Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Aims: This prospective, two-centre study derived and validated predictive algorithms for the Siemens Atellica IM high-sensitivity cardiac troponin I (hs-cTnI) assay in the emergency department (ED).

Methods And Results: Algorithms for predicting 30-day myocardial infarction (MI) Types 1 and 2 and death or non-ST-elevation MI (NSTEMI, Types 1 and 2) at index admission were developed from a derivation cohort of 1896 patients and validated using a synthetic data set with nearly 1 million patient cases. Performance was compared with the European Society of Cardiology algorithms for hs-cTnT (Roche Diagnostics) and hs-cTnI (Abbott Diagnostics). An admission hs-cTnI concentration < 5 ng/L had a negative predictive value (NPV) and sensitivity for 30-day MI or death of 99.5-99.7% and 98.1-98.8%, respectively, in the derivation cohort and validation data set. The NPV and sensitivity were ≥99.7% and ≥98.8% for ruling out index NSTEMI. A 0- to 1-h algorithm with baseline hs-cTnI concentration < 10 ng/L and Δ change < 3 ng/L had NPV of ≥99.5% and sensitivity ≥ 97.3% for predicting 30-day MI or death and a ≥99.5% sensitivity and NPV for index NSTEMI. Rule-in algorithms of either 0-h hs-cTnI ≥ 120 ng/L or 0- to 1-h Δ change ≥ 12 ng/L had positive predictive value ≥ 73% and specificity > 96% for 30-day MI or death and index NSTEMI. The results were comparable with established hs-cTn algorithms.

Conclusion: This study presents Siemens Atellica hs-cTnI algorithms for diagnosis and risk prediction in the ED with performance comparable with established hs-cTnT (Roche) and hs-cTnI (Abbott) algorithms.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11929528PMC
http://dx.doi.org/10.1093/ehjacc/zuaf017DOI Listing

Publication Analysis

Top Keywords

siemens atellica
12
30-day death
12
rule-in algorithms
8
algorithms siemens
8
atellica high-sensitivity
8
high-sensitivity cardiac
8
cardiac troponin
8
predicting 30-day
8
derivation cohort
8
data set
8

Similar Publications

Objectives: Serum thyroglobulin (Tg) is a key biomarker in the post-surgical monitoring of differentiated thyroid cancer (DTC). However, inter-assay variability among different immunoassay platforms can impact clinical interpretation, particularly at low Tg concentrations. This study aimed to compare the analytical performance and concordance of three widely used Tg immunoassays, Access (Beckman Coulter, Tg-B), Atellica (Siemens, Tg-A), and Liaison (Diasorin, Tg-L), with a focus on their agreement across clinically relevant Tg ranges.

View Article and Find Full Text PDF

Method choice for investigation of macrotroponin interference with the Siemens Atellica high sensitivity troponin I assay.

Clin Biochem

September 2025

Clinical Chemistry, Department of Pathology and Laboratory Medicine, Vancouver General Hospital, Vancouver, Canada; Department of Pathology and Laboratory Medicine, The University of British Columbia, Vancouver, Canada.

Objectives: Macrotroponin refers to circulating immunoglobulin-bound cardiac troponin species that may elevate troponin results in patients with or without myocardial injury, causing diagnostic confusion. Clinical laboratories have been recommended to provide a service for troponin interference investigation. We evaluated the applicability of a Protein A/G IgG-depletion procedure as well as polyethylene glycol (PEG) precipitation for detecting macrotroponin interference with the Siemens Atellica troponin I (TnIH) assay.

View Article and Find Full Text PDF

Background: Hepatitis B envelope Antigen (HBeAg) and anti-hepatitis B envelope Antigen (anti-HBe) are crucial markers for evaluating hepatitis B virus infection status and guiding clinical decisions. Considering the increasing prevalence of HBeAg-negative variants, accurate detection of both markers is essential. This study aimed to examine the analytical performance of four fully automated immunoanalyzers for the simultaneous detection of HBeAg and Anti-HBe and to assess the inter-platform concordance.

View Article and Find Full Text PDF

Background: Direct measurement of low-density lipoprotein cholesterol (LDL-C) is widely used and recommended by professional society guidelines despite its potential limitations in patients with hypertriglyceridemia and low LDL-C. This study evaluated the performance of 3 direct LDL-C (LDL-CD) assays, 2 modern LDL-C calculation methods [LDL-C Martin (LDL-CM), LDL-C modified Sampson (LDL-CS)] and the conventional Friedewald (LDL-CF) method against the reference method, beta-quantification (LDL-CBQ).

Methods: A total of 181 remnant sera from patients with standard lipid panel orders or from patients with LDL-CBQ orders with triglycerides (TG) ≥ 400 mg/dL (4.

View Article and Find Full Text PDF

Objectives: To assess the imprecision and stability of the point-of-care troponin I assay in the Emergency Department and its correlation and bias to two central laboratory troponin I assays (Siemens Atellica and Abbott Alinity).

Methods: Imprecision and stability testing was performed on opportunistically selected samples using whole blood in the emergency department by non-laboratory trained personnel. Assay comparisons were undertaken on samples taken from participants of the Mersey Acute Coronary syndrome Rule Out Study.

View Article and Find Full Text PDF