choking events in young children.

View Article and Find Full Text PDF