Effects of reduction in intraocular pressure after trabeculotomy on axial length and intraocular lens selection.

J Cataract Refract Surg

From the Saitama Red Cross Hospital, Cyuo-ku, Saitama-shi Saitama, Japan (Kanda, Karakawa, Nakagawa, Ishii); Department of Ophthalmology, University of Tokyo Graduate School of Medicine, Bunkyo-ku, Tokyo, Japan (Kanda, Fujishiro, Karakawa, Nakagawa, Ishii).

Published: July 2024


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To investigate the relationship between intraocular pressure (IOP) and axial length (AL) and to compare the refractive predicted error in patients who have undergone cataract surgery alone or in combination with trabeculotomy.

Setting: Hospital.

Design: Single-center, retrospective, case-control.

Methods: The medical records of patients who had undergone cataract surgery alone or in combination with trabeculotomy using the Tanito microhook were retrospectively reviewed. Patients were grouped into cataract surgery alone (CAT) or cataract surgery combined with trabeculotomy (LOT) groups. Demographic data, preoperative and postoperative IOP and AL, and surgically induced astigmatism (SIA) were analyzed before and 1 month postoperatively to evaluate the interplay between IOP, AL, and refractive outcomes.

Results: 52 eyes (52 patients) underwent LOT, and 67 eyes (67 patients) underwent CAT. The mean IOP at baseline did not differ between the groups; the change in IOP (dIOP) was significantly higher in the LOT group than in the CAT group. The mean AL at baseline and the change in AL (dAL) were 24.0 ± 1.2 mm and 0.16 ± 0.11 mm, respectively, in the LOT group, and 23.8 ± 1.1 mm and 0.11 ± 0.070 mm, respectively, in the CAT group. The difference in dAL was also significant. In the LOT group, dIOP and dAL were significantly correlated. The mean SIA vectors did not significantly differ between the groups.

Conclusions: AL decreased because of the reduction in IOP after cataract surgery combined with trabeculotomy. Consequently, the refractive target error was greater, and the postoperative refractive outcome showed a tendency toward hyperopia.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000001447DOI Listing

Publication Analysis

Top Keywords

cataract surgery
20
lot group
12
intraocular pressure
8
axial length
8
patients undergone
8
undergone cataract
8
surgery combination
8
surgery combined
8
combined trabeculotomy
8
eyes patients
8

Similar Publications

Purpose: To evaluate the time required for refractive error (RE) stabilization after standard phacoemulsification cataract surgery and identify preoperative factors influencing this duration.

Methods: This prospective case series study enrolled patients who had undergone phacoemulsification cataract surgery. RE stabilization was defined as 0.

View Article and Find Full Text PDF

Incision capsular bag herniation during subluxated lens surgery: mechanisms, management, and prevention.

Front Med (Lausanne)

August 2025

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.

Purpose: To investigate the mechanism, intraoperative characteristics, management, and prevention of incision capsular bag herniation (ICBH), a previously unreported complication during cataract surgery in eyes with lens subluxation.

Methods: A retrospective observational case series was conducted on five male patients who developed ICBH during phacoemulsification with intraocular lens (IOL) implantation between January 2019 and December 2024. Among 867 subluxated-lens surgeries performed during this period, the estimated incidence of ICBH was 0.

View Article and Find Full Text PDF

Purpose: To report clinical outcomes in patients implanted with an isofocal optic-design intraocular lens (IOL) with double C-loop haptics following cataract surgery.

Methods: This was a multicentre-prospective-study involving 108 eyes (54 subjects) implanted with the Isopure Serenity (BVI, Inc). IOL.

View Article and Find Full Text PDF

ObjectivePseudoexfoliation syndrome is a systemic disease of unknown etiology, seen in advanced ages, characterized by extracellular material accumulation in ocular tissues and visceral organs. Autophagy, which is a basic metabolic pathway, provides macromolecule recycling of the cell and maintains cell homeostasis by adapting to the cell's stress environment. The aim of this study was to examine the relationship between specific mechanisms of autophagy and pseudoexfoliation syndrome.

View Article and Find Full Text PDF

PurposeTo evaluate the safety and ability of an ophthalmic solution containing Poloxamer 407 and Polyquaternium 133 to reduce conjunctival bacterial load before cataract surgery.MethodsPatients (n = 74) were randomized to 2 groups: treatment (n = 37) or placebo (treatment's vehicle; (n = 37)) BID from V1 to V3. Patients were also given standard postoperative treatment from V2 to V3.

View Article and Find Full Text PDF