A Comparison of Optical Biometers Used in Children for Myopia Control.

Klin Monbl Augenheilkd

Internationale Innovative Ophthalmochirurgie GbR c/o Breyer Kaymak and Klabe Augenchirurgie, Düsseldorf, Germany.

Published: November 2023


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To assess the reproducibility (i.e., inter-device reliability) of the biometers Topcon MYAH, Oculus Myopia Master, and Haag-Streit Lenstar LS900 with the Carl Zeiss IOLMaster 700 and the intra-subject repeatability in myopic children in order to reliably detect axial growth for myopia management.

Methods: Twenty-two myopic children (11.1 ± 2.4 yr) with a spherical equivalent of - 3.53 ± 2.35 D were examined with each of the biometers to assess axial length (AL) and corneal parameters (steepK, flatK, meanK, vectors J0, J45), and16 of these children agreed to a second round of measurements. Reproducibility of the first measurements between the IOLMaster and every other biometer was assessed employing a Bland-Altman approach and paired Student's t-test. Repeatability was assessed as intra-subject standard deviation and was used to estimate the minimum time interval required between two AL measurements to reliably detect axial growth of an eye of at least 0.1 mm/year.

Results: Repeatability for AL measurements was as follows: IOLMaster: 0.05 mm, Myopia Master: 0.06 mm, Myah: 0.06 mm, Lenstar: 0.04 mm; the respective minimal time interval for axial growth assessment in myopia management was estimated as 5.6, 6.6, 6.7, and 5.0 months, respectively. Best reproducibility of the AL measurement was found between IOLMaster and Lenstar [95% Limits of Agreement (LoA) for reproducibility - 0.06 to 0.02]. As regards to the measured means, Lenstar gave measures of AL that were longer than with the IOLMaster by 0.02 mm (p < 0.001). Myopia Master measures of meanK were significantly lower (by 0.21 D with p < 0.001) than the values from the IOLMaster. As for J0, all biometers deviated significantly from IOLMaster measurements (p < 0.05).

Conclusion: Generally good agreement was observed between all the biometers. When assessing myopia progression in children, a time frame of at least 6 months between the AL measurements is advisable in order to reliably determine any deviation from a normal growth pattern.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10651350PMC
http://dx.doi.org/10.1055/a-2117-9335DOI Listing

Publication Analysis

Top Keywords

axial growth
12
myopia master
8
myopic children
8
reliably detect
8
detect axial
8
measurements iolmaster
8
time interval
8
myopia
5
iolmaster
5
comparison optical
4

Similar Publications

Background: Despite population growth, the UAE is experiencing a decline in the utilization of plain radiography (X-ray) due to changes in clinical priorities and advancements in imaging technologies like CT and MRI. This decrease impacts medical education, creating a gap between training and practical application. Ensuring evidence-based justification for radiological procedures is crucial to prevent overutilization and strain of healthcare resources, highlighting the need for updated curricula.

View Article and Find Full Text PDF

Measuring the stress tensor in nitrogen-doped CVD diamond using solid-state quantum sensor.

Sci Technol Adv Mater

August 2025

Research Center for Electronic and Optical Materials, National Institute for Materials Science, Tsukuba, Ibaraki, Japan.

We measured the residual stress tensor in a nitrogen-doped chemical vapor deposition (001) diamond film. The stress tensor was evaluated from the amount of the shift in optically detected magnetic resonance (ODMR) spectra of NV center in the diamond. A confocal microscopy setup was used to observe the spatial variation of the stress tensor in the diamond film.

View Article and Find Full Text PDF

Dome-shaped macula (DSM) is a distinctive anatomical entity characterized by an inward convexity of the macula, initially described in highly myopic eyes within posterior staphyloma, but it is now recognized as occurring across a broader spectrum of refractive conditions, including mild myopia and even emmetropia. Since its initial description in 2008, advances in imaging technologies and longitudinal studies have significantly improved our understanding of DSM. This review analyzed the recent literature, focusing on publications from the last 10 years.

View Article and Find Full Text PDF

Mitochondrial Complex V Deficiency Caused by a Homozygous Splice Variant in ATP5PO.

Am J Med Genet A

September 2025

Division of Clinical and Metabolic Genetics, Department of Pediatrics, Hospital for Sick Children, University of Toronto, Toronto, Ontario, Canada.

Most complex V subunits are nuclear encoded and so far, were not found in association with recognized Mendelian disorders. ATP5PO is a candidate gene for complex V mitochondrial disease. It encodes the oligomycin sensitivity-conferring protein (OSCP), an essential component of the "stalk" region that links the F1 and F0 domains of the ATP synthase complex.

View Article and Find Full Text PDF

Baseline ocular biometrics in children with orthokeratology-induced axial length reduction or elongation.

BMC Ophthalmol

September 2025

Optometry Center, Shaanxi Key Laboratory of Ophthalmology, Xi'an No.1 Hospital (the First Affiliated Hospital of Northwest University), Shaanxi Institute of Ophthalmology, No.30, Fenxiang Alley, South Street, Xi'an, Shaanxi Province, 710002, China.

Purpose: To determine the baseline ocular biometrics in children with axial length reduction after orthokeratology for better myopia control.

Methods: Changes in ocular parameters in 164 subjects who completed one year of Ortho-K treatment were retrospectively reviewed. The subjects were divided into AL reduction (n = 105, ≤ 0 mm/year axial growth) and AL elongation (n = 59, ≥ 0.

View Article and Find Full Text PDF