Article Abstract

Purpose: To evaluate the accuracy of intraocular pressure (IOP) monitoring by a novel contact lens sensor system (CLS) in human eyes.

Design: Cross sectional study.

Participants: Eighty eyes of 80 participants were recruited and divided into 3 groups: (1) 40 normal eyes; (2) 30 eyes with primary open-angle glaucoma (POAG) or ocular hypertension (OHT) with normal IOP (<21 mmHg), including 27 POAG eyes and 3 OHT eyes; and (3) 10 POAG/OHT eyes with high IOP (≥21 mmHg), comprising 4 POAG eyes and 6 OHT eyes.

Methods: Participants wore the CLS to enable continuous monitoring of IOP while they assumed both seated and supine positions, with each position maintained for 10 minutes. Intraocular pressure was also measured by the Goldmann applanation tonometer (GAT) while participants were seated and the Perkins applanation tonometer (PAT) in supine, both before and after CLS wear.

Main Outcome Measures: The average IOP measured by CLS during the final 1-minute of seated and supine positions was compared with IOP measured by GAT and PAT before and after CLS wear. Also, intraclass correlation coefficient and Bland-Altman analyses were performed.

Results: No significant differences were found between pre-CLS GAT and CLS in normal eyes or between all comparisons in POAG/OHT eyes with high IOP ( > 0.5). Contact lens sensor system IOP was higher than pre-CLS PAT and post-CLS GAT/PAT IOP in normal eyes ( < 0.01), and higher than pre-CLS GAT and post-CLS PAT in POAG/OHT eyes with normal IOP ( < 0.05). All IOP differences were within ± 2 mmHg. Intraclass correlation coefficient showed moderate to very strong consistency (0.51 ≤ ≤ 0.95, < 0.05) except for that between sitting CLS and post-CLS GAT in POAG/OHT eyes with high IOP. Bland-Altman analysis showed that over 80% of points were within ± 5 mmHg and over 60% within ± 3 mmHg.

Conclusions: With good agreement in IOP measurement compared with applanation tonometry in seated and supine positions, across normal and POAG/OHT eyes, the CLS can be used for fairly accurate continuous IOP monitoring.

Financial Disclosures: Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12209922PMC
http://dx.doi.org/10.1016/j.xops.2025.100826DOI Listing

Publication Analysis

Top Keywords

contact lens
12
lens sensor
12
sensor system
12
poag/oht eyes
12
iop
9
novel contact
8
intraocular pressure
8
eyes
8
normal eyes
8
normal iop
8

Similar Publications

Abtract: PURPOSE: To evaluate the correlation between corneal backscatter and visual function in patients with Fuchs endothelial corneal dystrophy (FECD).

Study Design: Prospective case series.

Methods: This study included 53 eyes from 38 patients with FECD.

View Article and Find Full Text PDF

Purpose: To report a case of acute and transient accommodative insufficiency after laser in situ keratomileusis (LASIK) due to coronavirus disease 2019 (COVID-19).

Methods: Case report and literature review.

Results: A 36-year-old man complained of acute blurred near vision 7 days after uneventful bilateral hyperopic LASIK, concurrent with the onset of COVID-19 infection.

View Article and Find Full Text PDF

Purpose: To evaluate visual outcomes after bacterial keratitis (BK) and identify predictive factors for poor prognosis at a tertiary referral center in Southern California.

Methods: This is a cross-sectional retrospective review of patients' medical records with culture-positive BK at University of California Los Angeles from January 1, 2014, to December 31, 2019. Main outcome measure was change in best-corrected visual acuity (BCVA) at 12 weeks posttreatment.

View Article and Find Full Text PDF

The ocular surface microbiome (OSM) is a low-density, low-diversity microbial ecosystem influenced by host and environmental factors, including age, hygiene, contact lens use, and systemic disease. Unlike other mucosal sites, the eye lacks a well-defined core microbiome, and its microbial composition varies significantly between individuals. Advances in metagenomics have revealed that commensals such as  and  may contribute to immune regulation and homeostasis.

View Article and Find Full Text PDF

As emerging technologies reshape both the body and how we represent it, anatomical education stands at a threshold. Virtual dissection tools, AI-generated images, and immersive platforms are redefining how students learn anatomy, while real-world bodies are becoming hybridized through implants, neural interfaces, and bioengineered components. This Viewpoint explores what it means to teach human anatomy when the body is no longer entirely natural, and the image is no longer entirely real.

View Article and Find Full Text PDF