We describe a multicenter study using the European Spine Phantom (ESP) to compare the accuracy, linearity and precision of QCT measurements of spine vBMD between different brands of scanner, different models of the same brand and identical units of the same model. Ten scans of the same ESP with repositioning were performed on forty CT scanners from five manufacturers in different hospitals across China, all calibrated with the Mindways QCT system. The three ESP vertebral bodies simulating low (L1), medium (L2) and high (L3) vBMD and their average (L1-3 vBMD) were compared with phantom values. Linearity was assessed using the standard error of the estimate derived from linear regression. Precision errors were expressed as the standard deviation of the ten measurements on each scanner. Median (IQR) vBMD over all forty CT scanners compared with phantom values were: L1: 52.2 (49.9-56.4) vs 51.0; L2: 104.4 (101.2-108.6) vs 102.2; L3: 201.4 (195.0-204.9) vs 200.4; L1-3: 119.3 (116.6-123.2) vs 117.9 mg/cm. Statistically significant differences in L1-3 vBMD were found between different brands (p= 0.005) and between different models of the same brand and identical units of the same model (both p< 0.001). Cross-calibration using linear regression gave a good fit for all forty systems with a median standard error of the estimate of 1.7 mg/cm. The median precision error for L1-3 vBMD was 0.61 mg/cm. Statistically significant differences in spine vBMD measurements between different scanners reinforce the importance of cross-calibration in multi-center studies. Cross-calibration can be reliably performed using linear regression equations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jocd.2021.02.004DOI Listing

Publication Analysis

Top Keywords

l1-3 vbmd
12
linear regression
12
accuracy linearity
8
linearity precision
8
vbmd
8
brands scanner
8
spine vbmd
8
vbmd brands
8
models brand
8
brand identical
8

Similar Publications

Single-energy QCT (SEQCT) scans to measure volumetric BMD (vBMD) are susceptible to errors caused by variability in the amount of marrow adipose tissue (MAT). We developed a three-material model that uses dual-layer spectral CT (DLCT) technology to measure bone matrix (BM), yellow marrow (YM), and red marrow (RM) and compared the results with measurements of proton density fat fraction (PDFF) by MRI and vBMD by SEQCT. Hounsfield units (HU) were measured in the L1-3 vertebrae on 50 and 150 keV mono-energy images in a training set of 100 Chinese adults.

View Article and Find Full Text PDF

Objective: To investigate the changes in volumetric bone mineral density (vBMD) assessed by quantitative computed tomography (QCT) in chronic kidney disease (CKD) patients on maintenance dialysis.

Study Design: Descriptive study. Place and Duration of the Study: Department of Radiology, The Second Affiliated Hospital of Anhui Medical University, Hefei, China, from March to July 2022.

View Article and Find Full Text PDF

Purpose: This study evaluated the prevalence of vertebral fractures (VF) in middle-aged and elderly Chinese men and women and explored the differences in lumbar spine volumetric bone mineral density (vBMD) derived from quantitative CT (QCT) between those with a grade 1 vertebral fracture and non-fractured individuals.

Materials And Methods: 3,457 participants were enrolled in the China Action on Spine and Hip Status (CASH) study and had upper abdominal CT examinations. Vertebral fractures were identified by Genant's semi-quantitative method from lateral CT scout views or CT sagittal views.

View Article and Find Full Text PDF

We describe a multicenter study using the European Spine Phantom (ESP) to compare the accuracy, linearity and precision of QCT measurements of spine vBMD between different brands of scanner, different models of the same brand and identical units of the same model. Ten scans of the same ESP with repositioning were performed on forty CT scanners from five manufacturers in different hospitals across China, all calibrated with the Mindways QCT system. The three ESP vertebral bodies simulating low (L1), medium (L2) and high (L3) vBMD and their average (L1-3 vBMD) were compared with phantom values.

View Article and Find Full Text PDF