Article Abstract

Objective: To evaluate the agreement between 3 point-of-care (POC) devices and a reference laboratory for measuring β-hydroxybutyrate (β-HB) concentration in African penguin (Spheniscus demersus) whole blood (WB) and plasma samples and the precision of each POC device for measuring β-HB concentration in plasma samples.

Animals: 48 healthy African penguins.

Procedures: Blood was obtained from the right jugular vein of each penguin, and β-HB concentration was measured on each POC device using fresh WB and heparinized plasma and at the reference laboratory using plasma. β-HB concentration was measured in plasma on each POC device.

Results: All devices overestimated serum β-HB concentrations on average by 0.46 mM relative to the reference laboratory. WB samples had less error than plasma for meters A and C. Meter A had the lowest total error observed (26.4%) and the lowest mean difference (0.19 mmol/L) relative to the reference laboratory. Controlling for other factors, the magnitude of disagreement was not affected by sex, age, packed cell volume, or serum total solids concentration.

Clinical Relevance: WB, not plasma, should be used for measurement of β-HB concentration on the POC meters tested. Meter A showed good correlation with the reference laboratory for WB. The use of POC devices for the measurement of β-HB concentration may be acceptable when laboratory analyzers are not available. Further research is needed for clinical application and the diagnostic value of POC meters compared with reference laboratories.

Download full-text PDF

Source
http://dx.doi.org/10.2460/ajvr.21.11.0192DOI Listing

Publication Analysis

Top Keywords

β-hb concentration
24
reference laboratory
20
healthy african
8
spheniscus demersus
8
poc devices
8
poc device
8
concentration measured
8
relative reference
8
measurement β-hb
8
poc meters
8

Similar Publications

Selenium is an essential trace element in many organisms but becomes toxic at elevated concentrations. At moderately increased, non-lethal levels, selenite triggers both selenium utilization and stress responses in microorganisms. However, the thresholds of such responses in archaea remain poorly understood.

View Article and Find Full Text PDF

Ambient Air Pollution and the Severity of Alzheimer Disease Neuropathology.

JAMA Neurol

September 2025

Translational Neuropathology Research Laboratory, Department of Pathology and Laboratory Medicine, Perelman School of Medicine at the University of Pennsylvania, Philadelphia.

Importance: Exposure to fine particulate matter air pollution (PM2.5) may increase risk for dementia. It is unknown whether this association is mediated by dementia-related neuropathologic change found at autopsy.

View Article and Find Full Text PDF

Simulations in three dimensions and time provide guidance on implantable, electroenzymatic glutamate sensor design; relative placement in planar sensor arrays; feasibility of sensing synaptic release events; and interpretation of sensor data. Electroenzymatic sensors based on the immobilization of oxidases on microelectrodes have proven valuable for the monitoring of neurotransmitter signaling in deep brain structures; however, the complex extracellular milieu featuring slow diffusive mass transport makes rational sensor design and data interpretation challenging. Simulations show that miniaturization of the disk-shaped device size below a radius of ∼25 μm improves sensitivity, spatial resolution, and the accuracy of glutamate concentration measurements based on calibration factors determined .

View Article and Find Full Text PDF

In pig production, weaning is a critical period where piglets face several environmental stressors. This transition leads to a significant growth reduction and can result in digestive disorders, including diarrhea. To formulate a feed that meets zinc (Zn) and copper (Cu) requirements during the weaning period while minimizing their release into the environment, it became evident that a more bioavailable micro-mineral supplement is necessary.

View Article and Find Full Text PDF

Background And Aims: Inflammatory bowel disease (IBD), including Crohn's disease (CD) and ulcerative colitis (UC), remain heterogeneous disorders with variable response to biologics. Post-operative recurrence in CD is common despite surgery and prophylactic biotherapies. Understanding the inflammatory mediators associated with recurrence and treatment response could pave the way for personalized strategies.

View Article and Find Full Text PDF