98%
921
2 minutes
20
Background A curve-shaped compression paddle could reduce the pain experienced by some women at breast cancer screening. Purpose To compare curved and standard compression systems in terms of pain experience and image quality in mammography screening. Materials and Methods In this randomized controlled trial conducted between October 2021 and February 2022, participants screened at three screening sites in the Netherlands were randomized to either a curved-paddle or sham-paddle group. The sham paddle was a standard paddle that was presented as a new paddle. At a standard screening examination, one additional image was acquired with a curved or sham paddle. Pain was measured on a numerical rating scale (range, 0-10). Participants provided a pain score after compression with the standard and test paddles, resulting in two scores per participant. Differences in pain scores were compared between groups using analysis of covariance, adjusting for pain score after standard-paddle compression. Two radiographers and two radiologists performed unblinded paired comparisons of curved-paddle vs standard-paddle images, using standard image quality criteria (radiographers evaluated 1246 image pairs using 12 criteria; radiologists evaluated 320 image pairs using six criteria). The one-sample Wilcoxon signed-rank test was used to determine if there was a significant preference for either paddle. Results In total, 2499 female participants (mean age, 61.6 years ± 7.1 [SD]) were studied; 1250 in the curved-paddle group and 1249 in the sham-paddle group. The mean pain score decreased by an additional 0.19 points in the curved-paddle group compared with the sham-paddle group (95% CI: 0.09, 0.28; < .001). In terms of image quality, the observers showed no preference or a preference for the standard paddle. Decreased image contrast (range Bonferroni-corrected values: < .001 to > .99) and visibility of structures were the main concerns for curved-paddle images. Conclusion The use of the curved paddle resulted in a minimal pain reduction during mammography breast compression but image quality was reduced. © RSNA, 2024
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1148/radiol.232680 | DOI Listing |
J Appl Clin Med Phys
September 2025
Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
Purpose: Real‑time magnetic resonance-guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.
View Article and Find Full Text PDFJ Appl Clin Med Phys
September 2025
Clinical Imaging Physics Group, Duke University Health System, Durham, North Carolina, USA.
Introduction: Medical physicists play a critical role in ensuring image quality and patient safety, but their routine evaluations are limited in scope and frequency compared to the breadth of clinical imaging practices. An electronic radiologist feedback system can augment medical physics oversight for quality improvement. This work presents a novel quality feedback system integrated into the Epic electronic medical record (EMR) at a university hospital system, designed to facilitate feedback from radiologists to medical physicists and technologist leaders.
View Article and Find Full Text PDFJ Appl Clin Med Phys
September 2025
Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.
Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning.
View Article and Find Full Text PDFEur Radiol Exp
September 2025
Center for MR-Research, University Children's Hospital Zurich, University of Zurich, Zurich, Switzerland.
Background: Fetal MRI is increasingly used to investigate fetal lung pathologies, and super-resolution (SR) algorithms could be a powerful clinical tool for this assessment. Our goal was to investigate whether SR reconstructions result in an improved agreement in lung volume measurements determined by different raters, also known as inter-rater reliability.
Materials And Methods: In this single-center retrospective study, fetal lung volumes calculated from both SR reconstructions and the original images were analyzed.
Radiother Oncol
September 2025
Dept of Radiation Oncology, Centre Léon Bérard, Lyon, France. Electronic address:
Background And Purpose: To date, no consensus guidelines have been published that systematically guide delineation of primary and nodal Clinical Target Volumes (CTVs) in patients who require post-operative radiotherapy (PORT) for mucosal Head and Neck squamous cell carcinoma (HNSCC). As a result, significant individual, institutional and national variation exists in the way that CTVs are delineated in the post-operative setting, leading to considerable heterogeneity in radiotherapy treatment.
Methods: A multi-disciplinary group of experts convened by the European Society for Radiotherapy and Oncology (ESTRO) set-out principles for the multi-disciplinary management of oral cavity squamous cell carcinoma (OCSCC).