Objective: We hypothesized that a new minimal ventricular pacing mode (MVP) that provides AAI/R pacing with ventricular monitoring and back-up DDD/R pacing as needed during AV block (AVB) would significantly reduce cumulative percent ventricular pacing compared to DDD/R.
Background: Conventional DDD/R mode often results in high cumulative percent ventricular pacing that may adversely affect ventricular function and increase risk of heart failure and atrial fibrillation.
Methods: MVP was made operational in 30 patients with DDD/R implantable cardioverter-defibrillators (ICDs) and no history of AVB. Patients were randomized to one week each in DDD/R and MVP. Holter monitor recordings (ECG, intracardiac electrograms, and event markers) and device diagnostics were analyzed for cumulative % atrial paced (Cum%AP), cumulative percent ventricular pacing, and frequency and duration of DDD/R pacing back-up. Diaries were used to report symptoms.
Results: Age of the study population was 61 years +/- 12 years and 83% were male. Baseline PR interval was 204 ms +/- 32 ms and programmed AV intervals (DDD/R) were 200 ms +/- 50 ms (paced)/167 ms +/- 54 ms (sensed). Cum%AP was similar between MVP and DDD/R (47.9 +/- 37 vs 46.3 +/- 36). Cumulative percent ventricular pacing was significantly lower in MVP vs DDD/R (3.79 +/- 16.3 vs 80.6 +/- 33.8, P < .0001). Back-up DDD/R pacing during MVP operation due to transient AVB occurred in 10% of patients (9.3 +/- 7.4 [range 1-15] episodes/patient-day, duration 39.7 minutes +/- 156 minutes). Fifteen percent of AV intervals during MVP operation exceeded 300 ms. No significant symptoms were reported during MVP operation.
Conclusions: MVP dramatically reduced cumulative percent ventricular pacing compared to DDD/R while maintaining AV synchrony and providing sensor-modulated atrial pacing support. Intermittent oscillations between MVP and DDD/R during transient AV block appeared safe and well tolerated.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.hrthm.2004.03.059 | DOI Listing |
Eur J Heart Fail
September 2025
Cardiology Department, University Medical Centre Ljubljana, Ljubljana, Slovenia.
Aims: There is a lack of data from randomized clinical trials comparing treatment outcomes between conduction system pacing (CSP) modalities and biventricular pacing (BVP) in symptomatic patients with refractory atrial fibrillation (AF) scheduled for atrioventricular node ablation (AVNA). The CONDUCT-AF investigates whether CSP is non-inferior to BVP in improving left ventricular ejection fraction (LVEF) and clinical outcomes in heart failure (HF) patients with symptomatic AF undergoing AVNA.
Methods: This study is an investigator-initiated, prospective, randomized, multicentre clinical trial conducted across 10 European centres, enrolling 82 patients with symptomatic AF, HF with reduced LVEF, and narrow QRS.
Heart Rhythm O2
August 2025
Department of Cardiology, Rouen University Hospital, UNIROUEN, INSERM U1096, Rouen, France.
Background: A high burden of right ventricular pacing (RVP) increases the risk of hospitalization because of heart failure. Data on predictive factors for high burden of RVP in patients with permanent pacemaker implantation (PPI) after transcatheter aortic valve replacement (TAVR) are limited.
Objective: This study aimed to identify predictors of high RVP burden in patients with current indications for PPI after TAVR.
Heart Rhythm O2
August 2025
Department of Cardiology, Triemli Hospital Zürich, Zürich, Switzerland.
Background: Leadless pacemakers (LPs) can reduce long-term complications compared with conventional devices. However, previous studies have primarily focused on single chamber right ventricular LPs.
Objective: This study aimed to evaluate the implantation, safety, and device performance characteristics in a first real-world European use of an active fixation atrial LP for either dual chamber or single chamber pacing.
Can J Cardiol
September 2025
Department of Cardiology, Ningbo No.2 Hospital, Ningbo, Zhejiang, China.
Background: During the electrode screwing process in left bundle branch pacing (LBBP), the significance of the S wave in lead V6 remains elusive. Our study analyzes the change of the S wave in lead V6 under different patterns of capture and explores its mechanisms.
Methods: This study included 243 cases with criterion of selective LBBP (SLBBP), we performed continuous pacing technique and classified the electrophysiological characteristics observed during the screwing process into four patterns: left ventricular septal pacing (LVSP), non-selective LBBP (NSLBBP) in low output and in the lower output, selective LBBP.
IEEE J Biomed Health Inform
September 2025
Identifying the onset of the QRS complex is an important step for localizing the site of origin (SOO) of premature ventricular complexes (PVCs) and the exit site of Ventricular Tachycardia (VT). However, identifying the QRS onset is challenging due to signal noise, baseline wander, motion artifact, and muscle artifact. Furthermore, in VT, QRS onset detection is especially difficult due to the overlap with repolarization from the prior beat.
View Article and Find Full Text PDF