Injection pressure monitoring during peripheral nerve blocks: from bench to operating theatre.

Anaesth Crit Care Pain Med

Department of Anaesthesiology and Critical Care Medicine, Lapeyronie University Hospital, 34295 Montpellier Cedex 5, France; Inserm Unit 1051 Montpellier NeuroSciences Institute, Montpellier University, 34295 Montpellier Cedex 5, France. Electronic address:

Published: October 2020


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Nerve damage can occur after ultrasound-guided peripheral nerve block (PNB). Injection pressure monitoring could improve the safety of PNB. The aim was to analyse parameters affecting pressure measurements during PNB.

Methods: The flow characteristics of needles connected to a pressure-sensing device were evaluated. Needles were placed under ultrasound guidance extra or epineurally in nerves/plexus of fresh cadavers. Using three flow rates, 4 mL of saline was injected and plateau pressure was measured. Finally, orthopaedic surgery patients receiving PNB were enrolled for an observational real-time pressure monitoring study. During PNB, periods with pressure > 50 mmHg were noted (high pressure ≥ 750 mmHg). A blinded investigator recorded injection pressure curves and peak pressure.

Results: The needle diameter influenced the injection pressure (β = 66.8; P <  0.0001). Non-echogenic needles increased the injection pressure (β = 82; P =  0.0009) compared with echogenic needles. Cadaver injection pressure was higher for intraneural (255 [122.5-555] mmHg) versus extraneural needle tip location (90 [50-158] mmHg; P < 0001); for high flow (9.6 mL/min; 470 [265; 900] mmHg) versus low flow (1.2 mL/min; 120 [71-250] mmHg) (P <  0.001) and for cervical roots (900 mmHg, intraneurally) compared with nerves (300 mmHg, intraneurally). In 37 patients and 61 procedures, there were 7 [1-18] peaks of injection pressure per procedure. Pressure was noted > 750 mmHg during 13.80% of the procedural time.

Conclusions: Needle diameter, needle tip location, type of nerve/plexus, flow rates, and the anaesthetist can have a significant effect on injection pressure values and monitoring.

Trial Registration: ClinicalTrials.gov ID: NCT03430453.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.accpm.2020.03.022DOI Listing

Publication Analysis

Top Keywords

injection pressure
16
pressure monitoring
12
peripheral nerve
8
pressure
8
injection
4
monitoring peripheral
4
nerve blocks
4
blocks bench
4
bench operating
4
operating theatre
4

Similar Publications

Incision capsular bag herniation during subluxated lens surgery: mechanisms, management, and prevention.

Front Med (Lausanne)

August 2025

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.

Purpose: To investigate the mechanism, intraoperative characteristics, management, and prevention of incision capsular bag herniation (ICBH), a previously unreported complication during cataract surgery in eyes with lens subluxation.

Methods: A retrospective observational case series was conducted on five male patients who developed ICBH during phacoemulsification with intraocular lens (IOL) implantation between January 2019 and December 2024. Among 867 subluxated-lens surgeries performed during this period, the estimated incidence of ICBH was 0.

View Article and Find Full Text PDF

A 12-month-old female presented with a friable, hemorrhagic papule on the right lateral inferior eyelid, clinically consistent with a pyogenic granuloma. Although pyogenic granulomas are known to bleed, the extent of hemorrhage in this case was unusual and occurred after minor trauma. The patient's father, an emergency medicine resident, injected lidocaine with epinephrine and applied pressure at home to control the bleeding.

View Article and Find Full Text PDF

A numerical investigation of the kinematic and fluid dynamic behaviour of an intramuscular autoinjector designed for optimising injection efficiency.

Med Eng Phys

October 2025

Department of Mechanical Engineering, University of Cape Town, 7701, South Africa; Centre for Research in Computational and Applied Mechanics (CERECAM), University of Cape Town, 7701, South Africa.

The usability and versatility of autoinjectors in managing chronic and autoimmune diseases have made them increasingly attractive in medicine. However, investigations into autoinjector designs require an understanding of the kinematic properties and fluid behaviour during injection. To optimise injection efficiency, this study develops a mathematical and computational fluid dynamics (CFD) model of an IM autoinjector by investigating the effects of viscosity, needle length, needle diameter, and medication volume on the injection process.

View Article and Find Full Text PDF

Amorphous silicon resistors enable smaller pixels in photovoltaic retinal prosthesis.

J Neural Eng

September 2025

Hansen Experimental Physics Laboratory, Stanford University, 452 Lomita Mall, Stanford, California, 94305, UNITED STATES.

Clinical trials of the photovoltaic subretinal prosthesis PRIMA demonstrated feasibility of prosthetic central vision with resolution matching its 100 μm pixel width. To improve prosthetic acuity further, pixel size should be decreased. However, there are multiple challenges, one of which is related to accommodating a compact shunt resistor within each pixel that discharges the electrodes between stimulation pulses and helps increase the contrast of the electric field pattern.

View Article and Find Full Text PDF

Introduction: Distraction methods such as virtual reality and cold vibration devices (Buzzy) are recommended during vascular access. Few studies focused on distraction during intramuscular injection.

Methods: This study evaluated the effect of distraction methods on procedure-related pain, fear, and anxiety during the intramuscular injection in children aged 5 to 12 years in the pediatric emergency department.

View Article and Find Full Text PDF