Article Abstract

Purpose: To compare patient preferences for eyeglasses prescribed using a low-cost, portable wavefront autorefractor versus standard subjective refraction (SR).

Design: Randomized, cross-over clinical trial.

Participants: Patients aged 18 to 40 years presenting with refractive errors (REs) to a tertiary eye hospital in Southern India.

Methods: Participants underwent SR followed by autorefraction (AR) using the monocular version of the QuickSee device (PlenOptika Inc). An independent optician, masked to the refraction approach, prepared eyeglasses based on each refraction approach. Participants (masked to refraction source) were randomly assigned to use SR- or AR-based eyeglasses first, followed by the other pair, for 1 week each. At the end of each week, participants had their vision checked and were interviewed about their experience with the eyeglasses.

Main Outcome Measures: Patients preferring eyeglasses were chosen using AR and SR.

Results: The 400 participants enrolled between March 26, 2018, and August 2, 2019, had a mean (standard deviation) age of 28.4 (6.6) years, and 68.8% were women. There was a strong correlation between spherical equivalents using SR and AR (r = 0.97, P < 0.001) with a mean difference of -0.07 diopters (D) (95% limits of agreement [LoA], -0.68 to 0.83). Of the 301 patients (75.2%) who completed both follow-up visits, 50.5% (n = 152) and 49.5% (n = 149) preferred glasses prescribed using SR and AR, respectively (95% CI, 45.7-56.3; P = 0.86). There were no differences in demographic or vision characteristics between participants with different preferences (P > 0.05 for all).

Conclusions: We observed a strong agreement between the prescriptions from SR and AR, and eyeglasses prescribed using SR and AR were equally preferred by patients. Wider use of prescribing based on AR alone in resource-limited settings is supported by these findings.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ophtha.2021.05.030DOI Listing

Publication Analysis

Top Keywords

low-cost portable
8
eyeglasses prescribed
8
masked refraction
8
refraction approach
8
eyeglasses
5
participants
5
investigation accuracy
4
accuracy low-cost
4
portable autorefractor
4
autorefractor provide
4

Similar Publications

Alpine streams represent some of the most challenging yet ecologically valuable freshwater environments to study, due to their remoteness, fast flows and extreme climatic conditions. Traditional fish survey methods are often impractical or invasive in these habitats. This study presents a lightweight, low-cost, T-shaped remote underwater video (RUV) system optimized for fish monitoring in small, high-altitude streams of the European Alps.

View Article and Find Full Text PDF

Construction of an Ag-functionalized structural color hydrogel sensor for colorimetric detection of glutathione.

Mikrochim Acta

September 2025

Hunan Key Laboratory of Typical Environmental Pollution and Health Hazards, School of Public Health, Hengyang Medical School, University of South China, Hengyang, 421001, China.

An Ag-functionalized structural color hydrogel (Ag-SCH) sensor is constructed for colorimetric detection of glutathione (GSH). The hydrogel is prepared by using the coordination of Ag and 1-vinylimidazole (1-VI) as cross-linking network. GSH acts as a competitive ligand to break the coordination between Ag and 1-VI, leading to the expansion and structural color change of the hydrogel.

View Article and Find Full Text PDF

Transposing intensive care innovation from modern warfare to other resource-limited settings.

Eur J Trauma Emerg Surg

September 2025

French Military Medical Service Academy - École du Val-de-Grâce, Paris, France.

Background: Delivering intensive care in conflict zones and other resource-limited settings presents unique clinical, logistical, and ethical challenges. These contexts, characterized by disrupted infrastructure, limited personnel, and prolonged field care, require adapted strategies to ensure critical care delivery under resource-limited settings.

Objective: This scoping review aims to identify and characterize medical innovations developed or implemented in recent conflicts that may be relevant and transposable to intensive care units operating in other resource-limited settings.

View Article and Find Full Text PDF

An electrochemiluminescence device powered by streaming potential for the detection of amines in flowing solution.

Nat Commun

September 2025

Department of Chemical Science and Engineering, School of Materials and Chemical Technology, Institute of Science Tokyo, Nagatsuta-cho, Midori-ku, Yokohama, Japan.

The research and implementation of portable and low-cost analytical devices that possess high reproducibility and ease of operation is still a challenging task, and a growing field of importance, within the analytical research. Herein, we report the concept, design and optimization of a microfluidic device based on electrochemiluminescence (ECL) detection that can be potentially operated without electricity for analytical purposes. The device functions exploiting the concept of streaming potential-driven bipolar electrochemistry, where a potential difference, generated from the flow of an electrolyte through a microchannel under the influence of a pressure gradient, is the driving force for redox reactions.

View Article and Find Full Text PDF

This study explores deep feature representations from photoplethysmography (PPG) signals for coronary artery disease (CAD) identification in 80 participants (40 with CAD). Finger PPG signals were processed using multilayer perceptron (MLP) and convolutional neural network (CNN) autoencoders, with performance assessed via 5-fold cross-validation. The CNN autoencoder model achieved the best results (recall 96.

View Article and Find Full Text PDF