Purpose: We evaluated the effect of changing the scan mode of the Elekta X-ray volume imaging cone beam computed tomography (CBCT) on the accuracy of dose calculation, which may be affected by computed tomography (CT) value errors in three dimensions.
Methods: We used the electron density phantom and measured the CT values in three dimensions. CT values were compared with planning computed tomography (pCT) values for various materials. The evaluated scan modes were for head and neck (S-scan), chest (M-scan), and pelvis (L-scan) with various collimators and filter systems. To evaluate the effects of the CT value error of the CBCT on dose error, Monte Carlo calculations of dosimetry were performed using pCT and CBCT images.
Results: The L-scan had a CT value error of approximately 800 HU at the isocenter compared with the pCT. Furthermore, inhomogeneity in the longitudinal CT value profile was observed in the bone material. The dose error for ±100 HU difference in CT values for the S-scan and M-scan was within ±2%. The center of the L-scan had a CT error of approximately 800 HU and a dose error of approximately 6%. The dose error of the L-scan occurred in the beam path in the case of both single field and two parallel opposed fields, and the maximum error occurred at the center of the phantom in the case of both the 4-field box and single-arc techniques.
Conclusions: We demonstrated the three-dimensional CT value characteristics of the CBCT by evaluating the CT value error obtained under various imaging conditions. It was found that the L-scan is considerably affected by not having a unique bowtie filter, and the S-scan without the bowtie filter causes CT value errors in the longitudinal direction. Moreover, the CBCT dose errors for the 4-field box and single-arc irradiation techniques converge to the isocenter.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8425939 | PMC |
http://dx.doi.org/10.1002/acm2.13384 | DOI Listing |
Nutr Clin Pract
September 2025
Department of Clinical Nutrition, Hubei Cancer Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China.
Objective: The cachexia index (CXI) demonstrates potential as both a diagnostic tool for cachexia and a prognostic tool for survival in cancer. However, CXI's predictive value has not been verified in cervical cancer. The purpose of this study is to investigate the prognostic value of the CXI in patients with cervical cancer treated with radiotherapy.
View Article and Find Full Text PDFClin Oral Investig
September 2025
Department of Stomatology, Shengli Oilfield Central Hospital, No. 31, Jinan Road, Dongying, 257034, China.
Objective: Progesterone (PG) and its target, progesterone receptor (PGR), are important regulators in inflammatory diseases. This study aimed to investigate the specific role of PG in periodontitis and to elucidate the underlying mechanisms involving PGR.
Methods: Women with periodontitis, including 250 with PG deficiency, 250 with PG supplementation, and 245 controls (normal PG) were enrolled.
J Imaging Inform Med
September 2025
Department of Biomedical Engineering, Gachon University, Seongnam-Si 13120, Gyeonggi-Do, Republic of Korea.
To develop and validate a deep-learning-based algorithm for automatic identification of anatomical landmarks and calculating femoral and tibial version angles (FTT angles) on lower-extremity CT scans. In this IRB-approved, retrospective study, lower-extremity CT scans from 270 adult patients (median age, 69 years; female to male ratio, 235:35) were analyzed. CT data were preprocessed using contrast-limited adaptive histogram equalization and RGB superposition to enhance tissue boundary distinction.
View Article and Find Full Text PDFJ Orthop Sci
September 2025
Department of Orthopaedic Surgery, NHO Saga Hospital, 1-20-1 Hinode, Saga 849-0923, Japan.
Background: Hounsfield units (HU) on computed tomography (CT) are strongly correlated with bone mineral density (BMD) and may aid in osteoporosis screening. However, there is no standardized method for assessing bone density in displaced femoral head fractures. This study aimed to measure HU values in the femoral head using preoperative post-fracture CT images of patients with intertrochanteric femoral fractures and investigate whether it correlated with BMD measured by dual-energy X-ray absorptiometry (DXA).
View Article and Find Full Text PDFJ Cardiovasc Comput Tomogr
September 2025
Depertament of Cardiology, Hitit University, Faculty of Medicine, Çorum, Turkey.