Background: A new algorithm has been developed, which combines information on gait bradykinesia and dyskinesia provided by a single kinematic sensor located on the waist of Parkinson disease (PD) patients to detect motor fluctuations (On- and Off-periods).
Objective: The goal of this study was to analyze the accuracy of this algorithm under real conditions of use.
Methods: This validation study of a motor-fluctuation detection algorithm was conducted on a sample of 23 patients with advanced PD. Patients were asked to wear the kinematic sensor for 1 to 3 days at home, while simultaneously keeping a diary of their On- and Off-periods. During this testing, researchers were not present, and patients continued to carry on their usual daily activities in their natural environment. The algorithm's outputs were compared with the patients' records, which were used as the gold standard.
Results: The algorithm produced 37% more results than the patients' records (671 vs 489). The positive predictive value of the algorithm to detect Off-periods, as compared with the patients' records, was 92% (95% CI 87.33%-97.3%) and the negative predictive value was 94% (95% CI 90.71%-97.1%); the overall classification accuracy was 92.20%.
Conclusions: The kinematic sensor and the algorithm for detection of motor-fluctuations validated in this study are an accurate and useful tool for monitoring PD patients with difficult-to-control motor fluctuations in the outpatient setting.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5943625 | PMC |
http://dx.doi.org/10.2196/rehab.8335 | DOI Listing |
J Neuroeng Rehabil
September 2025
Institute for Neuromodulation and Neurotechnology, University Hospital and University of Tübingen, 72076, Tübingen, Germany.
Innovative technology allows for personalization of stimulation frequency in dual-site deep brain stimulation (DBS), offering promise for challenging symptoms in advanced Parkinson's disease (PD), particularly freezing of gait (FoG). Early results suggest that combining standard subthalamic nucleus (STN) stimulation with substantia nigra pars reticulata (SNr) stimulation may improve FoG outcomes. However, patient response and the optimal SNr stimulation frequency vary.
View Article and Find Full Text PDFFront Neurol
August 2025
Department of Neurology, Charité-Universitätsmedizin Berlin, Berlin, Germany.
Background: Gait deficits and leg spasticity are frequent symptoms in Primary and Secondary Progressive Multiple Sclerosis (PPMS and SPMS). Transcutaneous spinal cord stimulation (tSCS) may alleviate these symptoms through the reduction of spinal hyperexcitability. We conducted a single-center, randomized, sham-controlled clinical crossover study (German Clinical Trials Register: DRKS00023357, https://www.
View Article and Find Full Text PDFJ Pain
September 2025
Cyber-physical Health and Assistive Robotics Technologies Research Group, University of Nottingham, United Kingdom; School of Computer Science, University of Nottingham, Nottingham, United Kingdom.
Neck pain is among the most prevalent musculoskeletal conditions worldwide. The underlying cause mostly remains unidentified, classified as non-specific neck pain. Pain can alter movement patterns and physiological responses, suggesting that certain biomechanical and physiological changes may serve as objective biomarkers for non-specific neck pain.
View Article and Find Full Text PDFISA Trans
July 2025
Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun, 130033, China. Electronic address:
Coordinated operation of dual-arm manipulators is essential for enhancing the load capacity and adaptability of robotic systems. However, the precise control of the internal and external forces during the coordinated operation of dual-arm manipulators can pose considerable challenges owing to factors such as force interactions, kinematic constraints, positional inaccuracies, and external disturbances. This study focused on precise force-tracking control for a dual-arm manipulator system in the presence of external disturbances and uncertainties.
View Article and Find Full Text PDFACS Appl Mater Interfaces
September 2025
The Institute of Precision Machinery and Smart Structure, College of Engineering, Zhejiang Normal University, Jinhua, Zhejiang 321004, China.
Flexible sensors integrating motion detection and tactile perception capabilities demonstrate significant potential in aerospace biomechanics and medical rehabilitation. Here, we report a biomimetic inflatable chamber sensor that synergistically integrates pneumatic-auxiliary and electronic sensing for elbow joint health monitoring. The device architecture combines multiwalled carbon nanotube-reinforced silicone composites with embedded electrode arrays integrated within the inner lining of inflatable chambers, achieving high sensitivity while maintaining signal stability under electromagnetic interference.
View Article and Find Full Text PDF