Article Abstract

Novel traceable analytical methods and reference gas standards were developed for the detection of trace-level ammonia in biogas and biomethane. This work focused on an ammonia amount fraction at an upper limit level of 10 mg m (corresponding to approximately 14 μmol mol) specified in EN 16723-1:2016. The application of spectroscopic analytical methods, such as Fourier transform infrared spectroscopy, cavity ring-down spectroscopy, and optical feedback cavity-enhanced absorption spectroscopy, was investigated. These techniques all exhibited the necessary ammonia sensitivity at the required 14 μmol mol amount fraction. A 29-month stability study of reference gas mixtures of 10 μmol mol ammonia in methane and synthetic biogas is also reported.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.analchem.2c01951DOI Listing

Publication Analysis

Top Keywords

μmol mol
12
fourier transform
8
transform infrared
8
cavity ring-down
8
ring-down spectroscopy
8
spectroscopy optical
8
optical feedback
8
feedback cavity-enhanced
8
cavity-enhanced absorption
8
absorption spectroscopy
8

Similar Publications

Humidity-responsive materials have attracted increasing attention for their potential use in various applications, e.g., sensors, soft robotics, and human-machine interfaces.

View Article and Find Full Text PDF

The concentration of various phospholipids (PLs) and sphingomyelin in platelets and the amount of [14C] arachidonic acid ([14C]-AA) esterified in phosphatidylinositol (PI), phosphatidylserine (PS), phosphatidylethanolamine (PE), and phosphatidylcholine (PC) were measured. The platelet-rich plasmas from unmedicated patients with psychiatric disorders and healthy controls were incubated for 30 min with 1 microM [14C]-AA. Platelets from patients with a schizoaffective disorder according to RDC criteria, a schizophreniform disorder (DSM III criteria) or an atypical phasic psychosis according to FC criteria contained twice as much PI and had significantly increased concentrations of PC as compared to controls (p less than 0.

View Article and Find Full Text PDF