98%
921
2 minutes
20
Background: Balance parameters derived from wearable sensor measurements during postural sway have been shown to be sensitive to experimental variables such as test duration, sensor number, and sensor location that influence the magnitude and frequency-related properties of measured center-of-mass (COM) and center-of-pressure (COP) excursions. In this study, we investigated the effects of test duration, the number of sensors, and sensor location on the reliability of standing balance parameters derived using body-mounted accelerometers.
Methods: Twelve volunteers without any prior history of balance disorders were enrolled in the study. They were asked to perform two 2-min quiet standing tests with two different testing conditions (eyes open and eyes closed). Five inertial measurement units (IMUs) were employed to capture postural sway data from each participant. IMUs were attached to the participants' right legs, the second sacral vertebra, sternum, and the left mastoid processes. Balance parameters of interest were calculated for the single head, sternum, and sacrum accelerometers, as well as, a three-sensor combination (leg, sacrum, and sternum). Accelerometer data were used to estimate COP-based and COM-based balance parameters during quiet standing. To examine the effect of test duration and sensor location, each 120-s recording from different sensor locations was segmented into 20-, 30-, 40-, 50-, 60-, 70-, 80-, 90-, 100-, and 110-s intervals. For each of these time intervals, time- and frequency-domain balance parameters were calculated for all sensor locations.
Results: Most COM-based and COP-based balance parameters could be derived reliably for clinical applications (Intraclass-Correlation Coefficient, ICC ≥ 0.90) with a minimum test duration of 70 and 110 s, respectively. The exceptions were COP-based parameters obtained using a sacrum-mounted sensor, especially in the eyes-closed condition, which could not be reliably used for clinical applications even with a 120-s test duration.
Conclusions: Most standing balance parameters can be reliably measured using a single head- or sternum-mounted sensor within a 120-s test duration. For other sensor locations, the minimum test duration may be longer and may depend on the specific test conditions.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10763154 | PMC |
http://dx.doi.org/10.1186/s12938-023-01196-7 | DOI Listing |
IEEE Trans Biomed Circuits Syst
September 2025
Neuroprostheses capable of providing Somatotopic Sensory Feedback (SSF) enables the restoration of tactile sensations in amputees, thereby enhancing prosthesis embodiment, object manipulation, balance and walking stability. Transcutaneous Electrical Nerve Stimulation (TENS) represents a primary noninvasive technique for eliciting somatotopic sensations. Devices commonly used to evaluate the effectiveness of TENS stimulation are often bulky and main powered.
View Article and Find Full Text PDFIEEE Trans Neural Netw Learn Syst
September 2025
In industrial scenarios, semantic segmentation of surface defects is vital for identifying, localizing, and delineating defects. However, new defect types constantly emerge with product iterations or process updates. Existing defect segmentation models lack incremental learning capabilities, and direct fine-tuning (FT) often leads to catastrophic forgetting.
View Article and Find Full Text PDFBiogerontology
September 2025
School of Pharmacy and Biomolecular Sciences, Liverpool John Moores University, Liverpool, UK.
Ashwagandha (Withania somnifera), a revered herb in Ayurvedic medicine, has gained significant scientific recognition for its potential to promote healthy aging. Traditionally used as a Rasayana or rejuvenator, this potent adaptogen helps the body manage stress and enhance vitality. This review synthesises extensive evidence for its multifaceted anti-aging capabilities, which target key hallmarks of the aging process.
View Article and Find Full Text PDFReproduction
October 2025
State Key Laboratory of Reproductive Medicine and Offspring Health, The Affiliated Taizhou People's Hospital of Nanjing Medical University, Taizhou School of Clinical Medicine, Nanjing Medical University, Nanjing, China.
In Brief: Lipid homeostasis is vital for maintaining testicular function and male fertility, but the specific contributions of lipid-regulating enzymes remain unclear. This study shows that DGKη, although highly expressed in the testis, is not essential for spermatogenesis but modulates testicular lipid metabolism in response to dietary conditions.
Abstract: Diacylglycerol kinase eta (DGKη), encoded by the Dgkh gene, catalyzes the phosphorylation of diacylglycerol to phosphatidic acid, both of which are key lipid second messengers.
PLoS One
September 2025
Department of design fundamentals, Faculty of Mechanical Engineering, Industrial University of Ho Chi Minh City, Ho Chi Minh City, Vietnam.
The slider-crank mechanism (SCM) is fundamental to various mechanical systems. However, optimizing its dynamic performance remains a pressing challenge due to excessive torque, joint reactions, and energy consumption. This study introduces two key innovations to address these challenges: (1) the integration of springs into SCM to optimize dynamic performance and (2) a novel hybrid optimization approach combining the Conjugate Direction with Orthogonal Shift (CDOS) method and Parameter Space Investigation (PSI).
View Article and Find Full Text PDF