Axial Length Measurement Failure Rates With Biometers Using Swept-Source Optical Coherence Tomography Compared to Partial-Coherence Interferometry and Optical Low-Coherence Interferometry.

Am J Ophthalmol

School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, China; Key Laboratory of Vision Science, Ministry of Health Peoples Republic of China, Wenzhou, China. Electronic address:

Published: January 2017


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To compare a new swept-source optical coherence tomography (SSOCT)-based biometer (OA-2000) with the IOLMaster v5.4 (partial-coherence interferometry) and Aladdin (optical low-coherence interferometry) biometers in terms of axial length measurement and failure rate in eyes with cataract.

Design: Reliability study.

Methods: A total of 377 eyes of 210 patients were scanned with the 3 biometers in a random order. For each biometer, the number of unobtainable axial length measurements was recorded and grouped as per the type and severity of cataract based on the Lens Opacities Classification System III by the same experienced ophthalmologist. The Bland-Altman limits-of-agreement (LoA) method was used to assess the agreement in axial length measurements between the 3 biometers.

Results: The failure rate was 0 eyes (0%) with the OA-2000, 136 eyes (36.07%) with the IOLMaster, and 51 eyes (13.53%) with the Aladdin. χ analyses indicated a significant difference in failure rate between all 3 devices (P < .001). Logistic regression analysis highlighted a statistically significant trend of higher failure rates with increasing severity of nuclear, cortical, and posterior subcapsular cataracts. Bland-Altman statistics indicated small mean differences and narrow LoA (OA-2000 vs IOLMaster -0.09 to 0.08 mm; OA-2000 vs Aladdin -0.10 to 0.07 mm; IOLMaster vs Aladdin -0.05 to 0.04 mm).

Conclusions: The OA-2000, a new SSOCT-based biometer, outperformed both the IOLMaster and Aladdin biometers in very advanced cataracts of various morphologies. The use of SSOCT technology may be the reason for the improved performance of the OA-2000 and may lead to this technology becoming the gold standard for the measurement of axial length.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2016.09.019DOI Listing

Publication Analysis

Top Keywords

axial length
20
failure rate
12
length measurement
8
measurement failure
8
failure rates
8
swept-source optical
8
optical coherence
8
coherence tomography
8
partial-coherence interferometry
8
optical low-coherence
8

Similar Publications

Objective: Our study aimed to compare premenopausal and postmenopausal women in terms of choroidal thickness and choroidal vascularity index.

Methods: This cross-sectional study included 96 eyes of 96 participants, comprising 48 premenopausal and 48 postmenopausal women. Enhanced depth image optical coherence tomography (EDI-OCT) was used to visualize the choroid.

View Article and Find Full Text PDF

Purpose: The purpose of this study was to investigate the focal relationship between choroidal thickness and retinal sensitivity in myopic eyes.

Methods: Participants underwent swept-source optical coherence tomography (SS-OCT) imaging and microperimetry testing. Choroidal thicknesses were obtained by segmenting the SS-OCT scans using a deep-learning approach.

View Article and Find Full Text PDF

Background And Aim: The incisive (nasopalatine) canal is an important anatomical structure of the anterior maxilla. It holds significance for surgeries and implant placement in the central incisor region. The size, shape, and relation with surrounding bones may vary by age, gender, and ethnicity.

View Article and Find Full Text PDF

Purpose: To compare refractive prediction accuracy using simulated keratometry (SimK) measurements obtained from a Scheimpflug tomographer (Pentacam AXL, Oculus) versus keratometry (K) measurements obtained from an optical biometer utilizing telecentric keratometry (IOLMaster 700 (IOLM700), Carl Zeiss Meditec AG) applied to modern IOL power calculation formulas.

Design: Retrospective accuracy and validity analysis METHODS: Setting: Private practice center STUDY POPULATION: Five hundred eighty-nine eyes with preoperative SimK and K measurements undergoing phacoemulsification and implantation of monofocal IOL (Clareon SY60WF IOL, Alcon Laboratories, Inc.).

View Article and Find Full Text PDF

Objective: This study evaluated changes in ocular parameters in patients with progressive myopia receiving 0.01% atropine.

Material And Methods: The study included 35 children (18 girls, 51.

View Article and Find Full Text PDF