Article Abstract

Background: Endovascular revascularization (ER) via percutaneous transluminal angioplasty (PTA) and stenting are viable options for revascularization in below-the-knee (BTK) peripheral arterial disease. Two-dimensional angiography has been the standard of practice for estimating vessel size and selecting treatment devices during ER. However, in other vascular territories, intravascular ultrasound (IVUS) offers better visualization of the lumen dimensions.

Purpose: To compare angiographic and intravascular ultrasound reference vessel (lumen) measurements in below-the-knee peripheral artery interventions.

Methods: Twenty consecutive patients were enrolled in the BTK Calibration study from 2 sites in the United States and Australia. Patients with at least one diseased segment in a native infra-popliteal artery (below-the-knee) and a clinical indication for endovascular therapy (EVT) were included with no limitations with regard to vessel diameter or lesion length. Digital subtraction angiography and IVUS imaging were collected pre- and post-PTA and images were sent to an independent core lab for standardized quantitative analysis of the normal-looking reference vessel dimensions when available. The results were presented as least square means with 95% confidence intervals and a p-value of <0.05 was considered significant.

Results: The overall (N = 19) mean reference vessel diameter for QVA was 2.98 ± 1.24 mm vs. 3.47 ± 0.72 mm for IVUS (mean difference was -0.50 mm, (95% CI: -0.80, -0.20; p = 0.14). As expected, in the proximal segments (N = 12), the mean reference vessel diameters were larger: for QVA, it was 3.17 ± 1.34 mm vs. 3.55 ± 0.76 mm in IVUS, (mean difference was -0.38 mm, (95% CI: -0.79, 0.03; p = 0.40); while in the distal segments (N = 7), mean reference vessel diameters were smaller: for QVA, it was 2.64 ± 1.06 mm vs. 3.33 ± 0.67 mm in IVUS, (mean difference was -0.69 mm, (95% CI: -1.04, 0.34; p = 0.17). We observed a greater degree of acute gain in cases where the treatment balloon size correlated with the IVUS measured reference size.

Conclusion: Angiography underestimates infrapopliteal reference vessel lumen size even when quantitatively assessed. Adjunctive IVUS imaging use in guiding BTK procedures could help ensure adequate sizing and possibly impact immediate post-procedure indices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.carrev.2021.09.004DOI Listing

Publication Analysis

Top Keywords

intravascular ultrasound
12
angiographic intravascular
8
calibration study
8
reference vessel
8
vessel
5
comparison angiographic
4
ultrasound vessel
4
vessel measurements
4
measurements infra-popliteal
4
infra-popliteal endovascular
4

Similar Publications

Prospective Evaluation of Thrombotic Complications After Internal Jugular Vein Cannulation for External Bypass.

J Cardiothorac Vasc Anesth

August 2025

Goethe-University Frankfurt, University Hospital Frankfurt, Department of Anaesthesiology, Intensive Care Medicine and Pain Therapy, Frankfurt, Germany; Department of Anesthesiology and Intensive Care Medicine, Rheinische Friedrich-Wilhelms-University, University Hospital Bonn, Bonn, Germany.

Objectives: To determine the incidence and subsequent complications of internal jugular vein (IJV) thrombosis after cannulation performed during cardiopulmonary bypass (CPB) to ensure adequate venous drainage during minimally invasive cardiac surgery.

Design: Single-center observational trial SETTINGS: Intensive care postoperative monitoring of cardiac surgery patients and diagnosis of IJV thrombi at a university tertiary hospital during the 13-month study period from December 1, 2022, to January 11, 2024.

Participants: 44 patients undergoing catheterization of the IJV for total CPB.

View Article and Find Full Text PDF

Sonographic examination of major vessels can be a valuable bedside tool for perioperative hemodynamic assessment. In the present review, we present the anatomic and physiological aspects of internal jugular vein ultrasonography, its utility in assessing central venous pressure, intravascular volume status, fluid responsiveness, and its predictive value regarding post-spinal anesthesia hypotension. The existing literature is primarily comprised of small, observational studies with great heterogeneity in their methodology and shortcomings in data development and analysis, rendering the generalization of their results difficult to interpret for daily clinical practice.

View Article and Find Full Text PDF

Background: An inward force is experienced by the guide catheter during device retrieval resulting in potential risk of deep engagement into the ostio-proximal coronary segment. This undesired movement can result in coronary injury. There is no systematic data or reports of techniques to prevent such inadvertent guide movement during difficult retrieval of devices.

View Article and Find Full Text PDF

Background: Cardiac issues following radiotherapy are increasingly prevalent among patients with thoracic cancer and coronary disease. However, the mechanisms underlying radiotherapy-induced plaque instability and changes in plaque characteristics on imaging remain unclear. This study used single-cell RNA sequencing to identify key features of vulnerable plaques following radiotherapy.

View Article and Find Full Text PDF

A 33-year-old male developed pulmonary vein (PV) stenosis 3 months after PV isolation for atrial fibrillation. Stents were implanted in the left superior and inferior PVs, but 2 years later, in-stent restenosis occurred. Intravascular ultrasound and nonobstructive general angioscopy (NOGA) revealed severe ostial stenosis and neointimal hyperplasia.

View Article and Find Full Text PDF