Article Abstract

Background: The labeled sizes of surgical valve prostheses and their discordance with the physical internal valve orifice sizes has long been a controversy in the cardiac surgery community, leading many to believe it to be a contributing factor in prosthesis-patient mismatch following valvular replacement surgery. In an attempt to address this issue, the International Organization for Standardization (ISO) 5840-2:2021 standard for surgical valve prostheses recommends that a new sizing parameter, namely, the effective orifice diameter, be provided in labeling by all manufacturers as an indicator of the true flow-passing capacity of a prosthetic valve.

Methods: The ISO Cardiac Valves Working Group conducted a multi-laboratory round-robin study to investigate whether the effective orifice diameter of a prosthetic surgical valve could be derived repeatably and reproducibly through steady forward-flow testing. A total of seven valve models, each with multiple sizes, were tested, including a mechanical heart valve and multiple biological heart valves.

Results: The round-robin study confirmed that the steady forward-flow test had good intra-laboratory repeatability and inter-laboratory reproducibility in deriving the effective orifice diameters of surgical valve prostheses. On average, among the participating laboratories, the experimentally derived effective orifice diameter of a prosthetic heart valve was 3-12 mm smaller than its labeled size.

Conclusions: The effective orifice diameter provides better characterization of the hydrodynamic characteristics of a surgical valve prosthesis and can be derived using a validated steady forward-flow test method. This new sizing parameter will soon be adopted by surgical valve manufacturers and provided in device labeling to inform valve selection by surgeons.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11148754PMC
http://dx.doi.org/10.21037/acs-2024-aae-0057DOI Listing

Publication Analysis

Top Keywords

surgical valve
28
effective orifice
24
orifice diameter
20
valve prostheses
16
valve
13
sizing parameter
12
steady forward-flow
12
inform valve
8
valve selection
8
round-robin study
8

Similar Publications

Severe tricuspid regurgitation (TR) can lead to significant enlargement of the right atrium (RA) and poses unique clinical challenges. We report this case of a 17-year-old boy previously misdiagnosed with Ebstein anomaly who presented with dyspnea and palpitations. Initial examination revealed irregular heart rhythm, distended neck veins, and a significant murmur.

View Article and Find Full Text PDF

We report a 64-year-old woman who developed symptomatic vasospasm on postoperative day 7 after clipping of an unruptured right middle cerebral artery (MCA) aneurysm. Imaging revealed right MCA vasospasm, which resolved with oral antiplatelets and intravenous vasodilators. She was discharged without neurological deficits.

View Article and Find Full Text PDF

Background: Long-term outcomes of transcatheter mitral valve edge-to-edge repair (TEER) are compared with medical therapy remain under investigation. This study evaluated the 3-year effects of MitraClip on mitral regurgitation (MR) severity, ventricular remodeling, and clinical outcomes in high surgical-risk patients.

Methods: A single-center retrospective cohort included 31 MitraClip patients (2016-2023) and 30 contemporaneous controls on maximally tolerated guideline-directed medical therapy.

View Article and Find Full Text PDF

Objective: Valve selection in acute type A aortic dissection (ATAAD) requiring aortic root replacement is challenging given the clinical acuity, unknown patient preferences, risk of surgical bleeding, and limited life expectancy. We sought to identify long-term outcomes of mechanical versus bioprosthetic aortic root replacement in young patients with ATAAD.

Methods: Retrospective review of our institution's database of ATAAD was conducted to identify patients aged 65 years and younger who underwent mechanical Bentall (mech-Bentall) or bioprosthetic Bentall (bio-Bentall) for ATAAD from 2002 to 2022.

View Article and Find Full Text PDF