Article Abstract

Purpose: Anaesthesia providers often complain that quantitative neuromuscular monitoring does not accurately assess neuromuscular function, a problem that can be mitigated by appropriate calibration. However, there are only very limited recommendations for the calibration of quantitative neuromuscular monitoring in clinical routine. Therefore, this multicentre prospective agreement study compared the precision of electromyography using three different calibration approaches.

Methods: Sixty patients were assigned to one of three investigational calibration approaches: calibration before anaesthesia induction, calibration during anaesthesia induction, i.e., at loss of consciousness and state entropy < 85, or uncalibrated. All patients received electromyography calibration under deep anaesthesia on the second arm (control as recommended for research). The primary endpoint was the repeatability coefficient, which describes the fluctuation of the following train-of-four (TOF) reading. It therefore provides an estimate of the precision of a measurement method. Secondary endpoints included agreement with control calibration and pain at induction.

Results: The repeatability coefficient at TOF ratios ≥ 0.8 indicated that electromyography monitoring was less precise when TOF readings were uncalibrated (0.124 ± 0.130) or with calibration during induction (0.087 ± 0.104) but was acceptable after calibration before induction (0.075 ± 0.036) compared to those measured after calibration on the contralateral arm (control: 0.072 ± 0.027, 0.061 ± 0.021, and 0.083 ± 0.063, respectively). Recall of pain at anaesthesia induction did not differ between investigational groups.

Conclusion: The findings underline the importance of thoroughly performed calibration for precise TOF readings to reliably exclude residual neuromuscular blockade. Electromyography was most precise when calibration was performed under deep anaesthesia (control). If that approach is not possible in the clinical setting, our data suggest that calibration before anaesthesia induction can be considered if previously discussed with the patient.

Clinical Trial Registration: Clinical Trials NCT04911088, registered January 6, 2021.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10877-025-01304-zDOI Listing

Publication Analysis

Top Keywords

neuromuscular monitoring
12
precision electromyography
8
prospective agreement
8
agreement study
8
quantitative neuromuscular
8
calibration anaesthesia
8
anaesthesia induction
8
calibration
7
electromyography calibration
4
calibration approach
4

Similar Publications

Background: Assessing human movement is essential for diagnosing and monitoring movement-related conditions like neuromuscular disorders. Timed function tests (TFTs) are among the most widespread types of assessments due to their speed and simplicity, but they cannot capture disease-specific movement patterns. Conversely, biomechanical analysis can produce sensitive disease-specific biomarkers, but it is traditionally confined to laboratory settings.

View Article and Find Full Text PDF

Neuromuscular diseases are often accompanied by various types of sleep-related breathing disorders, which can exacerbate the underlying condition and are associated with a poor prognosis. Early identification is essential, and interventions such as non-invasive ventilation, oxygen therapy, and respiratory rehabilitation should be initiated promptly to mitigate disease progression and improve outcomes. Nevertheless, the rates of missed and misdiagnosed cases remain common in clinical practice.

View Article and Find Full Text PDF

Purpose: Fat fraction (FF) quantification in individual muscles using quantitative MRI is of major importance for monitoring disease progression and assessing disease severity in neuromuscular diseases. Undersampling of MRI acquisitions is commonly used to reduce scanning time. The present paper introduces novel unrolled neural networks for the reconstruction of undersampled MRI acquisitions.

View Article and Find Full Text PDF

Background: Velocity-Based Training (VBT) is an emerging method in resistance training for objectively prescribing and monitoring training intensity and neuromuscular function. Given its growing popularity, assessing the validity and reliability of VBT devices is critical for strength and conditioning coaches.

Objective: The primary purpose of this review was twofold: (1) to identify and address methodological gaps in current assessments of VBT device validity and reliability, and (2) to propose and apply a novel, multi-layered, criterion-based framework-developed in collaboration with statisticians and domain experts-for evaluating these devices.

View Article and Find Full Text PDF

We report the management of a 64-year-old male with newly diagnosed bulbar-onset myasthenia gravis (MG) who was hospitalized with acute neuromuscular respiratory insufficiency. This case highlights the challenges in monitoring respiratory function in MG patients, especially in the presence of bulbar and nuchal weakness, and emphasizes the potential utility of single breath-hold time (SBHT) over forced vital capacity (FVC) as a reliable bedside monitoring tool. Despite initial stabilization with intravenous immunoglobulin (IVIG), the patient deteriorated, requiring escalation to the intensive care unit (ICU), and the clinical worsening corresponded with the SBHT rather than with FVC.

View Article and Find Full Text PDF