Background And Purpose: The CATCH (Coil Application Trial in China) trial was designed to assess the safety and efficacy of the Numen Coil Embolization System in the treatment of intracranial aneurysms in comparison with the Axium coil (ev3/Medtronic). Although the endovascular treatment of small (< 5 mm) intracranial aneurysms has been reported with favorable long-term clinical and angiographic outcomes, randomized trials are still lacking. Data for aneurysms smaller than 5 mm were extracted from the CATCH trial.
Materials And Methods: A randomized, prospective, multicenter trial was conducted at ten centers throughout China. Enrolled subjects with small intracranial aneurysms were randomly assigned to receive treatment with the Numen Coil or the Axium coil. The primary outcome was successful aneurysm occlusion at the 6-month follow-up. In contrast, the secondary outcomes included complete aneurysm occlusion, recurrence rate, clinical deterioration, and safety data at the 6-month and 12-month follow-ups.
Results: A total of 124 patients were enrolled in the study. Overall, 58 patients were assigned to the Numen group, and 66 were assigned to the Axium group. At the 6-month follow-up, the successful aneurysm occlusion rate was 93.1% (54/58) in the MicroPort NeuroTech group and 97.0% (64/66) in the Axium group, with a common odds ratio of 0.208 (95% confidence interval, 0.023-1.914; P = 0.184). Complications were comparable between the groups.
Conclusions: Compared with the Aixum coil, the Numen coil is safe and effective in treating small intracranial aneurysms.
Trial Registration: (13/12/2016, NCT02990156).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10276443 | PMC |
http://dx.doi.org/10.1186/s12893-023-02049-9 | DOI Listing |
BMC Surg
June 2023
Department of Neurointervention, Zhengzhou University First Affiliated Hospital, No.1, Jianshe East Road, Erqi District, 450052, Zhengzhou, Henan, China.
Background And Purpose: The CATCH (Coil Application Trial in China) trial was designed to assess the safety and efficacy of the Numen Coil Embolization System in the treatment of intracranial aneurysms in comparison with the Axium coil (ev3/Medtronic). Although the endovascular treatment of small (< 5 mm) intracranial aneurysms has been reported with favorable long-term clinical and angiographic outcomes, randomized trials are still lacking. Data for aneurysms smaller than 5 mm were extracted from the CATCH trial.
View Article and Find Full Text PDFWorld Neurosurg
April 2022
Department of Neurosurgery, The First Affiliated Hospital of Nanchang University, Nanchang City, Jiangxi Province.
Objective: We investigated the safety and efficacy of the Numen coil compared with the Axium coil in the treatment of intracranial aneurysms.
Methods: Because CATCH (Coil Application Trial in China) is a prospective randomized controlled open-label noninferiority trial conducted in 10 centers across China, patients who fulfilled the inclusion and exclusion criteria were randomized 1:1 to either a test group (Numen) or a control group (Axium). The primary outcome was based on successful aneurysm occlusion at 6 months follow-up, whereas secondary outcomes included technical success, the recanalization and retreatment rates, and the rate of serious adverse events (SAEs) at 6 months and 12 months follow-up.