Article Abstract

Objective: To prospectively evaluate the association between hearing preservation after cochlear implantation (CI) and intracochlear electrocochleography (ECochG) amplitude parameters.

Study Design: Multi-institutional, prospective randomized clinical trial.

Setting: Ten high-volume, tertiary care CI centers.

Patients: Adults (n = 87) with sensorineural hearing loss meeting CI criteria (2018-2021) with audiometric thresholds of ≤80 dB HL at 500 Hz.

Methods: Participants were randomized to CI surgery with or without audible ECochG monitoring. Electrode arrays were inserted to the full-depth marker. Hearing preservation was determined by comparing pre-CI, unaided low-frequency (125-, 250-, and 500-Hz) pure-tone average (LF-PTA) to LF-PTA at CI activation. Three ECochG amplitude parameters were analyzed: 1) insertion track patterns, 2) magnitude of ECochG amplitude change, and 3) total number of ECochG amplitude drops.

Results: The Type CC insertion track pattern, representing corrected drops in ECochG amplitude, was seen in 76% of cases with ECochG "on," compared with 24% of cases with ECochG "off" ( p = 0.003). The magnitude of ECochG signal drop was significantly correlated with the amount of LF-PTA change pre-CI and post-CI ( p < 0.05). The mean number of amplitude drops during electrode insertion was significantly correlated with change in LF-PTA at activation and 3 months post-CI ( p ≤ 0.01).

Conclusions: ECochG amplitude parameters during CI surgery have important prognostic utility. Higher incidence of Type CC in ECochG "on" suggests that monitoring may be useful for surgeons in order to recover the ECochG signal and preventing potentially traumatic electrode-cochlear interactions.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0000000000004286DOI Listing

Publication Analysis

Top Keywords

ecochg amplitude
24
amplitude parameters
12
hearing preservation
12
ecochg
12
amplitude
8
intracochlear electrocochleography
8
lf-pta activation
8
insertion track
8
magnitude ecochg
8
cases ecochg
8

Similar Publications

Purpose: The causes of residual hearing loss during cochlear implantation are generally poorly understood. This study sought to understand whether the complexity of this surgical approach related to a loss of residual hearing.

Method: Thirty-four adults underwent implantation with a sheathed perimodiolar electrode, via an extended round window approach.

View Article and Find Full Text PDF

Can Electrocochleography Predict Pure Tone Thresholds Without Correction Factors?

J Int Adv Otol

July 2025

Department of Otorhinolaryngology, Gülhane Faculty of Medicine, University of Health Science, Ankara, Türkiye.

BACKGROUND: Electrocochleography (EcochG) is a valuable method for determining frequency-specific objective hearing thresholds. While pure tone audiometry (PTA) is the gold standard, it can be inconclusive in certain populations, such as infants and non-responders. Auditory brainstem response is the primary electrophysiological test for threshold estimation in these groups.

View Article and Find Full Text PDF

Background/objectives: Electrocochleography (ECochG) is a promising tool to monitor preservation of cochlear structures and function during cochlear implant (CI) surgery. However, the interpretation of ECochG signal changes during insertion of the CI electrode array remains controversial. This study investigates the influence of the degree and localization of cochlear trauma on ECochG signal changes using a mouse model.

View Article and Find Full Text PDF

Objectives: The purpose of the study was (1) to retrospectively analyze ECochG data and compare it with the current literature. (2) to explore the relationship between ECochG and post-op audiological assessments.

Methods: This retrospective study included all patients with intraoperative and postoperative ECochG measurements and postoperative audiologic testing.

View Article and Find Full Text PDF

To investigate the application value of intraoperative electrocochleography(ECochG) monitoring technique and insertion techniques in cochlear implant(CI) and analyze its relationship with postoperative residual hearing(RH) preservation. Thirty-one patients(35 ears) who received CI in our hospital from June 2022 to July 2024 were enrolled. The Advanced Bionics Active Insertion Monitoring(AIM) system was used for real-time ECochG monitoring during surgery.

View Article and Find Full Text PDF