Background: Utilization of the direct anterior approach (DAA) for total hip arthroplasty (THA) has risen in popularity. Accurate implant placement is a critical factor that influences patient outcomes. The goal of this study was to compare the accuracy and precision of fluoroscopy with imageless optical navigation in DAA THA.
Methods: This was a cohort study of 640 consecutive primary DAA THAs performed with intraoperative fluoroscopy (n = 300 patients) or imageless optical navigation (n = 304 patients). Accuracy was compared by measuring acetabular cup inclination, anteversion, and leg-length discrepancy (LLD). The proportion of implants placed within the Lewinnek safe zone and those placed within a more precise target of 40 ± 5° inclination and 20 ± 5° anteversion was evaluated.
Results: According to the Lewinnek criteria, there was no difference in the percentage of implants placed within both inclination and anteversion targets (fluoroscopy: 90.3%; navigation: 88.8%, P = 0.519). Using the more precise targets, navigation increased the implants positioned correctly for both inclination and anteversion (fluoroscopy: 50.3%; navigation: 65.6%, P < 0.001). Navigation increased the proportion of implants positioned within the target anteversion zone (fluoroscopy: 71.3%; navigation: 83.8%, P < 0.001) but not inclination (fluoroscopy: 71.9%; navigation: 76.9%, P = 0.147). The mean LLD was higher with the use of fluoroscopy compared with navigation (5.5 mm, standard deviation: 4.1; 4.6 mm, SD: 3.4, P < 0.005). No difference in dislocation rate was observed ( P = 0.643).
Conclusion: Both fluoroscopy and imageless optical navigation demonstrated accurate acetabular implant positioning during DAA THA. Navigation was more precise and associated with improved acetabular anteversion placement and restoration of LLD. Navigation is an accurate alternative to fluoroscopy with decreased radiation exposure.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.5435/JAAOS-D-23-00790 | DOI Listing |
Comput Assist Surg (Abingdon)
December 2025
Orthopaedic Surgery Program, London Health Sciences Centre, London, ON, Canada.
Achieving optimal implant position and orientation during total knee arthroplasty (TKA) is a pivotal factor in long-term survival. Computer-assisted navigation (CAN) has been recognized as a trusted technology that improves the accuracy and consistency of femoral and tibial bone cuts. Imageless CAN offers advantages over image-based CAN by reducing cost, radiation exposure, and time.
View Article and Find Full Text PDFSensors (Basel)
November 2024
Sydney Musculoskeletal Health, Kolling Institute, Northern Clinical School, Faculty of Medicine and Health, University of Sydney, Camperdown, NSW 2064, Australia.
(1) Background: Imageless computer navigation systems have the potential to improve the accuracy of acetabular cup position in total hip arthroplasty (THA). Popular imageless navigation methods include locating the patient in a three-dimensional space (registration method) while using a baseline to angle the acetabular cup (reference plane). This study aims to compare the accuracy of different methods for determining postoperative acetabular cup positioning in THA via the direct anterior approach.
View Article and Find Full Text PDFJ Am Acad Orthop Surg
March 2024
From the Temerty Faculty of Medicine, University of Toronto, Toronto, Ontario, Canada (Manzo), the Department of Surgery, Division of Orthopaedic Surgery, Temerty Faculty of Medicine, University of Toronto, Toronto, Ontario, Canada, (Lex, Rodriguez-Elizalde, Perlus, and Cayen, and Chang) and the Div
Background: Utilization of the direct anterior approach (DAA) for total hip arthroplasty (THA) has risen in popularity. Accurate implant placement is a critical factor that influences patient outcomes. The goal of this study was to compare the accuracy and precision of fluoroscopy with imageless optical navigation in DAA THA.
View Article and Find Full Text PDFComput Assist Surg (Abingdon)
December 2023
Intellijoint Surgical, Kitchener, Ontario, Canada.
Purpose: To investigate the accuracy of an imageless, optical surgical navigation tool to assist with femoral and tibial bone cuts performed during TKA.
Patients And Methods: Six board-certified orthopedic surgeons participated in a laboratory cadaver investigation, performing femoral and tibial bone cuts with the assistance of a computer navigation tool. Femoral and tibial varus/valgus, tibial slope, femoral flexion, and both femoral and tibial rotation measurements from the device were compared with angular measurements calculated from computed tomography (CT) images of the knees.