Article Abstract

Objective: This study developed a feedback-enabled magnetic device for the management of blepharoptosis and conducted a clinical trial to evaluate its efficacy and safety.

Study Design: Prospective, two-center, single-arm, open-label clinical trial.

Setting: Tertiary referral centers.

Methods: A device primarily comprising an optical sensor, a permanent electromagnetic actuator, and an iron sheet was developed. The palpebral fissure height (PFH) was measured, and the corrective gain of the paralytic upper eyelid was calculated. Efficacy was evaluated through symmetrical blinks, the corrective gain, and the electrical activity of the frontalis muscle. Safety was evaluated through subjective sensations and complications.

Results: A total of 16 volunteers were recruited. After device-assisted correction, the patients reported subjective improvement of peripheral vision, and the mean PFH increased from 3.4 ± 1.6 to 7.2 ± 1.4 mm, with a mean gain of 3.7 ± 0.9 mm (range: 2.0-5.0 mm). The device could lift the paralytic eyelid without causing severe discomfort or complications. It also showed effects on the electrical activity of the frontalis muscle. Over the periorbital skin, posttest erythema was observed in one patient, mild hot sensation in two, and mild warm sensation in five. Subjective and objective evaluation of six patients who underwent two trials revealed no significant difference between the first and second trials (P > .05 for all seven items).

Conclusion: The concepts used in this device may provide the basis of a future device that may be of use in patients with blepharoptosis who are not suitable candidates for surgical correction. TRIAL REGISTERED AT CLINICALTRIALS.

Gov Prs: Registration number: NCT03812016.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ohn.1350DOI Listing

Publication Analysis

Top Keywords

magnetic device
8
management blepharoptosis
8
single-arm open-label
8
corrective gain
8
electrical activity
8
activity frontalis
8
frontalis muscle
8
device
6
efficacy safety
4
safety magnetic
4

Similar Publications

Multifunctional materials that simultaneously possess intrinsic magnetic and superhard properties, particularly those composed of light elements, have a wide range of applications in advanced sensors, shielding, durable devices, and other fields. However, research on the development and understanding of such materials remains limited. In this study, a series of 3D C covalent networks derived from the C fullerene precursor were theoretically designed.

View Article and Find Full Text PDF

Purpose: Real‑time magnetic resonance-guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.

View Article and Find Full Text PDF

Background And Purpose: Low-level light therapy (LLLT) has been shown to modulate recovery in patients with traumatic brain injury (TBI). However, the longitudinal impact of LLLT on brain metabolites has not been studied. The purpose of this study was to use magnetic resonance spectroscopic imaging (MRSI) to assess the metabolic response of LLLT in patients with moderate TBI at acute (within 1 week), subacute (2-3 weeks), and late-subacute (3 months) recovery phases.

View Article and Find Full Text PDF

Background MRI-derived arrhythmogenic substrate, including late gadolinium enhancement (LGE) and extracellular volume fraction (ECV), is indicative of sudden cardiac death (SCD) risk in nonischemic dilated cardiomyopathy (DCM). The relative prognostic value of LGE and ECV remains unclear. Purpose To evaluate the performance of LGE and T1 mapping in predicting SCD in patients with DCM and to explore clinical implementation.

View Article and Find Full Text PDF

A new whole-body exposure facility for a randomized, double-blind, cross-over provocation study investigating possible effects of 50 Hz magnetic field exposure on sleep and markers of Alzheimer's disease has been developed and dosimetrically analyzed. The exposure facility was custom-tailored for the sleep laboratory where the study was carried out and enables magnetic flux densities of up to 30 μT with a maximum field inhomogeneity of less than ± 20%. Exposure is applied fully software-controlled and in a blinded and randomized manner.

View Article and Find Full Text PDF