Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Flow diverters are a breakthrough treatment for large and giant intracranial aneurysms but carry a risk of periprocedural death or major stroke. Pipeline Flex is a second-generation device that is thought to have lower complication rates because of improvements in the delivery system as well as increased operator experience. Our objective was to analyze the risk of periprocedural death or major complications using Pipeline Flex for unruptured intracranial aneurysms.

Methods: A systematic search of three databases was performed for studies of ≥10 treatments using Pipeline Flex for unruptured intracranial aneurysms (2014-2019) using PRISMA guidelines. Random effects meta-analysis was used to pool the rates of periprocedural (<30 days) death, major ischemic stroke, symptomatic intracranial hemorrhage, and minor stroke/transient ischemic attack.

Results: We included eight studies reporting 901 treatments in 879 patients. Periprocedural mortality (<30 days) was 0.8% (5/901; 95% CI 0.4% to 1.5%; I=0%). Rate of major complications (death, major ischemic stroke, or symptomatic intracranial hemorrhage) was 1.8% (14/901; 95% CI 1.0% to 2.7%; I=0%). Aneurysm size ≥10 mm was a statistically significant predictor of a major complication (OR 6.4; 95% CI 2.0 to 20.7; p=0.002). Risk of a major complication in aneurysms <10 mm was 0.9% (95% CI 0.3% to 1.7%; I=0%). The meta-analysis was limited by the predominance of anterior circulation aneurysms.

Conclusion: Treatment of unruptured intracranial aneurysms using the Pipeline Flex flow diverter has a low periprocedural risk of death (0.8%) or major complication (1.8%). The risk of a major complication is significantly higher for large/giant aneurysms (4.4%) and is very low for aneurysms <10 mm (0.9%).

Download full-text PDF

Source
http://dx.doi.org/10.1136/neurintsurg-2019-014937DOI Listing

Publication Analysis

Top Keywords

pipeline flex
16
flex unruptured
12
unruptured intracranial
12
intracranial aneurysms
12
risk periprocedural
8
periprocedural death
8
death major
8
periprocedural
4
periprocedural complications
4
complications second-generation
4

Similar Publications

Traumatic intracranial pseudoaneurysm (PSA) is a rare yet dangerous subtype of cerebral aneurysm. The pipeline flex embolization device (PED Flex) has recently been applied to treat intracranial PSA. Herein, the authors report a middle-aged patient with intracranial subarachnoid hemorrhage and intracerebral hematoma after blunt cephalic trauma.

View Article and Find Full Text PDF

Single-cell transcriptomic profiling of eosinophils and airway immune cells in childhood asthma.

J Allergy Clin Immunol

July 2025

Center for Systems Immunology, Benaroya Research Institute at Virginia Mason, Seattle, Wash; Division of Allergy and Infectious Diseases, University of Washington School of Medicine, Seattle, Wash. Electronic address:

Background: Single-cell RNA sequencing has transformed our understanding of cellular heterogeneity but remains inadequate in capturing granulocytes, particularly in tissue compartments, owing to technical limitations.

Objective: To enhance granulocyte recovery in single-cell RNA sequencing, we used nasal lavage samples from children with asthma, leveraging the 10× Genomics Flex platform combined with a customized data processing pipeline.

Methods: Nasal lavage samples were processed without prior manipulation to avoid technical artifacts such as lysis or stimulation.

View Article and Find Full Text PDF

Dual antiplatelet therapy practices following Pipeline Shield embolization and their impact on adverse events in three large real-life registries.

J Neuroradiol

September 2025

Department of Interventional Neuroradiology - NEURI Brain Vascular Center, Bicêtre Hospital, Assistance Publique Hôpitaux de Paris, Paris-Saclay University, 78 Rue du General Leclerc, 94270 Le Kremlin-Bicêtre, France; INSERM U1195, Paris-Saclay University, Faculty of Medicine, 78 Rue du General L

Background: Dual Antiplatelet Therapy (DAPT) prescription following flow-diverter stent treatment of intracranial aneurysms is standard of care but evidence regarding the drugs of choice and duration of DAPT remains limited. We aim report to report practitioners' DAPT prescribing habits and to analysis their impact on adverse events in real-world use of Pipeline™ Flex with Shield Technology stent.

Methods: Data from 3 post-market core lab adjudicated studies were retrospectively analyzed.

View Article and Find Full Text PDF

The primate cerebral cortex relies on long-range connections to integrate information across spatially distributed and functionally specialized areas, yet tools for selectively modulating these pathways remain limited. Here, we present an optimized intersectional viral and optogenetic strategy for precisely exciting and inhibiting projection-specific neurons in the common marmoset. Building on a mouse-to-marmoset pipeline, we first validated that optogenetic activation of inhibitory neurons (via AAV9-Dlx-ChR2) enables robust local cortical inhibition.

View Article and Find Full Text PDF

Background: The Pipeline Flex embolization device with Shield technology (PED Shield, Medtronic, Irvine, CA, USA) is a flow diverter designed to reduce thrombogenicity through phosphorylcholine surface modification. Although in vitro and in vivo studies have demonstrated the thrombogenicity reducing effects of this technology, its effectiveness in real-world clinical practice remains unclear. This study aims to compare the number of post-procedure diffusion-weighted imaging (DWI)-positive lesions between PED Flex(Medtronic, Irvine, CA, USA) and PED Shield to assess the impact of surface modification.

View Article and Find Full Text PDF