Article Abstract

Objectives: An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.

Methods: To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.

Results: The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).

Conclusions: This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2024-0999DOI Listing

Publication Analysis

Top Keywords

protein-based calibration
12
cardiac troponin
8
isotope dilution
8
dilution liquid
8
liquid chromatography-tandem
8
chromatography-tandem mass
8
mass spectrometry
8
spectrometry id-lc-ms/ms
8
peptide-based calibration
8
calibration strategy
8

Similar Publications

The case for calibration-free concentration analysis.

Bioanalysis

June 2025

Bristol Myers Squibb, Precision Medicine, Bioanalytical, and Translational Sciences, Princeton, NJ, USA.

Accurate measurement of protein concentration is crucial in biological research, where protein-based reagents play a key role in assay performance and reliability. Traditional methods of protein quantification often measure total protein concentration, failing to account for the active portion capable of binding to its intended target. Calibration-free concentration analysis (CFCA), which uses surface plasmon resonance (SPR) technology, offers a solution by specifically measuring the active protein concentration of the sample.

View Article and Find Full Text PDF

Vaccination against SARS-CoV-2 remains a key measure to control COVID-19. Nuvaxovid, a recombinant Matrix-M-adjuvanted protein-based vaccine, showed similar efficacy to mRNA vaccines in clinical trials and real-world studies, with lower rates of reactogenicity. To support decision making on UK vaccine selection, a population-based compartmental dynamic transmission model with a cost-utility component was developed to evaluate the cost-effectiveness of Nuvaxovid compared with mRNA vaccines from a UK National Health Service perspective.

View Article and Find Full Text PDF

Objectives: An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.

View Article and Find Full Text PDF

Metabolism-related proteins as biomarkers for predicting prognosis in polycystic ovary syndrome.

Proteome Sci

December 2024

The addresses of the institutions: Reproductive Medicine Center, Lanzhou University Second Hospital, No.82, Cuiying Road, Chengguan District, Lanzhou City, Gansu Province, China.

Objective: The study aimed to explore the role of metabolism-related proteins and their correlation with clinical data in predicting the prognosis of polycystic ovary syndrome (PCOS).

Methods: This research involves a secondary analysis of proteomic data derived from endometrial samples collected from our study group, which includes 33 PCOS patients and 7 control subjects. A comprehensive identification and analysis of 4425 proteins were conducted to screened differentially expressed proteins (DEPs).

View Article and Find Full Text PDF

GlutaR: A High-Performance Fluorescent Protein-Based Sensor for Spatiotemporal Monitoring of Glutamine Dynamics In Vivo.

Angew Chem Int Ed Engl

January 2025

State Key Laboratory of Natural and Biomimetic Drugs and Department of Chemical Biology, Department of Chemical Biology, School of Pharmaceutical Sciences, Peking University, Beijing, 100191, China.

Glutamine is the most abundant amino acid in human blood and muscle, and is integral to a wide variety of functions in cancer cells. However, the inability to monitor the subcellular distribution of glutamine in real-time has obscured understanding of glutamine metabolism under physiological and pathological conditions. Here, we report the development of a genetically encoded fluorescent sensor and demonstrate how this GlnBP-cpYFP fusion "GlutaR sensor" undergoes glutamine-induced conformational changes reflected in detectable fluorescence responses.

View Article and Find Full Text PDF