Depth-based registration of 3D preoperative models to intraoperative patient anatomy using the HoloLens 2.

Int J Comput Assist Radiol Surg

Department of Radiology & Nuclear Medicine, Biomedical Imaging Group Rotterdam, Erasmus MC, Rotterdam, The Netherlands.

Published: May 2025


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: In augmented reality (AR) surgical navigation, a registration step is required to align the preoperative data with the patient. This work investigates the use of the depth sensor of HoloLens 2 for registration in surgical navigation.

Methods: An AR depth-based registration framework was developed. The framework aligns preoperative and intraoperative point clouds and overlays the preoperative model on the patient. For evaluation, three experiments were conducted. First, the accuracy of the HoloLens's depth sensor was evaluated for both Long-Throw (LT) and Articulated Hand Tracking (AHAT) modes. Second, the overall registration accuracy was assessed with different alignment approaches. The accuracy and success rate of each approach were evaluated. Finally, a qualitative assessment of the framework was performed on various objects.

Results: The depth accuracy experiment showed mean overestimation errors of 5.7 mm for AHAT and 9.0 mm for LT. For the overall alignment, the mean translation errors of the different methods ranged from 12.5 to 17.0 mm, while rotation errors ranged from 0.9 to 1.1 degrees.

Conclusion: The results show that the depth sensor on the HoloLens 2 can be used for image-to-patient alignment with 1-2 cm accuracy and within 4 s, indicating that with further improvement in the accuracy, this approach can offer a convenient alternative to other time-consuming marker-based approaches. This work provides a generic marker-less registration framework using the depth sensor of the HoloLens 2, with extensive analysis of the sensor's reconstruction and registration accuracy. It supports advancing the research of marker-less registration in surgical navigation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC12055921PMC
http://dx.doi.org/10.1007/s11548-025-03328-xDOI Listing

Publication Analysis

Top Keywords

depth sensor
16
sensor hololens
12
depth-based registration
8
surgical navigation
8
registration surgical
8
registration framework
8
registration accuracy
8
marker-less registration
8
registration
7
accuracy
7

Similar Publications

The sensing of Gram-negative Extracellular Vesicles (EVs) by the innate immune system has been extensively studied in the past decade. In contrast, recognition of Gram-positive EVs by innate immune cells remains poorly understood. Comparative genome-wide transcriptional analysis in human monocytes uncovered that S.

View Article and Find Full Text PDF

Inter-modality feature prediction through multimodal fusion for 3D shape defect detection.

Neural Netw

September 2025

School of Automation and Intelligent Sensing, Shanghai Jiao Tong University, Shanghai, 200240, China; Institute of Image Processing and Pattern Recognition, Shanghai Jiao Tong University, Shanghai, 200240, China; Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai, 200240, China.

3D shape defect detection plays an important role in autonomous industrial inspection. However, accurate detection of anomalies remains challenging due to the complexity of multimodal sensor data, especially when both color and structural information are required. In this work, we propose a lightweight inter-modality feature prediction framework that effectively utilizes multimodal fused features from the inputs of RGB, depth and point clouds for efficient 3D shape defect detection.

View Article and Find Full Text PDF

Mn-doped carbon dots-based fluorescent-colorimetric dual-mode probes for selective and sensitive detection of Cr(VI) ions and l-ascorbic acid via smartphone-integrated analytical platform.

Anal Chim Acta

November 2025

Guangxi Key Laboratory of Natural Polymer Chemistry and Physics, Key Laboratory of Nanobiosensor Analysis, College of Chemistry and Materials, Nanning Normal University, Nanning, 530001, PR China. Electronic address:

Background: Hexavalent chromium ions (Cr(VI)), a notorious toxic heavy metal pollutant with proven carcinogenicity, endangers human health and the environment. Meanwhile, l-ascorbic acid (L-AA), a vital biological antioxidant, has abnormal levels closely tied to various diseases. Developing efficient synchronous detection methods for these two key analytes is of great value in clinical and environmental monitoring.

View Article and Find Full Text PDF

Applying gerontechnology to dementia care: experiences from long-term care workers in Taiwan.

Disabil Rehabil Assist Technol

September 2025

Department of Senior Citizen Services, National Tainan Junior College of Nursing, Tainan, Taiwan.

Purpose: This study explored the experiences of long-term care workers in using assistive technologies in dementia-specific care facilities in Taiwan, with a focus on perceived benefits, challenges encountered, and required support strategies.

Methods: A qualitative research design was employed. Ten female care workers from five dementia-specific long-term care institutions, each with at least 1 year of experience using assistive technologies, participated in semi-structured in-depth interviews.

View Article and Find Full Text PDF

The development of flexible gas sensors is of growing interest in wearable electronics. However, developing a gas sensor with low operating temperature, high sensitivity, and rapid response remains a huge challenge. Herein, we first develop a polyacrylamide-sodium acrylate-sodium citrate (PAM-Na-SC) hydrogel electrolyte, and design a hydrogel-based nitrogen dioxide (NO) gas sensor enabled by zinc-air batteries (ZABs).

View Article and Find Full Text PDF