Article Abstract

Background: Telepathology utilizing high-throughput static whole slide image scanners is proposed to address the challenge of limited pathology services in resource-restricted settings. However, the prohibitive equipment costs and sophisticated technologies coupled with large amounts of space to set up the devices make it impractical for use in resource-limited settings. Herein, we aimed to address this challenge by validating a portable whole slide imaging (WSI) device against glass slide microscopy (GSM) using lymph node biopsies from suspected lymphoma cases from Sub-Saharan Africa.

Material And Methods: This was part of a multicenter prospective case-control head-to-head comparison study of liquid biopsy against conventional pathology. For the portable WSI scanner validation, the study pathologists evaluated 105 surgical lymph node specimens initially confirmed by gold-standard pathology between February and December 2021. The tissues were processed according to standard protocols for Hematoxylin and Eosin (H&E) and Immunohistochemistry (IHC) staining by well-trained histotechnicians, then digitalized the H& E and IHC slides at each center. The digital images were anonymized and uploaded to a HIPAA-compliant server by the histotechnicians. Three study pathologists independently accessed and reviewed the images after a 6-week washout. The agreement between diagnoses established on GSM and WSI across the pathologists was described and measured using Cohens' kappa coefficient (κ).

Results: On GSM, 65.5% (n=84) of specimens were lymphoma; 25% were classified as benign, while 9.5% were metastatic. Morphological quality assessment on GSM and WSI established that 79.8% and 53.6% of cases were of high quality, respectively. When diagnoses by GSM were compared to WSI, the overall concordance for various diagnostic categories was 93%, 100%, and 86% for lymphoma, metastases, and benign conditions respectively. The sensitivity and specificity of WSI for the detection of lymphoma were 95.2% and 85.7%, respectively, with an overall inter-observer agreement (κ) of 0.86; 95% CI (0.70-0.95).

Conclusions: We demonstrate that mobile whole slide imaging (WSI) is non-inferior to conventional glass slide microscopy (GSM) for the primary diagnosis of malignant infiltration of lymph node specimens. Our results further provide proof of concept that mobile WSI can be adapted to resource-restricted settings for primary surgical pathology and would significantly improve patient outcomes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9874079PMC
http://dx.doi.org/10.1016/j.jpi.2023.100188DOI Listing

Publication Analysis

Top Keywords

slide imaging
12
lymph node
12
portable slide
8
address challenge
8
resource-restricted settings
8
wsi
8
imaging wsi
8
glass slide
8
slide microscopy
8
microscopy gsm
8

Similar Publications

Artificial intelligence (AI) is a technique or tool to simulate or emulate human "intelligence." Precision medicine or precision histology refers to the subpopulation-tailored diagnosis, therapeutics, and management of diseases with its sociocultural, behavioral, genomic, transcriptomic, and pharmaco-omic implications. The modern decade experiences a quantum leap in AI-based models in various aspects of daily routines including practice of precision medicine and histology.

View Article and Find Full Text PDF

Diffuse large B-cell lymphoma is the most common type of non-Hodgkin lymphoma (NHL) in humans, accounting for about 30-40% of NHL cases worldwide. Canine diffuse large B-cell lymphoma (cDLBCL) is the most common lymphoma subtype in dogs and demonstrates an aggressive biologic behaviour. For tissue biopsies, current confirmatory diagnostic approaches for enlarged lymph nodes rely on expert histopathological assessment, which is time-consuming and requires specialist expertise.

View Article and Find Full Text PDF

Single camera estimation of microswimmer depth with a convolutional network.

J R Soc Interface

September 2025

Institute of Intelligent Systems and Robotics, Sorbonne Université, Paris, Île-de-France, France.

A number of techniques have been developed to measure the three-dimensional trajectories of protists, which require special experimental set-ups, such as a pair of orthogonal cameras. On the other hand, machine learning techniques have been used to estimate the vertical position of spherical particles from the defocus pattern, but they require the acquisition of a labelled dataset with finely spaced vertical positions. Here, we describe a simple way to make a dataset of images labelled with vertical position from a single 5 min movie, based on a tilted slide set-up.

View Article and Find Full Text PDF

Background: With the increasing incidence of skin cancer, the workload for pathologists has surged. The diagnosis of skin samples, especially for complex lesions such as malignant melanomas and melanocytic lesions, has shown higher diagnostic variability compared to other organ samples. Consequently, artificial intelligence (AI)-based diagnostic assistance programs are increasingly needed to support dermatopathologists in achieving more consistent diagnoses.

View Article and Find Full Text PDF

Aims: gene is amplified in 15%-20% of invasive breast cancers (IBCs), serving as critical prognostic and predictive marker. -targeted therapies have improved outcomes for -positive patients, highlighting the importance of accurate assessment. Immunohistochemistry is commonly used for screening overexpression, with equivocal cases reflex tested using in situ hybridisation (ISH) methods like fluorescence (FISH) or dual-colour dual ISH (D-DISH).

View Article and Find Full Text PDF