Article Abstract

Background: Long-lasting reconstruction after extensive resection involving peri-knee metaphysis is a challenging problem in orthopedic oncology. Various reconstruction methods have been proposed, but they are characterized by a high complication rate. The purposes of this study were to (1) assess osseointegration at the bone implant interface and correlated incidence of aseptic loosening; (2) identify complications including infection, endoprosthesis fracture, periprosthetic fracture, leg length discrepancy, and wound healing problem in this case series; and (3) evaluate the short-term function of the patient who received this personalized reconstruction system.

Methods: Between September 2016 and June 2018, our center treated 15 patients with malignancies arising in the femur or tibia shaft using endoprosthesis with a 3D-printed custom-made stem. Osseointegration and aseptic loosening were assessed with digital tomosynthesis. Complications were recorded by reviewing the patients' records. The function was evaluated with the 1993 version of the Musculoskeletal Tumor Society (MSTS-93) score at a median of 42 (range, 34 to 54) months after reconstruction.

Results: One patient who experienced early aseptic loosening was managed with immobilization and bisphosphonates infusion. All implants were well osseointegrated at the final follow-up examination. There are two periprosthetic fractures intraoperatively. The wire was applied to assist fixation, and the fracture healed at the latest follow-up. Two patients experienced significant leg length discrepancies. The median MSTS-93 score was 26 (range, 23 to 30).

Conclusions: A 3D-printed custom-made ultra-short stem with a porous structure provides acceptable early outcomes in patients who received peri-knee metaphyseal reconstruction. With detailed preoperative design and precise intraoperative techniques, the reasonable initial stability benefits osseointegration to osteoconductive porous titanium, and therefore ensures short- and possibly long-term durability. Personalized adaptive endoprosthesis, careful intraoperative operation, and strict follow-up management enable effective prevention and treatment of complications. The functional results in our series were acceptable thanks to reliable fixation in the bone-endoprosthesis interface and an individualized rehabilitation program. These positive results indicate this device series can be a feasible alternative for critical bone defect reconstruction. Nevertheless, longer follow-up is required to determine whether this technique is superior to other forms of fixation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8349501PMC
http://dx.doi.org/10.1186/s12957-021-02355-7DOI Listing

Publication Analysis

Top Keywords

aseptic loosening
12
custom-made ultra-short
8
ultra-short stem
8
stem porous
8
porous structure
8
structure acceptable
8
peri-knee metaphysis
8
leg length
8
3d-printed custom-made
8
msts-93 score
8

Similar Publications

Introduction: the progressive increase in the volume of primary knee arthroplasty surgery is leading in parallel to an increase in revision surgeries, and this trend is expected to continue in the coming decades.

Material And Methods: a retrospective observational study with 203 interventions in 201 patients with knee arthroplasty revision surgery at our center between 2018-2022, including the variables age, sex, body mass index, primary implant, time until revision surgery, mechanical alignment of the primary implant, cause of revision, revision implant, and need for secondary revision.

Results: the average age of the patients was 73.

View Article and Find Full Text PDF

Bilateral Osteonecrosis of the Femoral Head and Proximal Femur After Corticosteroid Injections: A Case Report.

J Am Acad Orthop Surg Glob Res Rev

September 2025

From the Mayo Clinic Alix School of Medicine, Scottsdale, AZ (Ms. Hiredesai and Mr. Holle), and the Department of Orthopedic Surgery, Mayo Clinic, Phoenix, AZ (Dr. Van Schuyver, Dr. Deckey, Dr. Probst, and Dr. Spangehl).

Atraumatic bilateral osteonecrosis of the femoral head (ONFH) is a rare phenomenon whose etiology is not fully understood. In this report, we describe the case of a 75-year-old female patient who developed rapidly onset bilateral ONFH after intra-articular corticosteroid injections. She was treated with staged bilateral total hip arthroplasty.

View Article and Find Full Text PDF

Soft-Tissue Sources of Extra-articular Pain Following Total Hip Arthroplasty: A Comprehensive Review.

J Am Acad Orthop Surg Glob Res Rev

September 2025

From the American Hip Institute Research Foundation (Dr. Quesada-Jimenez, Dr. Kahana-Rojkind, and Dr. Domb), and the American Hip Institute, Chicago, IL (Dr. Domb).

Hip pain after a total hip arthroplasty is a prevalent condition. Once aseptic loosening and infection have been ruled out, the possible entities are vast. Accurate diagnosis in this patient population is challenging because they might present in different stages of their recovery process and the potential overlap of some conditions.

View Article and Find Full Text PDF

Robotic-Assisted Versus Manual Total Hip Arthroplasty Performed for Hip Dysplasia.

J Am Acad Orthop Surg

September 2025

From the Department of Orthopedic Surgery, Hospital for Special Surgery, New York, NY (Neitzke, O'Donnell, Buchalter, Chandi, Westrich, and Gausden), the Department of Orthopedic Surgery, University of Wisconsin-Madison, Madison, WI (O'Donnell), and Somers Orthopaedic Surgery & Sports Medicine Group

Introduction: Developmental dysplasia of the hip (DDH) poses challenges for component positioning during total hip arthroplasty (THA) secondary to abnormal bone morphology, soft-tissue contractures, and hip center migration. The objective of this study was to evaluate the radiographic and clinical outcomes of THA for DDH performed with robotic assistance versus manual (M) technique.

Methods: A retrospective review identified 115 patients with Crowe II to IV dysplasia undergoing primary THA at a single institution from 2016 to 2022.

View Article and Find Full Text PDF

Objective: Durable and biologically integrated fixation is critical for long-term implant survival in patients with primary bone tumors. However, limited evidence exists regarding the long-term outcomes of uncemented stem designs in this population. Specifically, we investigated: (1) the long-term patient and implant survivorship rates; (2) the influence of factors such as resection length and patient age on implant survival; and (3) the incidence and types of complications, particularly those requiring implant removal or revision.

View Article and Find Full Text PDF